PURPOSE: The purpose of this study is to evaluate our first experience with 3D image-guided breast brachytherapy and to compare dose distribution parameters between Paris dosimetry system (PDS) and image-based plans.Entities:
Keywords: Paris dosimetry; boost; breast cancer; image-based brachytherapy
Year: 2015 PMID: 26816505 PMCID: PMC4716128 DOI: 10.5114/jcb.2015.56492
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Printout of pre-planning procedure with ribs counted and numbered and distances marked in all three dimensions from surgery clips to distinctive elements of patient's body (sternum and nipple)
Mean values of dosimetric parameters and doses at organs at risk (OARs) of both plans
| Dosimetric parameter | PDS plan | OPT plan |
|
|---|---|---|---|
| VPTV90 [%] | 88.6 ± 7.0 | 98.1 ± 2.4 | < 0.05 |
| VPTV100 [%] | 79.9 ± 7.2 | 93.9 ± 3.6 | < 0.05 |
| D90 [%] | 88.2 ± 9.3 | 105.9 ± 9.5 | < 0.05 |
| D100 [%] | 59.8 ± 9.4 | 75.7 ± 5.4 | < 0.05 |
| Skin D0.1cc [%] | 67.5 ± 11.6 | 73.9 ± 10.0 | < 0.05 |
| Skin D2cc [%] | 52.8 ± 7.4 | 59.2 ± 7.7 | < 0.05 |
| Rib D0.1cc [%] | 24.5 ± 17.6 | 28.1 ± 17.0 | < 0.05 |
| Lung D2cc [%] | 18.1 ± 11.2 | 21.0 ± 11.6 | < 0.05 |
| DNR | 0.29 ± 0.05 | 0.42 ± 0.07 | < 0.05 |
| DHI inside PTV | 0.71 ± 0.05 | 0.54 ± 0.09 | < 0.05 |
| COIN | 0.69 ± 0.10 | 0.76 ± 0.06 | < 0.05 |
| TRAK [Gym–1] | 0.0038 ± 0.0014 | 0.0044 ± 0.0017 | < 0.05 |
VPTV90 [%], VPTV100 [%] – relative percentage of planning target volume, which receives 90% and 100% of prescribed dose, respectively; D90 [%], D100 [%] – dose to the 90% and 100% volume of interest relative to the prescribed dose, respectively; D0.1cc, D2cc – dose at the most exposed 0.1cc and 2cc of volume of interest, respectively, DNR – dose nonuniformity ratio, DHI inside PTV – dose homogeneity index inside planning target volume, COIN – conformity index, TRAK [Gym–1] – total reference air kerma, PDS – Paris dosimetry system, OPT – image-based plan using graphical optimization
Mean values of dosimetric parameters divided into two groups of consecutive patients
| Dosimetric parameter | PDS plans | OPT plans | ||
|---|---|---|---|---|
| First 20 patients | Next 29 patients | First 20 patients | Next 29 patients | |
| VPTV100 [%] | 79.5 | 80.3 | 93.5 | 94.2 |
| D90 [%] | 88.8 | 87.8 | 105.0 | 106.4 |
| D100 [%] | 62.0 | 58.3 | 76.7 | 76.0 |
| DNR | 0.33 | 0.26 | 0.44 | 0.4 |
| DHI inside PTV | 0.68 | 0.74 | 0.54 | 0.55 |
| COIN | 0.69 | 0.67 | 0.75 | 0.76 |
VPTV100 [%] – relative percentage of planning target volume, which receives 100% of prescribed dose, D90 [%], D100 [%] – dose to the 90% and 100% volume of interest relative to the prescribed dose, respectively, DNR – dose nonuniformity ratio, DHI inside PTV – dose homogeneity index inside planning target volume, COIN – conformity index, PDS – Paris dosimetry system, OPT – image-based plan using graphical optimization
Fig. 2Correlation between total reference air kerma (TRAK) values for Paris dosimetry system (PDS) (orange) and graphical optimization (OPT) (blue) plans, and volume of planning target volume (PTV)
Fig. 3Correlation between dose homogeneity index for graphical optimization (OPT) plans and value of prescribed isodose (F-factor) chosen for dose specification in these plans