Long-Term Durability of Carpentier-Edwards Magna Ease Valve: A One Billion Cycle In Vitro Study.

Vrishank Raghav1, Ikechukwu Okafor2, Michael Quach3, Lynn Dang3, Salvador Marquez3, Ajit P Yoganathan4.   

Abstract

BACKGROUND: Durability and hemodynamic performance are top considerations in selecting a valve for valve replacement surgery. This study was conducted in order to evaluate the long-term mechanical durability and hydrodynamic performance of the Carpentier-Edwards PERIMOUNT Magna Ease Bioprostheses, through 1 billion cycles (equivalent to 25 years).
METHODS: In vitro valve hydrodynamic performance, durability, and quantitative flow visualization were conducted in accordance with ISO 5840:2005 heart valve standard. The study valves were subjected to accelerated valve cycling to an equivalent of 25 years of wear. Hydrodynamic evaluations at intervals of 100 million cycles (2.5 years) were performed on the study valves. New uncycled Magna Ease valves were used as hydrodynamic controls in this study. A quantitative assessment of the fluid motion downstream of the control and study valves was performed using particle image velocimetry. The results between the test and control valves were compared to assess valve performance after an equivalent of 25 years of wear.
RESULTS: All study valves met the ISO 5840 requirements for effective orifice area, 1.81 ± 0.06 cm(2) and 2.06 ± 0.17 cm(2), and regurgitant fraction, 1.11% ± 0.87% and 2.5% ± 2.34%, for the 21 mm and 23 mm study valves, respectively. The flow characterization of the control valves and the billion-cycle valves demonstrated that the valves exhibited similar flow characteristics. The velocity and shear stress fields were similar between the control and study valves.
CONCLUSIONS: The Magna Ease valves demonstrated excellent durability and hydrodynamic performance after an equivalent of 25 years of simulated in vitro wear. All study valves successfully endured 1 billion cycles of simulated wear, 5 times longer than the standard requirement for a tissue valve as stipulated in ISO 5840.
Copyright © 2016 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2016        PMID: 26806168     DOI: 10.1016/j.athoracsur.2015.10.069

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  11 in total

1.  Beyond Bernoulli: Improving the Accuracy and Precision of Noninvasive Estimation of Peak Pressure Drops.

Authors:  Fabrizio Donati; Saul Myerson; Malenka M Bissell; Nicolas P Smith; Stefan Neubauer; Mark J Monaghan; David A Nordsletten; Pablo Lamata
Journal:  Circ Cardiovasc Imaging       Date:  2017-01       Impact factor: 7.792

2.  Integrated strategy for in vitro characterization of a bileaflet mechanical aortic valve.

Authors:  Francesca Maria Susin; Stefania Espa; Riccardo Toninato; Stefania Fortini; Giorgio Querzoli
Journal:  Biomed Eng Online       Date:  2017-02-16       Impact factor: 2.819

3.  The effect of decellularisation on the real time mechanical fatigue of porcine aortic heart valve roots.

Authors:  Amisha Desai; Eileen Ingham; Helen E Berry; John Fisher; Louise M Jennings
Journal:  PLoS One       Date:  2022-04-01       Impact factor: 3.240

Review 4.  Current development of bovine jugular vein conduit for right ventricular outflow tract reconstruction.

Authors:  Chenggang Li; Bo Xie; Ruizhe Tan; Lijin Liang; Zhaoxiang Peng; Qi Chen
Journal:  Front Bioeng Biotechnol       Date:  2022-08-04

5.  A Bicentric Propensity Matched Analysis of 158 Patients Comparing Porcine Versus Bovine Stented Bioprosthetic Valves in Pulmonary Position.

Authors:  Bunty Ramchandani; Raúl Sánchez; Juvenal Rey; Luz Polo; Álvaro Gonzalez; Maria-Jesús Lamas; Tomasa Centella; Jesús Díez; Ángel Aroca
Journal:  Korean Circ J       Date:  2022-04-21       Impact factor: 3.101

6.  Long-term durability of a new surgical aortic valve: A 1 billion cycle in vitro study.

Authors:  Vahid Sadri; Phillip M Trusty; Immanuel David Madukauwa-David; Ajit P Yoganathan
Journal:  JTCVS Open       Date:  2021-11-04

7.  The Fluid Dynamical Performance of the Carpentier-Edwards PERIMOUNT Magna Ease Prosthesis.

Authors:  Philipp Marx; Wojciech Kowalczyk; Aydin Demircioglu; Gary Neil Brault; Hermann Wendt; Sharaf-Eldin Shehada; Konstantinos Tsagakis; Mohamed El Gabry; Heinz Jakob; Daniel Wendt
Journal:  Biomed Res Int       Date:  2018-01-10       Impact factor: 3.411

8.  Mid-term follow-up after aortic valve replacement with the Carpentier Edwards Magna Ease prosthesis.

Authors:  Taufiek K Rajab; Jason M Ali; Jules Hernández-Sánchez; Jennifer Mackie; Vincenzo Grimaudo; Silvia Sinichino; Christine Mills; Bushra Rana; John Dunning; Yasir Abu-Omar
Journal:  J Cardiothorac Surg       Date:  2020-08-03       Impact factor: 1.637

9.  Degeneration of Bioprosthetic Heart Valves: Update 2020.

Authors:  Alexander E Kostyunin; Arseniy E Yuzhalin; Maria A Rezvova; Evgeniy A Ovcharenko; Tatiana V Glushkova; Anton G Kutikhin
Journal:  J Am Heart Assoc       Date:  2020-09-21       Impact factor: 5.501

Review 10.  Use of Sutureless and Rapid Deployment Prostheses in Challenging Reoperations.

Authors:  Igor Vendramin; Andrea Lechiancole; Daniela Piani; Gaetano Nucifora; Giovanni Benedetti; Sandro Sponga; Daniele Muser; Uberto Bortolotti; Ugolino Livi
Journal:  J Cardiovasc Dev Dis       Date:  2021-06-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.