Literature DB >> 26833303

Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis.

Michael Schiff1, Joe Koo2, Erik Jin2, Eric Schiller2, Ashley Day3, Rebecca Stevens3, Christina Laskar2.   

Abstract

INTRODUCTION: The purpose of the present study was to determine whether the abatacept autoinjector can be used by the intended population without patterns of preventable use errors, and is acceptable when assessed against key user needs.
METHODS: Two independently conducted simulated-use studies, with no active drug administered, quantified use errors and evaluated the abatacept autoinjector and competitor devices on key attributes (comfort, control, ease of use, confidence of dose) and overall acceptability. Autoinjector preference was also assessed. Participants were patients with rheumatoid arthritis, caregivers, and healthcare professionals (HCPs). Participants were informed that a new rheumatoid arthritis autoinjector was being tested but were blinded to the intended drug and sponsor identity.
RESULTS: In the formative (pre-validation) study (n = 54), two high-priority use errors occurred, both of which resulted from protocol non-compliance rather than mental confusion or physical limitations. In the summative (validation) study (n = 99), one high-priority use error occurred; this was deemed a simulated-use study artifact as participant behavior was guided by actual experience associated with the feel of drug delivery into the skin rather than by protocol, so no mitigation steps were considered necessary. Across user groups, average scores were consistently high for the pre-defined key attributes. Overall acceptability scores (7-point scale) were significantly higher for the abatacept versus competitor autoinjectors-formative study: patients 6.7 vs 5.2 (P = 0.0001), caregivers 7.0 vs 4.6 (P = 0.0093), HCPs 6.8 vs 5.1 (P = 0.0020); summative study: patients 6.5 vs 5.9 (P = 0.0404), caregivers 6.8 vs 5.8 (P = 0.0047), HCPs 6.8 vs 5.1 (P = 0.0002). The abatacept autoinjector was preferred to competitor devices: patients 85.7% vs 14.3% (P = 0.00002), caregivers 84.2% vs 15.8% (P = 0.00443), HCPs 95.0% vs 5.0% (P = 0.00004). Positive experiences with the abatacept autoinjector were attributed to the rubberized grip, device size, visualization of dose progression, button ergonomics, and ease of use.
CONCLUSION: The abatacept autoinjector demonstrated usability without patterns of preventable use errors, and with high acceptability ratings across all key attributes assessed. Preference over competitor autoinjectors was due to device ergonomics, visualization of dose progression, confidence of dose delivery, and overall ease of use. FUNDING: Bristol-Myers Squibb.

Entities:  

Keywords:  Abatacept autoinjector; Failure modes and effects analysis; Human factors engineering; Usability; Validation testing

Mesh:

Substances:

Year:  2016        PMID: 26833303      PMCID: PMC4769728          DOI: 10.1007/s12325-016-0286-9

Source DB:  PubMed          Journal:  Adv Ther        ISSN: 0741-238X            Impact factor:   3.845


Introduction

Abatacept, a fully human fusion protein, is the only biologic for the treatment of rheumatoid arthritis (RA) that selectively modulates the CD80/CD86:CD28 co-stimulatory signal required for full T cell activation and is available in both intravenous and subcutaneous (SC) formulations. The intravenous formulation of abatacept has demonstrated efficacy in several patient populations, including methotrexate-naive patients with early RA [1] and patients with an inadequate response to methotrexate [2-5] or to anti-tumor necrosis factor therapy [6, 7]. SC abatacept has been shown to be non-inferior to intravenous abatacept [8, 9]. The SC formulation, available as a pre-filled syringe, was first approved in the US in July 2011 for the treatment of moderate-to-severe RA in adults. Since then, SC abatacept has received marketing approval for the treatment of adult RA in numerous regions, including the EU, Japan, Canada, and Australia. Although the SC delivery method affords users the benefit of self-injection at home, the pre-filled syringe requires several hand manipulations. This can be difficult for individuals with RA, as the disease often affects the small joints of the hand and impairs dexterity. To address this limitation and increase options for patients, a pre-filled, single-use autoinjector for abatacept (ClickJect®; Bristol-Myers Squibb, Princeton, NJ, USA) has been developed, with the aims of increasing the ease of the injection process and minimizing use error risk. The abatacept autoinjector was designed and developed using human factors engineering (HFE). The autoinjector has a large rubberized grip and a hidden needle, and uses an automated injection process including automated delivery of the entire syringe contents. It also incorporates visual feedback to indicate the end of the injection, and shields the needle after injection. These features were designed to facilitate the injection process by improving handling ergonomics and reducing the number of hand manipulations, as well as to help patients who are new to self-injection and to overcome barriers to self-injection such as needle phobia. This report describes the results of two studies that were conducted to assess the usability and acceptability of the abatacept autoinjector. A formative (pre-validation) study was performed to identify aspects of the product design and instructions for use (IFU) that could be further optimized to reduce the risk of user errors, while a summative study was performed as a final validation of the usability of the product and its labeling.

Methods

Two usability studies—formative and summative—were performed to measure the success of the HFE approach for device development (for details of the HFE, see Supplementary File 1). The goals of these studies were to determine: (1) whether the abatacept autoinjector, when provided with IFU, can be used without patterns of preventable use errors that would cause user or patient harm; and (2) if the autoinjector is acceptable for real-world use based on an assessment of key user needs (i.e., comfort, control, ease of use, confidence of dose) and overall acceptability. The summative and formative studies were conducted by an independent company, Ipsos-Insight, LLC, separate from the study sponsor (Bristol-Myers Squibb). Participants were informed that the studies were testing a new autoinjector for an RA therapy, but no other specifics were shared. Throughout the studies, participants were blinded to the drug for which the autoinjector was intended, as well as to any sponsor involvement. Informed consent was obtained from all patients for being included in the study; study risks were included in the informed consent.

Study Population

To replicate actual use, the study population included individuals from three defined user groups who met the following criteria: (1) patients with RA: both injection naive and experienced, with a minimum disease duration of 6 months and a diagnosis of moderate-to-severe RA (as evidenced by current treatment with a biologic or non-biologic disease-modifying anti-rheumatic drug); (2) caregivers: both injection naive and experienced, with a family member who has been diagnosed with RA; and (3) healthcare professionals (HCPs): injection experienced, who routinely work with and train patients with RA, and with a minimum of 2 years of experience. For the summative study, a minimum of 15 respondents per user group were recruited in accordance with the US Food and Drug Administration human factors guidance [10]. In each study, users were divided into two groups: trained and untrained (Fig. 1). To reduce the potential for use error, users should be trained, by an HCP, on the correct technique for preparing and performing injections when a therapy is prescribed, prior to the first injection. However, because there is no guarantee that all users will be adequately trained before their first use, the device was also validated with untrained participants. For the formative study (Fig. 1a), the trained arm comprised only patients; for the summative study (Fig. 1b), the trained arm included both patients and caregivers. Because most HCPs are not expected to receive any training on how to use the autoinjector, this group was not represented in the trained arm of either study. Prior to the studies, representatives from Bristol-Myers Squibb trained a nurse educator on the correct use of the autoinjector. The nurse educator in turn trained patients and caregivers using only the IFU and a device, to ensure the training was representative of that available in a routine clinical setting.
Fig. 1

Trained and untrained user groups by study type. HCPs healthcare professionals

Trained and untrained user groups by study type. HCPs healthcare professionals

Use Environment

As these were simulated-use studies, efforts were made to simulate an end user’s actual use environment, and the study materials were designed to replicate the ergonomics and steps of actual administration. Injections are typically performed in low-traffic, low-distraction, low-noise settings, with normal office environment lighting conditions, and this environment was therefore replicated at the market research facilities where the studies were conducted. Study materials included an injection pad, which was strapped to the desired injection location (abdomen or thigh) in place of the injection site, a refrigerator (unplugged) for device storage, a hand sanitizer to mimic hand washing, and production-grade devices filled with medical-grade silicone oil to simulate the viscosity of abatacept and, hence, the injection time. The devices did not contain active therapy.

Study Procedures and Assessments

During the formative and summative studies, each participant was provided with the IFU and a device, and was asked to perform the necessary steps per the IFU to simulate an injection. The investigators evaluated each step (based on a risk analysis conducted prior to the study; see Supplementary File 1) using an assessment checklist, based on the perceptual, cognitive, and physical requirements of each step. High-priority tasks were defined as those that required correct completion for successful dose delivery. The investigators then rated the observed performance as ‘correct performance’ (A), ‘performed with difficulty’ (D), or ‘use error’ (UE). A ‘performed with difficulty’ (D) classification included all close calls and near misses in which the user experienced confusion or difficulty, misinterpreted the IFU, or made an error that would result in mistreatment or harm, but then recovered and was able to continue so that no actual performance failure occurred. If a performance error occurred, but the user self-identified that the error had been made without prompting, it was also considered a ‘performed with difficulty’ (D) error, since the individual acknowledged the incorrect step and would self-correct with subsequent usage. A ‘use error’ (UE) classification included instances where the participant did not complete the step as appropriate. Following the usability assessment, the investigator and participant discussed the underlying causes of any difficulties (D) or use errors (UE) encountered during the simulated-use assessments, to ascertain if modifications could be incorporated into the product design or IFU to further optimize ease of use. Participants were also asked to rate the acceptability of the autoinjector against a series of key attributes (i.e., comfort, control, ease of use, confidence of dose) and overall acceptability using a 7-point scale (1 = very unacceptable, 4 = neutral, and 7 = very acceptable). If time allowed, those participants who were experienced with an RA autoinjector were provided with their current or most recent autoinjector, allowed to re-familiarize themselves with it, and asked to provide competitive ratings. The participants did not perform a simulated injection with the competitor autoinjector. Participants experienced with an RA autoinjector were also asked for their preference of autoinjector based on their experiences during the simulated-use assessment. A paired t test was used to identify statistical significance (P < 0.05) for the competitive rating analysis, while an exact binomial test was used for the user preference analysis. Finally, respondents were asked to name key positive features of the abatacept autoinjector, based on their simulated-use experience.

Results

The formative (pre-validation) study was conducted from February to March 2014 at market research centers in Fort Lee, NJ, USA; Fort Lauderdale, FL, USA; and San Francisco, CA, USA. A total of 34 patients with RA, 10 caregivers, and 10 HCPs were recruited (Fig. 1a). The summative (validation) study was conducted from August to September 2014 at market research centers in Fort Lee, NJ, USA; Baltimore, MD, USA; Stamford, CT, USA; and Dallas, TX, USA. A total of 51 patients with RA, 33 caregivers, and 15 HCPs were recruited (Fig. 1b). Respondents from Fort Lee, NJ, USA who had participated in the formative study were not eligible to participate in the summative study. The participant demographics are shown in Table 1.
Table 1

Participant demographics of (A) patients with RA, (B) caregivers, and (C) healthcare professionals

Parameter, %a Formative studySummative study
Untrained (n = 24)Trained (n = 10)Untrained (n = 35)Trained (n = 16)
(A) Patients with RA
Injection experienceNaive33302025
Experienced67708075
Injection experience device typeb Autoinjector56436875
Pre-filled syringe44573225
Age, years18–394066
40–4938102325
50–5933603450
60–6921302919
70–754090
SexFemale71807481
Male29202619
Disease duration, yearsc 61185
EducationHigh-school graduate21101119
Some college29604031
College graduate/post-graduate50304850
DexterityNo difficulty210116
Slight/moderate difficulty50605775
Considerable difficulty29403119
Current medicationMTX54504344
Prednisone8006

MTX methotrexate, RA rheumatoid arthritis

aPercentage values are presented, unless otherwise noted

bOf injection-experienced participants: formative untrained, n = 16; formative trained, n = 7; summative untrained, n = 28; summative trained, n = 12

cMedian values are presented

Participant demographics of (A) patients with RA, (B) caregivers, and (C) healthcare professionals MTX methotrexate, RA rheumatoid arthritis aPercentage values are presented, unless otherwise noted bOf injection-experienced participants: formative untrained, n = 16; formative trained, n = 7; summative untrained, n = 28; summative trained, n = 12 cMedian values are presented

Simulated-Use Usability Assessment

Most participants completed each individual step of the simulation task without use errors (Table 2). In the formative study, two use errors were observed with the high-priority injection steps, both related to the respondent [one untrained patient and one HCP (i.e., untrained)] not holding the device in place long enough to complete the injection. Although these errors were observed, post-study actions were not taken to modify the device design or labeling, since the root-cause analysis identified the outcome to be a result of the respondents’ non-compliance to the protocol and study methodology, rather than a result of mental confusion or physical limitations. However, minor layout changes were made to the IFU to improve compliance with two (low-priority risk) use steps: ‘pinch skin’ and ‘check the autoinjector for damage’.
Table 2

Frequency of completion of individual use steps during the formative and summative studies

Task, %Formative studySummative study
PatientsCaregiversHealthcare professionalsPatientsCaregiversHealthcare professionals
Untrained (n = 24)Trained (n = 10)Untrained (n = 10)Untrained (n = 10)Untrained (n = 35)Trained (n = 16)Untrained (n = 17)Trained (n = 16)Untrained (n = 15)
Take autoinjector out of refrigerator; wait 30 min9610090100971008810093
Wash hands92901001009410010010093
Check expiration date96100100100911009410093
Check autoinjector for damage7980100100891009410087
Check the liquid921001001009710094100100
Choose an injection site96100100100100100100100100
Clean the site96100100100971008210093
Remove needle covera 100 100 100 100 100 100 100 100 100
Position the autoinjector100100100100100100100100100
Pinch skin8810080100911001009493
Push down to unlock the autoinjectora 100 100 100 100 100 100 100 100 100
Press buttona 100 100 100 100 100 100 100 100 100
Hold for count and/or wait until blue indicator stops movinga 96 100 100 90 97 100 100 100 100
Dispose used autoinjector into sharps container1001009010094100100100100

Steps classified as ‘A—correct completion’ plus ‘D—performed with difficulty’

aTasks highlighted in bold are high-priority tasks, defined as those that required correct completion for successful dose delivery

Frequency of completion of individual use steps during the formative and summative studies Steps classified as ‘A—correct completion’ plus ‘D—performed with difficulty’ aTasks highlighted in bold are high-priority tasks, defined as those that required correct completion for successful dose delivery In the summative study, all participants except one (an untrained patient) completed all of the high-priority injection steps (Table 2). The exception related to the ‘hold for count and/or wait until blue indicator stops moving’ step. The use error was deemed an artifact of the study, as the patient was currently using a competitor autoinjector and was applying their expectations of the ‘feel’ in the skin of an injection to indicate the appropriate hold time. Since this tactile end-of-dose indication does not apply to a simulated-use setting, no mitigation steps were deemed necessary.

Simulated-Use Acceptability Assessment

Average acceptability scores for the device were consistently high for each user group for all measures (comfort, control, ease of use, confidence of dose, overall acceptability) (Table 3). In the formative study, all 54 participants responded (34 patients, 10 caregivers, and 10 HCPs); mean overall acceptability scores (out of 7) for patients, caregivers, and HCPs were 6.6, 7.0, and 6.8, respectively. In the summative study, 94 out of 99 participants responded (48 patients, 31 caregivers, and 15 HCPs); mean overall acceptability scores for patients, caregivers, and HCPs were 6.6, 6.8, and 6.8, respectively.
Table 3

Participant user experience data

RatingFormative studySummative study
Patients (n = 34)Caregivers (n = 10)Healthcare professionals (n = 10)Patients (n = 48)Caregivers (n = 31)Healthcare professionals (n = 15)
Comfort6.5 (0.8)6.6 (1.0)6.8 (0.6)6.4 (1.0)6.3 (1.0)6.9 (0.4)
Control6.6 (0.9)6.8 (0.4)6.7 (1.0)6.9 (0.4)6.8 (0.5)6.8 (0.4)
Ease of use6.7 (0.8)6.9 (0.3)6.8 (0.4)6.6 (0.8)6.9 (0.3)6.7 (0.6)
Confidence of dose6.8 (0.5)7.0 (0.0)6.4 (1.6)6.6 (0.7)6.9 (0.4)6.7 (0.6)
Overall acceptability6.6 (0.8)7.0 (0.0)6.8 (0.4)6.6 (0.6)6.8 (0.5)6.8 (0.4)

All values are expressed as mean (standard deviation)

Each score was quantified using a 7-point scale, where 1 = very unacceptable, 4 = neutral, and 7 = very acceptable

Participant user experience data All values are expressed as mean (standard deviation) Each score was quantified using a 7-point scale, where 1 = very unacceptable, 4 = neutral, and 7 = very acceptable

Current Autoinjector User Subgroup Analysis

Among patients who had previously administered injections with a competitor autoinjector, overall mean acceptability ratings were significantly greater for the abatacept autoinjector versus competitor autoinjectors across all user groups in both studies [formative study: patients 6.7 vs 5.2 (P = 0.0001), caregivers 7.0 vs 4.6 (P = 0.0093), HCPs 6.8 vs 5.1 (P = 0.0020); summative study: patients 6.5 vs 5.9 (P = 0.0404), caregivers 6.8 vs 5.8 (P = 0.0047), HCPs 6.8 vs 5.1 (P = 0.0002); Table 4]. In the individual categories of comfort, control, ease of use, and confidence of dose, all groups rated the abatacept autoinjector at least the same or superior to competitor autoinjectors; this difference was significant for almost all parameters, across all user groups, and in both studies (Table 4).
Table 4

Comparative user experience data following abatacept autoinjector use in (A) the formative study or (B) the summative study, from participants who had previously administered injections with an alternative autoinjector

RatingPatients (n = 12)Caregivers (n = 5)Healthcare professionals (n = 10)
Mean abatacept autoinjector ratingMean competitor autoinjector ratingMean paired differenceMean abatacept autoinjector ratingMean competitor autoinjector ratingMean paired differenceMean abatacept autoinjector ratingMean competitor autoinjector ratingMean paired difference
(A) Formative study
Comfort6.64.6 P = 0.00337.04.2 P = 0.01356.84.5 P < 0.0001
Control6.84.6 P = 0.00257.04.0 P = 0.01326.74.6 P = 0.0095
Ease of use7.05.9 P = 0.00067.05.2 P = 0.05336.85.2 P = 0.0046
Confidence of dose6.86.3 P = 0.08197.05.8 P = 0.20806.45.6 P = 0.2999
Overall acceptability6.75.2 P = 0.00017.04.6 P = 0.00936.85.1 P = 0.0020

Each score was quantified using a 7-point scale, where 1 = very unacceptable, 4 = neutral, and 7 = very acceptable

Analyses were restricted to those participants who provided both abatacept and competitor autoinjector ratings

Comparative user experience data following abatacept autoinjector use in (A) the formative study or (B) the summative study, from participants who had previously administered injections with an alternative autoinjector Each score was quantified using a 7-point scale, where 1 = very unacceptable, 4 = neutral, and 7 = very acceptable Analyses were restricted to those participants who provided both abatacept and competitor autoinjector ratings Participants demonstrated a strong preference for the abatacept autoinjector over competitor autoinjectors (Table 5). When preference data from the two studies were combined, a significantly greater number of participants in each user group preferred the abatacept autoinjector over their current or most recent autoinjector [patients 85.7% vs 14.3% (P = 0.00002), caregivers 84.2% vs 15.8% (P = 0.00443), HCPs 95.0% vs 5.0% (P = 0.00004)]. For the individual studies, this preference remained significant for patients in the formative (P = 0.00049) and summative (P = 0.01062) studies, and for HCPs in the summative study (P = 0.00049). Although the other comparisons were not significant, all preference data showed at least a trend in favor of the abatacept autoinjector in both studies.
Table 5

Autoinjector preferences from participants who had previously administered injections with an alternative autoinjector

Respondents, n (%)Formative studySummative studyTotal of formative and summative studies
Abatacept autoinjector preferenceCompetitor autoinjector preferenceTest preference of abatacept autoinjectora Abatacept autoinjector preferenceCompetitor autoinjector preferenceTest preference of abatacept autoinjectora Abatacept autoinjector preferenceCompetitor autoinjector preferenceTest preference of abatacept autoinjectora
Patients12 (100)0 (0.0)(73.54, 100.00); P = 0.0004918 (78.3)5 (21.7)(56.30, 92.54); P = 0.0106230 (85.7)5 (14.3)(69.74, 95.19); P = 0.00002
Caregivers5 (100)0 (0.0)(47.82, 100.00); P = 0.0625011 (78.6)3 (21.4)(49.20, 95.34); P = 0.0573716 (84.2)3 (15.8)(60.42, 96.62); P = 0.00443
HCPs7 (87.5)1 (12.5)(47.35, 99.68); P = 0.0703112 (100)0 (0.0)(73.54, 100.00); P = 0.0004919 (95.0)1 (5.0)(75.13, 99.87); P = 0.00004

aDisplayed as (95% confidence interval); P value

Autoinjector preferences from participants who had previously administered injections with an alternative autoinjector aDisplayed as (95% confidence interval); P value

Key Positive Features

In total, 31 formative study participants and 87 summative study participants reported on the positive features of the abatacept autoinjector. The most frequently noted positive features were: rubberized grip (formative study, noted by 58% of respondents; summative study, 56%), device size (formative, 55%; summative, 52%), visualization of dose progression (window size and location, colored plunger rod; formative, 45%; summative, 57%), button ergonomics (shape, reach, activation force; formative, 19%; summative, 37%), and ease of use (number of steps, simplicity; formative, 29%; summative, 30%) (Fig. 2).
Fig. 2

Key positive factors of the abatacept autoinjector reported by respondents

Key positive factors of the abatacept autoinjector reported by respondents

Discussion

The independently conducted studies reported here found that the newly designed autoinjector for the SC delivery of abatacept in patients with RA was easy to use, with low residual risk to users in a real-world setting. Most participants, whether patients, caregivers, or HCPs, trained or untrained, performed all high-priority use steps correctly. Participants rated the abatacept autoinjector highly for user experience in terms of comfort, control, ease of use, confidence of dose, and overall acceptability, with scores nearing the maximum of the 7-point scale used. All user groups across both studies rated the abatacept autoinjector significantly higher than competitor devices for overall acceptability, and when preference data from both studies were combined, significantly more participants in each user group preferred the abatacept autoinjector over their current or most recent device. The drivers for the positive ratings and preference could be grouped into three main categories: device ergonomics (size, rubberized grip, button reach, force to activate), which provides comfort and security during injection; visualization of dose progression (size and location of window, colored plunger rod), which provides confidence of dose delivery; and simplicity of the process, which contributes to the overall ease of use of the device. Because RA can affect the small joints of the hands, many patients with RA suffer from compromised dexterity. In addition to interfering with activities of daily living, poor dexterity can affect the ability to perform the steps required for self-injection in patients who would otherwise be eligible for an SC therapy. Autoinjectors are available for the administration of many SC therapies for chronic conditions, including those in which patients have impaired dexterity [11, 12]. The single-use, pre-filled autoinjector described here was developed following the principles of HFE, to help users to inject one dose of SC abatacept in a home environment. As evidenced by the results of the current studies, the new autoinjector is an important advance for patients receiving, or eligible to receive, SC abatacept, as it increases the ease of the injection process without introducing usability challenges. In both the formative and summative studies, the vast majority of respondents completed each individual step of the simulated injection without use errors. These results demonstrate that, overall, both trained and untrained users were able to operate the autoinjector safely and effectively to deliver the SC injection by following the IFU. Furthermore, the overall preference in favor of the abatacept autoinjector over other available injection devices across a range of attributes indicates that the device’s features were favored by users with experience of such devices. The fact that similar results were obtained across the two studies provides further credence to these findings. These data also indicate that the device features designed in early development (i.e., shape, rubberized grip, placement/size of window, colored plunger rod, button ergonomics) contributed to the favorable comfort, control, confidence of dose, ease of use, and acceptability ratings for the device, indicating value by the end users. Furthermore, as no patterns of error were observed in the studies, use risk was mitigated. A limitation of these evaluations is the simulated-use design of the studies, rather than autoinjector use in a clinical setting. However, according to US Food and Drug Administration guidance, simulation testing, as conducted here, is an acceptable method for assessing the safe and effective use of an autoinjector device. Greater support for the ease of use of the autoinjector may have been obtained by restricting the study population to patients with more severe hand deformity, for whom any improvement would be most beneficial. In addition, respondent numbers were low in the subgroup analyses. A further limitation is the lack of a simulated injection using the competitor autoinjector device; permitting a simulated injection with a competitor device may have allowed for a more accurate comparison with the abatacept autoinjector versus relying on participant memory. The time span between using and rating the competitor device may have influenced the comparative data; therefore, not capturing and adjusting for this time difference may affect the interpretation of these results. Lastly, the autoinjector was branded in order for participants to evaluate its colored design features. As such, HCPs may have ascertained which product the autoinjector was designed to deliver.

Conclusions

The abatacept autoinjector was found to be highly acceptable against key measures of comfort, control, ease of use, confidence of dose, and overall acceptability in two independently conducted simulated-use studies. High overall acceptability ratings were achieved, and these ratings were significantly greater than those for competitor devices. In addition, a significantly greater number of participants in each user group preferred the abatacept autoinjector over their current or most recent autoinjector. Participants’ positive experiences with the abatacept autoinjector can be attributed to the following key features: device ergonomics (size, shape, rubberized grip, button reach, and activation force), which provides comfort and security during injection; visualization of dose progression (size and location of window, colored plunger rod), which provides confidence of dose delivery; and simplicity of the process, which contributes to the overall ease of use of the device. These results show that HFE optimized the device design and IFU of the abatacept autoinjector to ensure its effective use by patients, caregivers, and HCPs without patterns of preventable use errors. Below is the link to the electronic supplementary material. Supplementary material 1 (PDF 313 kb) Supplementary material 2 (DOCX 1338 kb)
  8 in total

1.  Effects of abatacept in patients with methotrexate-resistant active rheumatoid arthritis: a randomized trial.

Authors:  Joel M Kremer; Harry K Genant; Larry W Moreland; Anthony S Russell; Paul Emery; Carlos Abud-Mendoza; Jacek Szechinski; Tracy Li; Zhiyu Ge; Jean-Claude Becker; Rene Westhovens
Journal:  Ann Intern Med       Date:  2006-06-20       Impact factor: 25.391

2.  Abatacept for rheumatoid arthritis refractory to tumor necrosis factor alpha inhibition.

Authors:  Mark C Genovese; Jean-Claude Becker; Michael Schiff; Michael Luggen; Yvonne Sherrer; Joel Kremer; Charles Birbara; Jane Box; Kannan Natarajan; Isaac Nuamah; Tracy Li; Richard Aranda; David T Hagerty; Maxime Dougados
Journal:  N Engl J Med       Date:  2005-09-15       Impact factor: 91.245

3.  Subcutaneous abatacept for the treatment of rheumatoid arthritis: longterm data from the ACQUIRE trial.

Authors:  Mark C Genovese; César Pacheco Tena; Arturo Covarrubias; Gustavo Leon; Eduardo Mysler; Mauro Keiserman; Robert Valente; Peter Nash; J Abraham Simon-Campos; Jane Box; Clarence William Legerton; Evgeny Nasonov; Patrick Durez; Ingrid Delaet; Julie Teng; Rieke Alten
Journal:  J Rheumatol       Date:  2014-03-01       Impact factor: 4.666

4.  Long-term safety, efficacy and inhibition of radiographic progression with abatacept treatment in patients with rheumatoid arthritis and an inadequate response to methotrexate: 3-year results from the AIM trial.

Authors:  Joel M Kremer; Anthony S Russell; Paul Emery; Carlos Abud-Mendoza; Jacek Szechinski; Rene Westhovens; Tracy Li; Xianhuang Zhou; Jean-Claude Becker; Richard Aranda; Charles Peterfy; Harry K Genant
Journal:  Ann Rheum Dis       Date:  2011-10       Impact factor: 19.103

5.  An open-label, multicenter study to evaluate the safe and effective use of the single-use autoinjector with an Avonex® prefilled syringe in multiple sclerosis subjects.

Authors:  J Theodore Phillips; Edward Fox; William Grainger; Dianne Tuccillo; Shifang Liu; Aaron Deykin
Journal:  BMC Neurol       Date:  2011-10-14       Impact factor: 2.474

6.  The 6-month safety and efficacy of abatacept in patients with rheumatoid arthritis who underwent a washout after anti-tumour necrosis factor therapy or were directly switched to abatacept: the ARRIVE trial.

Authors:  M Schiff; C Pritchard; J E Huffstutter; V Rodriguez-Valverde; P Durez; X Zhou; T Li; K Bahrt; S Kelly; M Le Bars; M C Genovese
Journal:  Ann Rheum Dis       Date:  2008-12-15       Impact factor: 19.103

7.  Clinical efficacy and safety of abatacept in methotrexate-naive patients with early rheumatoid arthritis and poor prognostic factors.

Authors:  R Westhovens; M Robles; A C Ximenes; S Nayiager; J Wollenhaupt; P Durez; J Gomez-Reino; W Grassi; B Haraoui; W Shergy; S-H Park; H Genant; C Peterfy; J-C Becker; A Covucci; R Helfrick; J Bathon
Journal:  Ann Rheum Dis       Date:  2009-01-05       Impact factor: 19.103

8.  Results of a human factors experiment of the usability and patient acceptance of a new autoinjector in patients with rheumatoid arthritis.

Authors:  Florence Schwarzenbach; Michèle Dao Trong; Laurent Grange; Philippe E Laurent; Herve Abry; Joël Cotten; Corinne Granger
Journal:  Patient Prefer Adherence       Date:  2014-02-11       Impact factor: 2.711

  8 in total
  8 in total

1.  Usability and Patient Preference Phase 3 Study of the Sarilumab Pen in Patients with Active Moderate-to-Severe Rheumatoid Arthritis.

Authors:  Alan Kivitz; Lydie Baret-Cormel; Hubert van Hoogstraten; Sheldon Wang; Janie Parrino; Christine Xu; Marina Stanislav
Journal:  Rheumatol Ther       Date:  2017-12-05

2.  Perceptions and Preferences of Two Etanercept Autoinjectors for Rheumatoid Arthritis: A New European Union-Approved Etanercept Biosimilar (Benepali®) Versus Etanercept (Enbrel®) - Findings from a Nurse Survey in Europe.

Authors:  Kunal Thakur; Anna Biberger; Alexandra Handrich; Mourad Farouk Rezk
Journal:  Rheumatol Ther       Date:  2016-05-23

Review 3.  Abatacept: A Review in Rheumatoid Arthritis.

Authors:  Hannah A Blair; Emma D Deeks
Journal:  Drugs       Date:  2017-07       Impact factor: 9.546

4.  Patient and Healthcare Professionals Preference for Brenzys vs. Enbrel Autoinjector for Rheumatoid Arthritis: A Randomized Crossover Simulated-Use Study.

Authors:  Marc Egeth; Jennifer Soosaar; Peter Nash; Denis Choquette; Ricardo Infante; Dena Rosen Ramey; Sevag Sahakian; Angela Lai; Jin Ju Kim; David Wu
Journal:  Adv Ther       Date:  2017-04-17       Impact factor: 3.845

5.  Usability of a novel disposable autoinjector device for ixekizumab: results from a qualitative study and an open-label clinical trial, including patient-reported experience.

Authors:  Kristina Callis Duffin; Michael Bukhalo; Margaret A Bobonich; David Shrom; Fangyi Zhao; James R Kershner; Anne Gill; Beth Pangallo; Catherine L Shuler; Jerry Bagel
Journal:  Med Devices (Auckl)       Date:  2016-10-12

6.  User-Centric Approach to Specifying Technical Attributes of Drug Delivery Devices: Empirical Study of Autoinjector-Cap Removal Forces.

Authors:  Andreas Schneider; Philipp Richard; Philippe Mueller; Christoph Jordi; Mary Yovanoff; Jakob Lange
Journal:  Patient Prefer Adherence       Date:  2021-02-02       Impact factor: 2.711

7.  Advances in subcutaneous injections: PRECISE II: a study of safety and subject preference for an innovative needle-free injection system.

Authors:  E Lynne Kelley; Richard H Smith; Gillian Corcoran; Sandra Nygren; Mary V Jacoski; Andrea Fernandes
Journal:  Drug Deliv       Date:  2021-12       Impact factor: 6.419

8.  Formative Study on the Wearability and Usability of a Large-Volume Patch Injector.

Authors:  Jakob Lange; Andreas Schneider; Christoph Jordi; Michael Lau; Timothy Disher
Journal:  Med Devices (Auckl)       Date:  2021-11-16
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.