| Literature DB >> 26904504 |
Haisong Liu1, David W Andrews2, James J Evans2, Maria Werner-Wasik1, Yan Yu1, Adam Paul Dicker1, Wenyin Shi1.
Abstract
OBJECTIVES: This study compares the dosimetry and efficiency of two modern radiosurgery [stereotactic radiosurgery (SRS)] modalities for multiple brain metastases [Gamma Knife (GK) and LINAC-based RapidArc/volumetric modulated arc therapy], with a special focus on the comparison of low-dose spread.Entities:
Keywords: Gamma Knife; RapidArc; SRS; VMAT; brain metastasis
Year: 2016 PMID: 26904504 PMCID: PMC4749694 DOI: 10.3389/fonc.2016.00026
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Prescription dose and tumor volumes of each target in this study.
| Patient no. | No. of targets | Total tumor vol. (cc) | Tumor vol. (cc) | Prescription dose (Gy) |
|---|---|---|---|---|
| 1 | 3 | 5.32 | 2.53 | 16 |
| 0.81 | 24 | |||
| 1.98 | 20 | |||
| 2 | 3 | 2.08 | 0.41 | 24 |
| 0.51 | 24 | |||
| 1.16 | 18 | |||
| 3 | 3 | 1.19 | 0.41 | 24 |
| 0.52 | 24 | |||
| 0.26 | 24 | |||
| 4 | 3 | 5.13 | 4.62 | 18 |
| 0.12 | 24 | |||
| 0.39 | 24 | |||
| 5 | 3 | 11.14 | 10.51 | 15 |
| 0.35 | 24 | |||
| 0.28 | 24 | |||
| 6 | 4 | 1.70 | 0.68 | 24 |
| 0.56 | 24 | |||
| 0.46 | 24 | |||
| 0.4 | 24 |
Planning parameters for Gamma Knife Perfexion plans.
| Patient no. | No. of targets | Number of shots | Beam-on time (min) | Est. total tx time (min) |
|---|---|---|---|---|
| 1 | 3 | 43 | 83.5 | 100 |
| 2 | 3 | 15 | 72.2 | 87 |
| 3 | 3 | 4 | 47 | 56 |
| 4 | 3 | 19 | 70 | 84 |
| 5 | 3 | 26 | 93 | 112 |
| 6 | 4 | 10 | 64 | 77 |
Planning parameters for single-isocenter VMAT plan.
| Patient no. | No. of targets | No. of arcs | Table angles (Varian IEC Scale) | Arc length | Monitor units (MU) | Beam-on time (min) | Est. total tx time (min) |
|---|---|---|---|---|---|---|---|
| 1 | 3 | 4 | 45, 0, 335, 300 | 170, 120, 140, 140 | 7000 | 5.0 | 15 |
| 2 | 3 | 6 | 70, 40, 10, 0, 335, 305 | 160, 150, 40, 130, 160, 150 | 8300 | 5.9 | 18 |
| 3 | 3 | 6 | 50, 25, 0, 340, 315, 290 | 160, 160, 130, 140, 140, 150 | 9600 | 6.9 | 21 |
| 4 | 3 | 5 | 80, 45, 20, 353, 300 | 140, 140, 150, 140, 150 | 10,130 | 7.2 | 22 |
| 5 | 3 | 5 | 60, 20, 345, 310, 270 | 120, 120, 120, 140, 140 | 8660 | 6.2 | 19 |
| 6 | 4 | 5 | 70, 40, 10, 0, 300 | 170, 170, 170, 140, 140 | 9750 | 7.0 | 21 |
Figure 1Isodose distribution of a single-isocenter VMAT (RapidArc) SRS plan (A) and Gamma Knife (B) Perfexion plan for the same patient (patient #1). Isodose lines are 24, 20, 16, 12, 6, and 3 Gy. (A) Isodose distribution of a single-isocenter VMAT (RapidArc) SRS plan for patient #1. (B) Isodose distribution of Gamma Knife Perfexion SRS plan for patient #1.
Figure 2Dose–volume histogram (DVH) comparison of a single-isocenter VMAT (RapidArc) SRS plan and Gamma Knife Perfexion plan for the same patient (patient #1) showing background dose to normal tissue and DVH for the 16, 20, and 24 Gy treatment of the larger to the smaller target, respectively.
Comparison of conformity index and normal tissue .
| Normal tissue | ||||||||
|---|---|---|---|---|---|---|---|---|
| Patient # | Target # | Target vol. (cc) | Conformity index | Gradient index | ||||
| GKP | RPA | GKP | PRA | GKP | RPA | |||
| 1 | 1 | 2.53 | 1.62 | 1.15 | 3.42 | 4.28 | 4.80 | 3.45 |
| 2 | 0.81 | 1.63 | 1.17 | 3.75 | 4.63 | 4.14 | 3.59 | |
| 3 | 1.98 | 1.60 | 1.14 | 3.21 | 3.92 | 5.76 | 4.51 | |
| 2 | 4 | 0.41 | 1.27 | 1.12 | 3.46 | 5.04 | 1.39 | 1.91 |
| 5 | 0.51 | 1.43 | 1.14 | 2.82 | 4.55 | 1.55 | 2.13 | |
| 6 | 1.16 | 1.78 | 1.38 | 2.75 | 4.54 | 2.72 | 3.21 | |
| 3 | 7 | 0.41 | 1.54 | 1.15 | 2.56 | 4.68 | 1.20 | 1.91 |
| 8 | 0.52 | 1.35 | 1.10 | 3.71 | 4.25 | 2.08 | 2.19 | |
| 9 | 0.26 | 1.58 | 1.23 | 3.88 | 5.34 | 1.33 | 1.54 | |
| 4 | 10 | 4.62 | 1.29 | 1.15 | 3.64 | 3.69 | 7.23 | 6.48 |
| 11 | 0.12 | 1.83 | 1.67 | 3.32 | 10.30 | 0.61 | 1.94 | |
| 12 | 0.39 | 1.51 | 1.26 | 2.76 | 5.14 | 1.24 | 2.13 | |
| 5 | 13 | 10.51 | 1.36 | 1.07 | 2.69 | 2.58 | 8.64 | 5.34 |
| 14 | 0.35 | 1.49 | 1.26 | 4.60 | 5.20 | 2.04 | 1.94 | |
| 15 | 0.28 | 1.68 | 1.21 | 6.40 | 5.38 | 2.73 | 1.55 | |
| 6 | 16 | 0.68 | 1.41 | 1.09 | 4.38 | 4.07 | 3.52 | 2.33 |
| 17 | 0.56 | 1.46 | 1.11 | 4.95 | 4.55 | 3.50 | 2.26 | |
| 18 | 0.46 | 1.41 | 1.17 | 4.45 | 4.19 | 2.43 | 1.80 | |
| 19 | 0.4 | 1.37 | 1.00 | 2.67 | 4.30 | 1.28 | 1.58 | |
| Mean | 1.50 | 1.19 | 3.65 | 4.77 | 3.06 | 2.73 | ||
| SD | 0.16 | 0.14 | 0.98 | 1.49 | 2.20 | 1.38 | ||
| <0.001 | 0.01 | 0.58 | ||||||
Dosimetric parameters of Gamma Knife SRS and RapidArc SRS radiation treatment plans in this study.
| Gamma Knife | RapidArc | ||||
|---|---|---|---|---|---|
| Mean | Std. | Mean | Std. | ||
| V12 (patient composite) | 10.85 | 7.2 | 9.7 | 5.1 | 0.63 |
| V6 | 36.9 | 16.9 | 36.3 | 14.7 | 0.96 |
| V4.5 | 86.7 | 29.8 | 99 | 27.3 | 0.15 |
| V3 | 160.8 | 55.7 | 224 | 53 | 0.1 |
| D100 | 4 | 0.9 | 4.6 | 0.6 | 0.3 |
| D200 | 2.6 | 0.6 | 3.2 | 0.5 | 0.06 |
| D300 | 2 | 0.4 | 2.6 | 0.4 | <0.05 |
| Dspread | 1.9 | 0.73 | 2.5 | 0.2 | 0.01 |
| Beam-on time | 71.6 | 15.9 | 6.4 | 0.8 | <0.01 |
| Est. total tx time | 85.9 | 19.1 | 19.3 | 2.6 | <0.01 |