Early Outcomes of Revision Surgery for Taper Corrosion of Dual Taper Total Hip Arthroplasty in 187 Patients.

Dimitris Dimitriou1, Ming Han Lincoln Liow1, Tsung-Yuan Tsai1, William A Leone2, Guoan Li1, Young-Min Kwon1.   

Abstract

BACKGROUND: Modular dual taper femoral neck designs have been associated with taper corrosion requiring revision surgery. However, outcomes after revision dual taper total hip arthroplasty in patients with symptomatic adverse local tissue reaction due to taper corrosion remain largely unknown.
METHODS: A total of 198 revision surgeries in 187 patients with dual taper femoral stem total hip arthroplasty with minimum 12-month follow-up were evaluated.
RESULTS: At mean follow-up of 18 months, at least 1 complication had occurred in 39 patients (20%) of 198 revisions. Single episode of dislocation, treated with close reduction, occurred in 16, whereas 2 patients required rerevision due to multiple dislocations. Infection requiring rerevision occurred in 3 patients. Adverse local tissue reaction recurrence requiring reoperation occurred in 6 patients. Implant survivorship for revision for any cause was 86% at 30 months. The reoperation rate of revised dual taper was 8% (16 out of 198 hips). The median serum levels of cobalt, chromium, and cobalt/chromium ratio decreased (P < .01) from 5.3 μg/L (range: 2.3-48.5 μg/L), 2.6 μg/L (range: 0.2-64 μg/L), and 4.7 (range: 2.1-35) prerevision to 1.4 μg/L (range: 0.2-8.8 μg/L), 0.7 μg/L (range: 0.1-3.9 μg/L), and 2.2 (range: 0.4-8.8) postrevision, respectively.
CONCLUSION: This pilot study demonstrates that intraoperative tissue necrosis was associated with a high rate of early complications (20%) and revisions (8%), suggesting the importance of a systematic evaluation of these patients including metal ion levels and metal artifact reduction sequence magnetic resonance imaging in optimizing revision outcome, as early diagnosis will facilitate the initiation of appropriate treatment before significant adverse tissue necrosis.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  clinical outcomes; dual taper; modular femoral neck stem; revision total hip arthroplasty; systematic evaluation; taper corrosion

Mesh:

Substances:

Year:  2016        PMID: 26895822     DOI: 10.1016/j.arth.2016.01.015

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  11 in total

1.  Long-term survivorship of an exchangeable-neck hip prosthesis with a Ti-alloy/Ti-alloy neck-stem junction.

Authors:  Massimiliano Baleani; Aldo Toni; Cristina Ancarani; Susanna Stea; Barbara Bordini
Journal:  Arch Orthop Trauma Surg       Date:  2022-09-30       Impact factor: 2.928

2.  Projections and Epidemiology of Revision Hip and Knee Arthroplasty in the United States to 2030.

Authors:  Andrew M Schwartz; Kevin X Farley; George N Guild; Thomas L Bradbury
Journal:  J Arthroplasty       Date:  2020-02-19       Impact factor: 4.757

3.  Pseudotumor recurrence in a post-revision total hip arthroplasty with stem neck modularity: A case report.

Authors:  Bhumit R Desai; Gonzalo E Sumarriva; George F Chimento
Journal:  World J Orthop       Date:  2020-02-18

Review 4.  Systemic and local toxicity of metal debris released from hip prostheses: A review of experimental approaches.

Authors:  Divya Rani Bijukumar; Abhijith Segu; Júlio C M Souza; XueJun Li; Mark Barba; Louis G Mercuri; Joshua J Jacobs; Mathew Thoppil Mathew
Journal:  Nanomedicine       Date:  2018-01-12       Impact factor: 5.307

5.  Outcome of revision surgery for adverse local tissue reactions in patients with recalled total hip arthroplasty.

Authors:  Christian Klemt; Sakkadech Limmahakhun; Georges Bounajem; Christopher M Melnic; Michael J Harvey; Young-Min Kwon
Journal:  Arch Orthop Trauma Surg       Date:  2021-04-09       Impact factor: 2.928

6.  Outcomes following revision surgery performed for adverse reactions to metal debris in non-metal-on-metal hip arthroplasty patients: Analysis of 185 revisions from the National Joint Registry for England and Wales.

Authors:  G S Matharu; A Judge; D W Murray; H G Pandit
Journal:  Bone Joint Res       Date:  2017-07       Impact factor: 5.853

7.  Exchangeable Femoral Neck (Dual-Modular) THA Prostheses Have Poorer Survivorship Than Other Designs: A Nationwide Cohort of 324,108 Patients.

Authors:  Sandrine Colas; Assia Allalou; Antoine Poichotte; Philippe Piriou; Rosemary Dray-Spira; Mahmoud Zureik
Journal:  Clin Orthop Relat Res       Date:  2017-02-13       Impact factor: 4.176

8.  Revision surgery of metal-on-metal hip arthroplasties for adverse reactions to metal debris.

Authors:  Gulraj S Matharu; Antti Eskelinen; Andrew Judge; Hemant G Pandit; David W Murray
Journal:  Acta Orthop       Date:  2018-03-01       Impact factor: 3.717

9.  The Chemical Form of Metal Species Released from Corroded Taper Junctions of Hip Implants: Synchrotron Analysis of Patient Tissue.

Authors:  Anna Di Laura; Paul D Quinn; Vasiliki C Panagiotopoulou; Harry S Hothi; Johann Henckel; Jonathan J Powell; Fitim Berisha; Fernanda Amary; J Fred W Mosselmans; John A Skinner; Alister J Hart
Journal:  Sci Rep       Date:  2017-09-08       Impact factor: 4.379

10.  Cobalt (II) ions and nanoparticles induce macrophage retention by ROS-mediated down-regulation of RhoA expression.

Authors:  Jing Xu; Junyao Yang; Agata Nyga; Mazdak Ehteramyan; Ana Moraga; Yuanhao Wu; Lingfang Zeng; Martin M Knight; Julia C Shelton
Journal:  Acta Biomater       Date:  2018-04-09       Impact factor: 8.947

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.