Effect of Variation in Test Methods on Performance of Ultraviolet-C Radiation Room Decontamination.

Jennifer L Cadnum1, Myreen E Tomas1, Thriveen Sankar1, Annette Jencson1, J Itty Mathew2, Sirisha Kundrapu2, Curtis J Donskey2.   

Abstract

OBJECTIVE: To determine the effect of variation in test methods on performance of an ultraviolet-C (UV-C) room decontamination device.
DESIGN: Laboratory evaluation.
METHODS: We compared the efficacy of 2 UV-C room decontamination devices with low pressure mercury gas bulbs. For 1 of the devices, we evaluated the effect of variation in spreading of the inoculum, carrier orientation relative to the device, type of organic load, type of carrier, height of carrier, and uninterrupted versus interrupted exposures on measured UV-C killing of methicillin-resistant Staphylococcus aureus and Clostridium difficile spores.
RESULTS: The 2 UV-C room decontamination devices achieved similar log10 colony-forming unit reductions in the pathogens with exposure times ranging from 5 to 40 minutes. On steel carriers, spreading of the inoculum over a larger surface area significantly enhanced killing of both pathogens, such that a 10-minute exposure on a 22-mm2 disk resulted in greater than 2 log reduction in C. difficile spores. Orientation of carriers in parallel rather than perpendicular with the UV-C lamps significantly enhanced killing of both pathogens. Different types of organic load also significantly affected measured organism reductions, whereas type of carrier, variation in carrier height, and interrupted exposure cycles did not.
CONCLUSIONS: Variation in test methods can significantly impact measured reductions in pathogens by UV-C devices during experimental testing. Our findings highlight the need for standardized laboratory methods for testing the efficacy of UV-C devices and for evaluations of the efficacy of short UV-C exposure times in real-world settings.

Entities:  

Mesh:

Year:  2016        PMID: 26809607     DOI: 10.1017/ice.2015.349

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  5 in total

Review 1.  Hospital Infection Control: Clostridioides difficile.

Authors:  Nicholas A Turner; Deverick J Anderson
Journal:  Clin Colon Rectal Surg       Date:  2020-02-25

2.  Killing of Candida auris by UV-C: Importance of exposure time and distance.

Authors:  Theun de Groot; Anuradha Chowdhary; Jacques F Meis; Andreas Voss
Journal:  Mycoses       Date:  2019-03-12       Impact factor: 4.377

Review 3.  Systematic review on use, cost and clinical efficacy of automated decontamination devices.

Authors:  Stephanie J Dancer; Marco-Felipe King
Journal:  Antimicrob Resist Infect Control       Date:  2021-02-12       Impact factor: 4.887

4.  Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards.

Authors:  Vincent Masse; Michael J Hartley; Michael B Edmond; Daniel J Diekema
Journal:  Antimicrob Resist Infect Control       Date:  2018-02-22       Impact factor: 4.887

Review 5.  Effectiveness of ultraviolet devices and hydrogen peroxide systems for terminal room decontamination: Focus on clinical trials.

Authors:  David J Weber; William A Rutala; Deverick J Anderson; Luke F Chen; Emily E Sickbert-Bennett; John M Boyce
Journal:  Am J Infect Control       Date:  2016-05-02       Impact factor: 2.918

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.