A new, linear accelerator based radiosurgical system has been designed and tested. It incorporates a mechanical system of precision bearings to control all patient and accelerator movements. The dosimetry system allows near real-time examination of the isodose distributions in any computed tomography plane such that any treatment plan can be quickly optimized. Extensive testing shows a radiation beam accuracy of .2 +/- .1 mm. The dose gradient between the 90% and 10% isodose distributions compares favorably with any previously published radiosurgical method.Entities:
Mesh:
Year: 1989 PMID: 2683164 DOI: 10.1016/0090-3019(89)90135-3