Literature DB >> 26894364

Automatic planning of head and neck treatment plans.

Irene Hazell1, Karl Bzdusek, Prashant Kumar, Christian R Hansen, Anders Bertelsen, Jesper G Eriksen, Jørgen Johansen, Carsten Brink.   

Abstract

Treatment planning is time-consuming and the outcome depends on the person performing the optimization. A system that automates treatment planning could potentially reduce the manual time required for optimization and could also provide a method to reduce the variation between persons performing radiation dose planning (dosimetrist) and potentially improve the overall plan quality. This study evaluates the performance of the Auto-Planning module that has recently become clinically available in the Pinnacle3 radiation therapy treatment planning system. Twenty-six clinically delivered head and neck treatment plans were reoptimized with the Auto-Planning module. Comparison of the two types of treatment plans were performed using DVH metrics and a blinded clinical evaluation by two senior radiation oncologists using a scale from one to six. Both evaluations investigated dose coverage of target and dose to healthy tissues. Auto-Planning was able to produce clinically acceptable treatment plans in all 26 cases. Target coverages in the two types of plans were similar, but automatically generated plans had less irradiation of healthy tissue. In 94% of the evaluations, the autoplans scored at least as high as the previously delivered clinical plans. For all patients, the Auto-Planning tool produced clinically acceptable head and neck treatment plans without any manual intervention, except for the initial target and OAR delineations. The main benefit of the method is the likely improvement in the overall treatment quality since consistent, high-quality plans are generated which even can be further optimized, if necessary. This makes it possible for the dosimetrist to focus more time on difficult dose planning goals and to spend less time on the more tedious parts of the planning process.

Entities:  

Mesh:

Year:  2016        PMID: 26894364      PMCID: PMC5690191          DOI: 10.1120/jacmp.v17i1.5901

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

A number of uncertainties and variations are present in radiotherapy such as absolute dose precision, delivery precision, , , , precision of calculated dose distributions, , , , and radioresponsiveness of the specific tumor and normal tissues. , , , , , Two of the largest variations within radiotherapy are the heterogeneity in target definition , , and the variation among treatment plans for a given geometry both intra‐ and interinstitutional. , Most treatment plans are likely to have sufficient dose coverage of the delineated targets, but large variations in dose to healthy tissues occur. The dose distribution depends on the dose objective defined by the dosimetrist, typically in accordance with institution‐specific guidelines. However, even guidelines do not ensure an optimal dose distribution for the specific anatomy, since the lower achievable dose limit to an OAR for a specific patient is unknown. This is the reason why treatment plans are optimized for the individual patient by trained dosimetrists. Moreover, the treatment optimization is labor‐intensive work with a very large solution space, which makes it difficult to ensure that the clinical treatment plan is the optimal plan. Therefore, there is a need to automate the treatment planning optimization procedure both to reduce the amount of time spent on the optimization and, more importantly, to reduce the interdosimetrist variation. If an automatically generated treatment plan of high clinical quality is available prior to manual optimization, it could serve as a quality reference and starting point for the specific treatment and thereby ensure a certain minimum quality. Furthermore, the automatic plans could potentially be a time‐saving tool during the treatment optimization, which would reduce one of the most tedious steps in the process. Sharing the optimization parameters between institutions could also provide a method to share knowledge and standardize plan quality. Previous documented solutions with somewhat different approaches have shown the potential of automation of the planning process. , , , , The current study validates the performance of a prototype version of the Auto‐Planning module which recently has been productized for clinical use in the Pinnacle3 treatment planning system from Philips Healthcare (Fitchburg, WI).

MATERIALS AND METHODS

The Auto‐Planning software was evaluated by replanning 26 previously delivered clinical head and neck IMRT treatment plans of the oropharynx. The plans were delivered over the 12 months prior to the study. The plans were created in accordance with the Danish Head And Neck Cancer Groups guidelines (DAHANCA ‐ Version 2004), and each dose plan included three dose levels of 50 Gy, 60 Gy, and 66 or 68 Gy in 33 treatment fractions with a simultaneous integrated boost technique. In the Auto‐Planning software, a template of configurable parameters known as a Technique (details in Appendix A) can be defined for each treatment protocol. The Techniques include definition of beam parameters and planning goals. The Auto‐Planning module uses the Technique definition to iteratively adjust IMRT planning parameters to best meet the planning goals. The Technique was defined according to local standards, including prioritization between target coverage and dose to organs at risk. The Technique definition was based on five additional pilot patients independent of the 26 study patients. Each of the 26 treatment plans were replanned with Auto‐Planning without knowledge of the clinically delivered treatment plans and without any dosimetrist postoptimization of the treatment plans. The only input to the replanning was the delineations of planning target and organs at risk and the positioning of the isocenter. Quantitative dosimetric evaluation of the performance of the treatment plans was performed on dose volume histograms (DVH) extracted from the planning system. CT scans had a slice thickness of 3 mm and in plane voxel size of . The dose plans were calculated using the Pinnacle3 collapsed cone algorithm with a dose grid resolution of 3 mm. Specific DVH values, as well as the overall shape of the DVH, was compared using average DVHs of the two types of treatment plans. The average DVH was calculated for each type of treatment plan as the average of the patient‐specific DVH values at each dose level. To specify dose regions for which statistically significant differences exist, a probability curve as in Bertelsen et al. was calculated. In short, the probability curve is a Wilcoxon matched‐pair, signed‐rank test performed at each dose level. Individual values of the curve are not strict statistical tests since the test is performed multiple times and on values that are not mutually independent. The probability curve is, therefore, primarily a tool that indicates regions for which the average DVHs deviate significantly. The DVH analysis was performed for all target volumes, as well as for parotid gland, submandibular gland, and spinal cord. To evaluate dose to other normal structures than the delineated, a DVH evaluation of all healthy tissues outside the PTV was performed. In contrast to the previous DVH evaluations, this evaluation was performed in absolute volume to compensate for a difference of the CT scanned volume of each patient (relative values would depend on the scanned volume). The dosimetric evaluation was extended with a blinded clinical evaluation of the treatment plans. Two senior head and neck radiation oncologist independently scored the treatment plans on a categorical scale from 1 to 6 ( and ). The scoring was performed for target coverage, sparing of healthy tissues, and an overall assessment of the treatment plan. Finally, based on the clinical evaluation, the radiation oncologists selected the plan they would favor for clinical treatment. The radiation oncologist had access to all clinical information and diagnostic scans. The evaluation was performed similar to the clinical procedure used for evaluation of two different proposals for a given clinical treatment. All information related to the production of the plans was blinded and the treatment plans were presented for evaluation in random order. All statistical tests were made by Wilcoxon matched‐pair, signed‐rank test using a significance level of 5%.

RESULTS

Average DVH comparison between automatic and the clinical delivered plans for the targets is shown in Fig. 1. In accordance with ICRU 83, PTV50 includes PTV60 and PTV 66/68 causing the long tail towards high doses. Likewise, PTV60 contains PTV66/68. For PTV50, the autoplans had a small, but statistically significant, lower dose than the clinical plans below 95% of the 50 Gy and also above of the 50 Gy.
Figure 1

Mean DVHs of the target volumes for the clinical (red) and auto (blue) plans normalized to the prescribed doses for each PTV. P‐value curves are shown in gray.

The PTV60 had a similar pattern as the PTV50 below of 60 Gy. A slightly higher average dose for the autoplans was also observed, while no differences were seen for higher doses. For the highest dose level, PTV66/68, the trend of a small, but statistical lower, dose below 95% of prescribed dose was observed for the autoplans, while no other differences were found. The observations from Fig. 1 are reflected in Table 1 which shows selected DVH parameters. The difference in steepness of the DVH for PTV50 is statistically significant, with the automatic plans having the steeper slope. Also, the minimum dose was different for the two types of plans, with the clinical plans having a slightly higher minimum dose. Nevertheless, for both types of plans, the average DVH covers the PTV50 with the 95% isodose lines for at least 98.5% of the target which meets the ICRU 83 recommendation, as well as the 2013 DAHANCA guidelines and QA evaluations. For PTV60, the only statistically significant differences observed were a slightly higher dose level for the automatic plans (Table 1). For the highest dose level, the automatic plans produce more conformal dose distributions (Conformity Index CI 95%), but slightly less maximum and minimum dose.
Table 1

Mean values of standard DVH parameters for PTVs. Mean dose and are percent of the dose prescribed to the specific PTV. The standard deviations (SD) of the dose to PTVs are in percent points of the prescribed dose and are in volume fractions. CI 95% is defined as the ratio of the 95%‐isodose‐volume and the volume of the PTV. Uncertainties are reported as 1 SD, except for the median volume where ranges are reported. Statistically significant differences are shown in bold.

PTV50 PTV60 PTV66/68
Parameter Clinical Auto P Clinical Auto P Clinical Auto P
Median volume (cm3)266 (91‐323)90 (36‐175)80 (20‐264)
Mean dose (%) 105.17±2.28 104.50±2.88 0.091 103.99±1.27 104.90±1.68 0.034 100.29±0.74 99.70±1.24 0.058
SD of dose to PTV (%) 6.98±2.15 5.92±1.52 <0.001 3.99±0.66 3.89±0.69 0.209 2.12±0.42 1.93±0.63 0.124
D5%‐D95% (%) 23.15±7.19 19.48±5.37 <0.001 13.06±2.15 12.52±2.21 0.078 6.89±1.33 6.18±1.99 0.078
D2% (%) 124.81±7.17 122.68±6.11 0.006 111.79±2.31 112.02±2.30 0.585 104.29±1.24 103.16±1.63 0.012
D50% (%) 103.37±1.60 103.17±2.70 0.469 104.13±1.35 105.06±1.70 0.023 100.33±0.77 99.84±1.20 0.159
D98% (%) 96.10±0.72 96.46±2.74 0.694 96.03±0.92 96.56±2.25 0.182 95.93±0.89 95.09±2.36 0.182
V95% 0.992±0.006 0.985±0.014 0.001 0.990±0.008 0.987±0.014 0.869 0.992±0.009 0.969±0.047 0.002
V107% 0.249±0.132 0.209±0.183 0.096 0.241±0.124 0.316±0.168 0.078 0.001±0.003 0.001±0.004 0.966
CI95% 1.48±0.15 1.42±0.23 1.000 1.64±0.29 1.65±0.15 0.844 1.74±0.20 1.28±0.17 0.031
Mean DVHs of the target volumes for the clinical (red) and auto (blue) plans normalized to the prescribed doses for each PTV. P‐value curves are shown in gray. Mean values of standard DVH parameters for PTVs. Mean dose and are percent of the dose prescribed to the specific PTV. The standard deviations (SD) of the dose to PTVs are in percent points of the prescribed dose and are in volume fractions. CI 95% is defined as the ratio of the 95%‐isodose‐volume and the volume of the PTV. Uncertainties are reported as 1 SD, except for the median volume where ranges are reported. Statistically significant differences are shown in bold. Average DVH curves for the organs at risk (OAR) are shown in Fig. 2. For all the delineated organs, the average DVH doses from the autoplans are either less than or equal to those of the clinical plans. There are no dose ranges for which the autoplans generated statistically significant higher doses to the delineated OAR.
Figure 2

Mean DVHs of the OAR for the clinical (red) and auto (blue) plans. P‐value curves are shown in gray. Figure 2(f) show the average absolute volume difference in cm3 between the DVHs for the healthy tissue for the clinical minus the autoplans.

To evaluate dose to nonde‐lineated healthy tissues, Fig. 2(f) shows the DVH difference for all tissues outside the PTV. The figure demonstrates the difference of the absolute volume irradiated above a specific dose level for the clinical plans minus the automatic plans. For all dose levels up to , the dose levels outside the PTVs were lower for the automatic plans. Above , the two types of plans were equal to each other. To quantify the DVH differences and their variation, selected DVH metrics for the organs at risk are shown in Table 2.
Table 2

Mean values for doses to organs at risk. Uncertainties of all values are reported as 1 SD. Statistically significant differences are shown in bold.

Clinical Auto p
Spinal Cord
Mean (Gy) 26.03±5.13 21.13±4.55 <0.001
Max (Gy) 42.35±1.56 39.60±3.62 <0.001
D20% (Gy) 40.50±1.45 36.23±3.60 <0.001
V35Gy 0.46±0.19 0.09±0.10 <0.001
Submandibular Contralateral
Mean (Gy) 47.33±7.66 40.41±6.87 <0.001
V30Gy 0.94±0.20 0.79±0.19 <0.001
V39Gy 0.85±0.25 0.56±0.21 <0.001
V60Gy 0.08±0.16 0.05±0.12 0.074
Submandibular Ipsilateral
Mean (Gy) 57.62±6.78 54.54±9.18 0.001
V30Gy 0.99±0.03 0.92±0.11 <0.001
V39Gy 0.94±0.12 0.84±0.20 <0.001
V60Gy 0.53±0.33 0.49±0.30 0.072
Parotid Contralateral
Mean (Gy) 30.48±5.71 26.55±4.02 <0.001
V5Gy 0.96±0.07 0.94±0.08 0.033
V26Gy 0.60±0.19 0.46±0.11 <0.001
V35Gy 0.42±0.16 0.30±0.12 <0.001
Parotid Ipsilateral
Mean (Gy) 35.54±6.84 29.73±5.57 <0.001
V5Gy 0.96±0.09 0.94±0.12 0.007
V26Gy 0.71±0.19 0.51±0.11 <0.001
V35Gy 0.52±0.16 0.37±0.12 <0.001
The overall result in Table 2 is as stated above, that the OAR receives less dose using the automatic generated plans than the previously delivered plans. Part of the observed dose differences was related to interdosimetrist variation. Figure 3 shows an example of all pairs of DVH for ipsilateral parotid, which illustrates that the autoplans are confined to a narrower region than the manual plans.
Figure 3

Pairs of DVH for ipsilateral parotid, illustrating the less variation of autoplans.

The evaluations by the oncologists are shown in Table 3.
Table 3

Clinical evaluation of the auto and clinical plans on a score from one to six ( – ). The observers scores are shown in each columns separated by a slash. At the lower part of the table, mean values for both the individual observers and a combined score is shown together with a test of statistically significant differences between the score for auto and clinical plans. For Patient 16 please see note in text. Statistically significant differences are shown in bold.

Target Coverage Organs at Risk Overall Quality Selected Plan
Patient Clinical Auto Clinical Auto Clinical Auto Clin=0;Auto=1
15/66/54/55/65/56/51/0
24/55/56/45/65/45/61/1
36/56/65/46/65/56/61/1
46/66/65/56/65/66/61/1
56/66/55/56/65/66/61/0
66/56/64/55/65/56/61/1
76/56/65/56/65/56/6NA/1
86/66/55/56/66/56/51/0
95/66/55/55/65/66/61/0
106/55/64/36/54/35/51/1
116/66/64/56/65/66/61/1
126/66/64/56/65/66/61/1
136/66/54/56/65/66/61/0
145/65/54/56/65/66/61/0
156/66/64/56/65/66/61/1
166/65/54/66/15/66/11/0
175/55/64/45/6NA/4NA/61/1
186/65/44/55/54/65/51/0
195/66/64/66/65/66/61/0
205/65/64/66/65/66/61/1
215/66/54/66/65/65/61/0
226/66/65/56/65/66/61/1
236/66/55/56/65/56/51/0
246/45/35/56/66/46/41/0
256/66/55/56/65/66/61/0
266/NA6/NA4/NA6/NA5/NA6/NA1/NA
Mean Observ5.65/5.685.69/5.364.46/4.965.77/5.725.00/5.405.84/5.521.00/0.48
Mean Total5.675.534.715.755.205.680.74
P Observ0.739/0.074 <0.001/0.001 <0.001/0.285
P total0.194 <0.001 <0.001
Mean DVHs of the OAR for the clinical (red) and auto (blue) plans. P‐value curves are shown in gray. Figure 2(f) show the average absolute volume difference in cm3 between the DVHs for the healthy tissue for the clinical minus the autoplans. Except for Patient 16, the main finding is that the qualities of both types of plans are good. Patient 16 failed to meet the maximum dose goal to the spinal cord which has a very high clinical priority, but for the other targets and organ delineations, the plan was evaluated as good. In terms of target coverage, the scores were not statistically significantly different between the two types of plans for any of the two observers; however, a p‐value of 0.079 for one observer could indicate a preference for the clinical plans in term of target coverage. The clinical evaluation of the dose to the OARs was clearly in favor of the autoplans for both observers, which is consistent with the dosimetric observations in Fig. 2 and Table 2. For the overall evaluation of the plans, one observer is clearly in favor of the autoplans, while the other observer showed no statistically significant differences. In 94% of the cases, the overall score for the autoplans were as good as, or better than, the clinical plans. The last column of Table 3 shows the preferred treatment plans by the clinicians. The column reflects the overall plan score, and shows that the observer with a difference in the overall evaluation only selected automatic plans while the observer who did not find a significant overall difference selected the automatic plans and the clinical plans evenly. Mean values for doses to organs at risk. Uncertainties of all values are reported as 1 SD. Statistically significant differences are shown in bold. Pairs of DVH for ipsilateral parotid, illustrating the less variation of autoplans. Clinical evaluation of the auto and clinical plans on a score from one to six ( – ). The observers scores are shown in each columns separated by a slash. At the lower part of the table, mean values for both the individual observers and a combined score is shown together with a test of statistically significant differences between the score for auto and clinical plans. For Patient 16 please see note in text. Statistically significant differences are shown in bold.

DISCUSSION

For all 26 patients it was possible to produce automatic plans of high plan quality. Small differences in the PTV dose coverage but a significant reduction in dose to OAR between automatic plans and clinical plans indicate that the Technique parameters used in the current study are biased towards normal tissue sparing relative to the clinical practice. It is likely that another set of Technique parameters could be determined which would focus more on dose coverage. In addition, in the clinical release of the Auto‐Planning software, modifications have been made to enhance the priority of maximum dose constraints to OAR such as the cord, a need highlighted by the physicians during plan evaluation. No postoptimization of the plans was performed in the study to make a pure validation of the Auto‐Planning software. If needed or requested, in a clinical situation automatic optimized plans could be further optimized just as any manually created plan since the automatic plans are delivered in exactly the same format as manually created plans. Thus, the automatic plans can either be a high quality starting point for further manual optimization or an attempt to produce plans of clinical quality without further user intervention. It is likely that the reduced dose to normal tissue of the sizes seen in Fig. 2 could be of clinical impact for the patients, while the small difference in tumor coverage will almost certainly have no clinical impact. However, without no definitive answer to which plan is the better in terms of tumor control and sparing of normal tissue (e.g., based on large randomized trials), the evaluation of the clinical quality of the plans will be somewhat subjective and depend upon individual views of the oncologist judging the plans. In the current study there are indications that the two oncologists have slight differences in their priorities. As stated in the results, both oncologists scores the OAR irradiation significantly better for the automatic plans compared to the manual plans, while none of the oncologists found statistically significant differences between the automatic and manual plans in terms of target coverage; although a p‐value of .074 for one of the oncologist indicates a favor of target coverage from the manual plans. These differences are also reflected in selection of treatment plan in which one oncologist only select the automatic plans while the other select an even mixture of automatic and manual plans. However, independent of the interoncologist variations, it is interesting to observe that in 94% of the cases (all except for one) the overall score given to the automatic plan is at least as high as the score for the manual plan, indicating that most of the plans could be used clinically without any user intervention. Only for one plan, one of the oncologists scored the automatic plan lower than the manual plan due to a violation of the maximum dose constraint of the spinal cord. With slight manual effort, this single treatment plan was later optimized to have an acceptable maximum spinal cord dose. A prerequisite to achieve high‐quality automatic plans compared to manually created plans is obviously a sound optimization and tuning algorithm (the Auto‐Planning engine). However, the quality of the manual plans is obviously also of importance in comparison with the automatic plans (poor quality manual plan would favor automatic plans). At the time the manually created plans were made, all clinical head and neck treatment plans in the department were created using IMRT; thus the department was experienced in creating “high quality” plans manually. As a result, given the significant experience with IMRT in the department, the interdosimetrist variation should be limited. Nevertheless, as seen in Fig. 3, there is quite a variation in manually optimized plans, a variation which is less for the automatic optimization. It therefore seems likely that one of the advantages of automatic plans is a reduction of the interdosimetrist variation which is present even within departments that use IMRT extensively. The reason of the interdosimetrist variation is related both to limited time to create the plan and to lack of knowledge of how much the plan in reality can be optimized. For a manually created plan, it is difficult to know the extent an OAR can be spared prior to actual plan optimization. Therefore, objectives for organs at risk are typically set relatively loose initially, in order to ensure dose coverage of the target. Having obtained dose coverage of target, the next step is to reduce the dose to organs at risk as much as possible. In a busy clinic, it can be hard to ensure that all constraints on organs at risk have been tightened as much as possible. This issue could be reduced significantly if an initial “high quality” plan — e.g., an automatic generated plan — were available such that the dosimetrist could focus on fine tuning of the treatment plan. Another potential benefit of Auto‐Planning could be a simple method to exchange planning knowledge and procedures between institutions since the Technique configuration of the Auto‐Planning software can easily be shared between institutions. This could help institutions with, for example, limited resources to quickly create IMRT or VMAT plans with similar quality as in more advanced institutions. Most previous work on automating the planning process has built on knowledge of previously treated patients. One approach of extracting information from previously treated patients is utilizing the overlap volume histogram method, which measures the position of an OAR relative to the target. , , , Knowledge of overlap volume histograms from previously treated patients can be used to predict the likely achievable irradiation level of specific OARs. A few published solutions on automating the planning process have been documented, , , , , and all build on the Pinnacle3 planning system. The published systems did show the feasibility of automating the planning process. However, in terms of flexibility, it could be a potential issue that “knowledge based” approaches require a database of “high” quality plans for each protocol. Changes to planning techniques, prescriptions, OAR sparing goals, and contouring style could, if not implemented in a very smart way, require a new “high” quality database. Such a change could be quite labor‐intensive to implement clinically, and might not be as flexible to interchange between institutions. This issue might be addressed within “knowledge based” algorithms, but is not present in the Auto‐Planning solution evaluated in this study since it only relies on a small set of Technique parameters. Finally, it should be mentioned that plan comparison studies are inherently difficult to perform since development of the treatment planning skill is continually ongoing within any department in order to optimize the treatments plans. Thus, if the current study was repeated, the results might be different since our dose planning team has learned new ways to improve the quality based on the results of this study. Similar statements could be made about the configuration of the automatic system. However, this does not change the fact that the current comparison between treatment plans that have been delivered clinically and the automatic treatment plans did show the autoplans to be superior at that time. Thus, the impact of Auto‐Planning seems likely to be a tool to increase the overall quality of dose planning, rather than a tool that could remove the need of manual optimization.

CONCLUSIONS

Comparison of autoplans and previous delivered clinical plans showed only small dosimetric differences in target coverage, but significant reduction in dose to OAR for the autoplans. The blinded clinical evaluation of the plans showed that, for 94% of the evaluations, the autoplans were similar to or better than the clinical plans. Auto‐Planning software will, therefore, be able to reduce the manual time spend per treatment plan since the most of the plans could potentially be used clinically without further optimization. Perhaps more importantly, Auto‐Planning could be used as a high quality starting point for further plan optimization. This could increase the overall quality of the treatments and reduce the interobserver variation present in manually created treatment plans.

ACKNOWLEDGMENTS

This work is related to AgeCare (Academy of Geriatric Cancer Research), an international research collaboration based at Odense University Hospital, Denmark. Research agreement with Philips Healthcare is acknowledged. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file. Supplementary Material Click here for additional data file.
  30 in total

1.  Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations.

Authors:  Tommy Knöös; Elinore Wieslander; Luca Cozzi; Carsten Brink; Antonella Fogliata; Dirk Albers; Håkan Nyström; Søren Lassen
Journal:  Phys Med Biol       Date:  2006-10-24       Impact factor: 3.609

2.  Sensitivity of NTCP parameter values against a change of dose calculation algorithm.

Authors:  Carsten Brink; Martin Berg; Morten Nielsen
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

3.  A methodology for automatic intensity-modulated radiation treatment planning for lung cancer.

Authors:  Xiaodong Zhang; Xiaoqiang Li; Enzhuo M Quan; Xiaoning Pan; Yupeng Li
Journal:  Phys Med Biol       Date:  2011-06-08       Impact factor: 3.609

4.  Increased organ sparing using shape-based treatment plan optimization for intensity modulated radiation therapy of pancreatic adenocarcinoma.

Authors:  Steven F Petit; Binbin Wu; Michael Kazhdan; André Dekker; Patricio Simari; Rachit Kumar; Russel Taylor; Joseph M Herman; Todd McNutt
Journal:  Radiother Oncol       Date:  2011-06-15       Impact factor: 6.280

5.  Investigation of respiration induced intra- and inter-fractional tumour motion using a standard Cone Beam CT.

Authors:  Karina Lindberg Gottlieb; Christian R Hansen; Olfred Hansen; Jonas Westberg; Carsten Brink
Journal:  Acta Oncol       Date:  2010-10       Impact factor: 4.089

6.  Reduction of observer variation using matched CT-PET for lung cancer delineation: a three-dimensional analysis.

Authors:  Roel J H M Steenbakkers; Joop C Duppen; Isabelle Fitton; Kirsten E I Deurloo; Lambert J Zijp; Emile F I Comans; Apollonia L J Uitterhoeve; Patrick T R Rodrigus; Gijsbert W P Kramer; Johan Bussink; Katrien De Jaeger; José S A Belderbos; Peter J C M Nowak; Marcel van Herk; Coen R N Rasch
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-09-28       Impact factor: 7.038

7.  Influence of dose calculation algorithms on the predicted dose distribution and NTCP values for NSCLC patients.

Authors:  Tine B Nielsen; Elinore Wieslander; Antonella Fogliata; Morten Nielsen; Olfred Hansen; Carsten Brink
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

8.  Effect of HPV-associated p16INK4A expression on response to radiotherapy and survival in squamous cell carcinoma of the head and neck.

Authors:  Pernille Lassen; Jesper G Eriksen; Stephen Hamilton-Dutoit; Trine Tramm; Jan Alsner; Jens Overgaard
Journal:  J Clin Oncol       Date:  2009-03-16       Impact factor: 44.544

9.  Patient geometry-driven information retrieval for IMRT treatment plan quality control.

Authors:  Binbin Wu; Francesco Ricchetti; Giuseppe Sanguineti; Misha Kazhdan; Patricio Simari; Ming Chuang; Russell Taylor; Robert Jacques; Todd McNutt
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

10.  Automated IMRT planning with regional optimization using planning scripts.

Authors:  Ilma Xhaferllari; Eugene Wong; Karl Bzdusek; Michael Lock; Jeff Chen
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

View more
  49 in total

1.  Automated IMRT planning in Pinnacle : A study in head-and-neck cancer.

Authors:  J M A M Kusters; K Bzdusek; P Kumar; P G M van Kollenburg; M C Kunze-Busch; M Wendling; T Dijkema; J H A M Kaanders
Journal:  Strahlenther Onkol       Date:  2017-08-02       Impact factor: 3.621

2.  An empirical method for automatic determination of maximum number of segments in DMPO-based IMRT for Head and Neck cases.

Authors:  Vaitheeswaran Ranganathan; K Joseph Maria Das
Journal:  Rep Pract Oncol Radiother       Date:  2016-09-30

3.  SBRT planning for spinal metastasis: indications from a large multicentric study.

Authors:  Marco Esposito; Laura Masi; Margherita Zani; Raffaela Doro; David Fedele; Cristina Garibaldi; Stefania Clemente; Christian Fiandra; Francesca Romana Giglioli; Carmelo Marino; Laura Orsingher; Serenella Russo; Michele Stasi; Lidia Strigari; Elena Villaggi; Pietro Mancosu
Journal:  Strahlenther Onkol       Date:  2018-10-23       Impact factor: 3.621

4.  Automated prediction of dosimetric eligibility of patients with prostate cancer undergoing intensity-modulated radiation therapy using a convolutional neural network.

Authors:  Tomohiro Kajikawa; Noriyuki Kadoya; Kengo Ito; Yoshiki Takayama; Takahito Chiba; Seiji Tomori; Ken Takeda; Keiichi Jingu
Journal:  Radiol Phys Technol       Date:  2018-08-14

Review 5.  Automation in intensity modulated radiotherapy treatment planning-a review of recent innovations.

Authors:  Mohammad Hussein; Ben J M Heijmen; Dirk Verellen; Andrew Nisbet
Journal:  Br J Radiol       Date:  2018-09-04       Impact factor: 3.039

6.  Artificial Intelligence in Radiation Therapy.

Authors:  Yabo Fu; Hao Zhang; Eric D Morris; Carri K Glide-Hurst; Suraj Pai; Alberto Traverso; Leonard Wee; Ibrahim Hadzic; Per-Ivar Lønne; Chenyang Shen; Tian Liu; Xiaofeng Yang
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-08-24

7.  Evaluation of plan quality improvements in PlanIQ-guided Autoplanning.

Authors:  Bojarajan Perumal; Harikrishna Etti Sundaresan; Vaitheeswaran Ranganathan; Natarajan Ramar; Gipson Joe Anto; Samir Ranjan Meher
Journal:  Rep Pract Oncol Radiother       Date:  2019-09-20

8.  Impact of dosimetric differences between CT and MRI derived target volumes for external beam cervical cancer radiotherapy.

Authors:  Vikneswary Batumalai; Siobhan Burke; Dale Roach; Karen Lim; Glen Dinsdale; Michael Jameson; Cesar Ochoa; Jacqueline Veera; Lois Holloway; Shalini Vinod
Journal:  Br J Radiol       Date:  2020-06-18       Impact factor: 3.039

9.  Comparative Study of Auto Plan and Manual Plan for Nasopharyngeal Carcinoma Intensity-Modulated Radiation Therapy.

Authors:  Xin Xin; Chuandong Cheng; Churong Li; Jie Li; Pei Wang; Gang Yin; Jinyi Lang
Journal:  Cancer Manag Res       Date:  2020-12-02       Impact factor: 3.989

10.  Pareto Optimal Projection Search (POPS): Automated Radiation Therapy Treatment Planning by Direct Search of the Pareto Surface.

Authors:  Charles Huang; Yong Yang; Neil Panjwani; Stephen Boyd; Lei Xing
Journal:  IEEE Trans Biomed Eng       Date:  2021-09-20       Impact factor: 4.756

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.