Universal pulses: A new concept for calibration-free parallel transmission.

Vincent Gras1, Alexandre Vignaud1, Alexis Amadon1, Denis Le Bihan1, Nicolas Boulant1.   

Abstract

PURPOSE: A calibration-free parallel transmission method is investigated to mitigate the radiofrequency (RF) field inhomogeneity problem in brain imaging at 7 Tesla (T). THEORY AND METHODS: Six volunteers were scanned to build a representative database of RF and static field maps at 7T. Small-tip-angle and inversion pulses were designed with joint kT -points trajectory optimization to work robustly on all six subjects. The returned "universal" pulses were then inserted in an MPRAGE sequence implemented on six additional volunteers without further field measurements and pulse optimizations. Similar acquisitions were performed in the circularly polarized mode and with subject-based optimizations for comparison. Performance of the different approaches was evaluated by means of image analysis and computation of the flip angle normalized root mean square errors (NRMSE).
RESULTS: For both the excitation and inversion, the universal pulses (NRMSE∼11%) outperformed the circularly polarized (NRMSE∼28%) and RF shim modes (NRMSE∼20%) across all volunteers and returned slightly worse results than for subject-based optimized pulses (NRMSE∼7%).
CONCLUSION: RF pulses can be designed to robustly mitigate the RF field inhomogeneity problem over a population class. This appears as a first step toward another plug and play parallel transmission solution where the pulse design can be done offline and without measuring subject-specific field maps. Magn Reson Med 77:635-643, 2017.
© 2016 International Society for Magnetic Resonance in Medicine. © 2016 International Society for Magnetic Resonance in Medicine.

Keywords:  RF pulse design; parallel transmission; plug and play; ultra-high field

Mesh:

Year:  2016        PMID: 26888654     DOI: 10.1002/mrm.26148

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  22 in total

1.  Human Connectome Project-style resting-state functional MRI at 7 Tesla using radiofrequency parallel transmission.

Authors:  Xiaoping Wu; Edward J Auerbach; An T Vu; Steen Moeller; Pierre-François Van de Moortele; Essa Yacoub; Kâmil Uğurbil
Journal:  Neuroimage       Date:  2018-09-17       Impact factor: 6.556

2.  Eliminating susceptibility induced hyperintensities in T1w MPRAGE brain images at 7 T.

Authors:  Ruoyun Ma; Thomas R Henry; Pierre-François Van de Moortele
Journal:  Magn Reson Imaging       Date:  2019-08-22       Impact factor: 2.546

Review 3.  Parallel Transmission for Ultrahigh Field MRI.

Authors:  Cem M Deniz
Journal:  Top Magn Reson Imaging       Date:  2019-06

4.  Machine learning RF shimming: Prediction by iteratively projected ridge regression.

Authors:  Julianna D Ianni; Zhipeng Cao; William A Grissom
Journal:  Magn Reson Med       Date:  2018-03-23       Impact factor: 4.668

5.  Toward imaging the body at 10.5 tesla.

Authors:  M Arcan Ertürk; Xiaoping Wu; Yiğitcan Eryaman; Pierre-François Van de Moortele; Edward J Auerbach; Russell L Lagore; Lance DelaBarre; J Thomas Vaughan; Kâmil Uğurbil; Gregor Adriany; Gregory J Metzger
Journal:  Magn Reson Med       Date:  2016-10-21       Impact factor: 4.668

6.  Specific absorption rate implications of within-scan patient head motion for ultra-high field MRI.

Authors:  Emre Kopanoglu; Cem M Deniz; M Arcan Erturk; Richard G Wise
Journal:  Magn Reson Med       Date:  2020-04-17       Impact factor: 4.668

Review 7.  Pulse sequences and parallel imaging for high spatiotemporal resolution MRI at ultra-high field.

Authors:  Benedikt A Poser; Kawin Setsompop
Journal:  Neuroimage       Date:  2017-04-06       Impact factor: 6.556

8.  Bench to bore ramifications of inter-subject head differences on RF shimming and specific absorption rates at 7T.

Authors:  Benjamin M Hardy; Rana Banik; Xinqiang Yan; Adam W Anderson
Journal:  Magn Reson Imaging       Date:  2022-07-13       Impact factor: 3.130

Review 9.  New acquisition techniques and their prospects for the achievable resolution of fMRI.

Authors:  Saskia Bollmann; Markus Barth
Journal:  Prog Neurobiol       Date:  2020-10-23       Impact factor: 11.685

10.  SAR and temperature distributions in a database of realistic human models for 7 T cardiac imaging.

Authors:  Bart R Steensma; Ettore F Meliadò; Peter Luijten; Dennis W J Klomp; Cornelis A T van den Berg; Alexander J E Raaijmakers
Journal:  NMR Biomed       Date:  2021-05-06       Impact factor: 4.044

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.