Device closure of secundum atrial septal defect's and the risk of cardiac erosion.

J D R Thomson1, S A Qureshi2.   

Abstract

Cardiac erosion related to transcatheter atrial septal defect closure devices is of increasing concern. Erosion is reported to have occurred with most of currently available occluder devices. Perhaps due to the very large number of implants worldwide, the Amplatzer (St Jude) occluder is associated with the majority of cardiac erosion events reported in the literature. Best current estimates of the incidence of erosion with the St Jude device are between one and three cases per 1000 implants. Most events occur early after implantation and it is rare, although not unheard of, for events to occur after a year following device insertion. It is important that those involved with closure programmes are vigilant for the problem, because device-related erosion is associated with a significant mortality risk. Despite considerable debate, the risk factors (either patient or device) for erosion remain unclear and require further investigation. Currently available data sets have focussed largely on erosion cohorts and are unable to place these cases in appropriate context with non-erosion closure cases. What is certain is that programmes implanting these devices must take care to implant appropriately sized devices and have in place plans to ensure that patients are both well informed and can access help and advice in the event of developing symptoms.

Entities:  

Keywords:  atrium; congenital heart defects; congenital heart disease; device closure

Year:  2015        PMID: 26796060      PMCID: PMC4676437          DOI: 10.1530/ERP-15-0023

Source DB:  PubMed          Journal:  Echo Res Pract        ISSN: 2055-0464


Background

The search for less invasive alternatives to surgery for closing secundum atrial septal defects (ASD) has occupied cardiologists for decades. Trans-catheter closure of secundum ASD was first described by King and Mills in 1976 (1). In the mid 1990s, the introduction of the Amplatzer septal occluder (ASO) (http://health.sjm.com/amplatzer-septal-occluder), a nitinol device with a central waist manufactured by the AGA Corporation (Fig. 1) brought this procedure into mainstream cardiological practice. The ASO device was reliable and consistent in its handling characteristics and could easily be recaptured and redeployed without destroying its structure if it was not in an acceptable position initially. Importantly, it was able to deal with the majority of secundum ASDs regardless of their size. It can be said without exaggeration that the introduction of the ASO was a game-changing event for congenital cardiac intervention. More than 200 000 patients worldwide (an estimate based on the number of shipped devices) have subsequently been treated with the device. In addition, partly through experience derived from ASD occlusion, cardiologists developed skills that led to a period of intense development in other aspects of this field.
Figure 1

Amplatzer septal occluder (ASO): Lateral aspect. Left and right discs (blue arrows), central core (red arrow).

Amplatzer septal occluder (ASO): Lateral aspect. Left and right discs (blue arrows), central core (red arrow). Over time, a number of other devices have become commercially available for ASD closure and deserve consideration with reference to cardiac erosion. These currently include: Occlutech Figulla occluder (http://www.occlutech.com). Cera (Lifetech) occluder (http://www.lifetechmed.com). PFM Nit-Occlud device (http://www.pfmmedical.com/en). Gore Cardioform septal occluder (formerly Gore Septal Occluder (GSO)) (http://www.goremedical.com/eu/cardioform/). Cardia Ultrasept (formerly Atriasept 2) (http://www.cardiainc.com). Cocoon septal occluder (http://sanaremedprod.com/cocoon-septal-occluder/).

Erosion

Cardiac erosion, as a result of mechanical trauma following device implantation, was initially recognised in case reports or small series and in direct communication from physicians to the AGA Corporation (subsequently purchased by St Jude Medical) (2, 3, 4). In most cases, clinical presentation after erosion is with sudden onset of symptoms, e.g., chest pain in association with haemodynamic instability, typically associated with a new pericardial effusion. In some patients, a fistula may be evident on echocardiography (Fig. 2).
Figure 2

Trans-thoracic echocardiogram (TTE): Short axis view. Amplatzer septal occluder positioned in the atrial septum. Fistula (due to device related erosion) between the aorta and right atrium.

Trans-thoracic echocardiogram (TTE): Short axis view. Amplatzer septal occluder positioned in the atrial septum. Fistula (due to device related erosion) between the aorta and right atrium. Over a 6-year period (1998–2004), 28 cases of haemodynamic instability, following device insertion, were reported to AGA (5). Data relating to these cases were considered by an AGA-appointed expert panel (5). The majority of patients (19/28, 67%) developed symptoms within 72 h of device implantation, 21/28 (75%) underwent surgery, and the subsequent perforation site was either the left or right atrial roof, the aorta or a combination of those sites. From these data, the AGA Corporation determined the erosion rate at that time in the US to be ∼0.1%. This can at best be considered speculative, given that neither the numerator nor the denominator was accurately known. Further insights were gained through analysis of the FDA (voluntary) adverse event website, the ‘Manufacturer and User Facility Device Experience’ (MAUDE), and this created additional concern. Cardiac erosion was responsible for a large proportion of events reported to MAUDE (15%). In the 2009 report from DiBardio et al. (6), based on these data, there were ten deaths associated with the 51 cases of cardiac perforation. Although most of the erosions occurred relatively early after device implantation, events were also reported in previously well patients up to 3 years after device closure of the ASD (4). Subsequent case reports have demonstrated that erosions can occur even later after device implantation (7, 8). Analysis of an updated contemporary MAUDE dataset identified over 100 potential erosions leading to a readjustment of the likely incidence of erosion to 1–3 cases per 1000 implants, based on approximate worldwide sales of 250 000 ASO devices (9, 10) As a result, the FDA issued guidelines to both physicians and patients: http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm371145.htm. In addition to the appropriate advice relating to explicit and thorough counselling of patients prior to implantation, the intervention of the FDA led to a rapid change in the instructions for use (IFU) of the ASO device by St Jude Medical in the US in 2013. Although the altered IFU applied primarily to the US, they were taken seriously in other countries and led to written advice from the company to implanters in other jurisdictions. Despite the lack of definitive evidence relating to the risk factors, the new IFU seemingly removed the option of an interventional approach for many patients and resulted in widespread confusion amongs implanting physicians worldwide, before being rapidly amended.

What are the risk factors for erosion?

A number of potential risk factors have been postulated for the occurrence of ASO-related cardiac erosion. These include deficient aortic and/or superior rim (the rim to the dome of the left atrium), movement of the device within the heart, the insertion of an unnecessarily large or oversized occluder, the insertion of an undersized occluder, patient age and device type.

Deficient rims

A deficient anterior and/or superior rim may bring an edge of a device into contact with either the aorta or the ‘roof’ of the atrium (Fig. 3, Fig. 4) This may then cause tissue trauma and subsequently an erosion. As a result, there is some logic in the argument that a device positioned within a superiorly placed defect may be at a higher risk for eventual erosion. Certainly deficient anterior or superior rims predominate in studies describing the anatomy after cardiac erosion and in AGA's review of erosion cases, deficiency in this area of the heart was present in almost 90% (5, 11). However it is crucially important to remember that deficient rims of this nature are common in ASDs in general, particularly larger ASDs where the prevalence may be as high as 70–80% (11, 12). Data currently in existence (essentially retrospective case series) are insufficient to draw definite conclusions in this regard, and the issue will only be resolved by large, carefully constructed and lengthy longitudinal studies.
Figure 3

(A) Trans-oesophageal echocardiogram (TOE): Mid-oesophagus, 25° view demonstrating a secundum atrial septal defect (ASD) (crosses). P, posterior margin; AS, antero-superior margin; Ao, Aorta; LA, Left atrium; RA, Right atrium. (B) Trans-oesophageal echocardiogram (TOE): Mid-oesophagus, 90° view demonstrating a secundum ASD. PI, Posterio-inferior margin; PS, Posterio-superior margin; SVC, superior vena cava.

Figure 4

Trans-oesophageal echocardiogram (TOE) with colour Doppler: Mid-oesophagus, 45° view demonstrating an ASD (crosses) with a deficient margin to the aorta. Note the complete lack of tissue to the aorta (Ao), LA, Left atrium; RA, Right atrium.

(A) Trans-oesophageal echocardiogram (TOE): Mid-oesophagus, 25° view demonstrating a secundum atrial septal defect (ASD) (crosses). P, posterior margin; AS, antero-superior margin; Ao, Aorta; LA, Left atrium; RA, Right atrium. (B) Trans-oesophageal echocardiogram (TOE): Mid-oesophagus, 90° view demonstrating a secundum ASD. PI, Posterio-inferior margin; PS, Posterio-superior margin; SVC, superior vena cava. Trans-oesophageal echocardiogram (TOE) with colour Doppler: Mid-oesophagus, 45° view demonstrating an ASD (crosses) with a deficient margin to the aorta. Note the complete lack of tissue to the aorta (Ao), LA, Left atrium; RA, Right atrium.

Device movement in the heart

In theory, a relatively mobile device may exert a ‘cheese wire’ effect within the myocardium. Studies demonstrate that there are different patterns of contact between the device and the myocardium and also that following closure of ASD, the relationship of a device with adjacent tissue changes as the device settles and the heart remodels (13, 14). In addition, particularly ‘dynamic’ ASDs (those that apparently change significantly in size with the cardiac cycle) were a feature of erosion cases in a recently published review (13). There are also anecdotal cases where erosion may have been associated with exaggerated cardiac motion related to exercise (15).

Device size

As with deficient tissue rims, the use of apparently oversized devices within defects predominate within case series of cardiac erosion (5, 11). There are potential issues relating to the retrospective assessment of the appropriateness of device size. However, it is easy to envisage how a device that is significantly too large for a particular defect might protrude into a vulnerable tissue plane, eventually causing damage. To add to the apparent confusion regarding aetiology and risk factors, in direct contrast a published survey of experienced implanters within the CCISC consortium (a US-based quality-assurance data submission group), El-Said and Moore found that in the opinion of over two-thirds of respondents, undersized devices which do not straddle the aorta were at the highest risk of erosion! (16). As a general principle, the use of an adequate- (but no more, and probably no less) sized device should be the aim during implantation; however, sizing ASDs is an imprecise process in which different operators have significantly divergent views. Those closing ASDs including echocardiographers guiding the procedure should be aware of the potential for significant oversizing if balloon interrogation of the ASD is used. As a minimum, if a balloon is deemed necessary, the principles of ‘stop flow,’ i.e., the obliteration of colour flow rather than producing a waist on the balloon due to tissue pressure, should be strictly adhered to.

Patient age

Although erosion is reported in paediatric patients, there appears to be a greater relative risk in adults (5, 10, 11). To our knowledge, there are currently no reports of deaths related to erosion with the ASO in paediatric patients. The reasons for this are not entirely clear, but tissue rigidity would seem to be a potential answer.

Occluder type

Erosions have been reported with the ASO, Occlutech occluder (17), Cardia device (18) as well as with devices that are no longer available on the commercial market (19). Given the relative disparity in the numbers of implants with each device, it is difficult to be absolutely certain that the risk of erosion with ASO is genuinely greater than with other devices. It is important to be aware that as the ASO increases in size, there are changes in wire thickness, leading to ‘jumps’ in stiffness at certain points in the device range, most notably beyond 24 mm. It is noteworthy that the 26 mm ASO device appears to be disproportionally associated with cardiac perforation, adding weight to the argument that device stiffness is an important issue. There have been more than 20 000 implants worldwide with the Gore family of occluders (the HELEX device and Gore Septal Occluder); these devices are relatively soft and compliant and have not been associated with any reported erosions.

Implantation technique

As yet, there is little or no information relating to the conduct of the occlusion procedure and the subsequent risk of early erosion. In all of the previous analyses, no information has been collected about the number of attempts made for implanting a device. It is possible that the risk of erosion may be exacerbated by multiple attempts at deployment of the device prior to the final delivery, leading to local trauma.

What does the echocardiographer need to know?

Experts in echocardiography are an essential part of the team in all areas of ASD closure from pre-procedural assessment, to guiding the procedure, as well as during follow-up after device implantation. It is clear that there are limited data upon which to delineate risk factors for cardiac erosion following transcatheter ASD occlusion. What information there is available seems to suggest the following: That the problem is primarily associated with the ASO. That there may be an association with deficient anterior or superior rims. That there is additional risk related to increasing device stiffness. That incorrect sizing (either over-sizing or under-sizing) of the device relative to the defect may be an issue. That the problem is more common and perhaps more serious in adult than in paediatric patients. These variables can be factored into patient selection for procedures, counselling of patients prior to a procedure and the choice of device used. However, a coherent argument can also be made that at present we do not have enough information to identify the risk factors and therefore material changes to implantation strategy based on uncertain stratification is perhaps an illogical reaction. For echocardiographers involved with guiding ASD closure procedures, a thorough and accurate delineation of the anatomy is essential. Agreed protocols should be in place with regard to the assessment, ensuring appropriate accurate defect sizing. The echocardiographer should also establish that the device is not unduly affecting the natural anatomy of the heart prior to release, e.g., exerting repetitive compressive force on the aorta. If this occurs, a discussion amongst the implanting team should take place regarding the advisability of device release. Finally, it is important that protocols are in place with regard to follow-up and surveillance of implanted devices. Vigilance for the presence of pericardial effusion, alterations in device position and any other abnormality, particularly when associated with patient symptoms, are essential. Current recommendations are that patients with devices should remain under indefinite review, even though erosions are by definition unexpected events that cannot necessarily be detected by ‘untargeted’ follow-up. However, continued contact with the implanting team means that patients are more likely to seek appropriate help in the event of worrying symptoms (pain, dizziness, collapse, etc.).

Implications for implantable device technology

Anything other than a cursory analysis of the current available data leads to an uncomfortable conclusion. Despite the fact that there are over 200 000 patients with implants going back almost two decades, previous data collection relating to the implantation and follow-up of ASD devices have been insufficiently robust to identify the risk factors for rare complications of this nature. As a result, discussions about changes to implant practice are based on current ‘received wisdom’ and not hard scientific fact. This is at odds with the issuing of a change in the instructions for use of the ASO device by St Jude Medical in many countries and markets during 2013, which was not based on solid evidence and was both confusing and threatening for physicians. Clearly, it is essential that future studies be conducted to examine the issue of erosion further. There are also important lessons for the wider industry involved with the implantation of devices (including physicians, the medical device industry and those purchasing these services on behalf of patients) in terms of mandatory data collection. Data collection of this nature will come at a cost but is in the best long-term interests of patients.
  17 in total

1.  [Perforation of the aorta after percutaneous closure of an atrial septal defect with an Amplatz prosthesis, presenting with acute severe hemolysis].

Authors:  Y Aggoun; B Gallet; P Acar; M Pulik; D Czitrom; A Lagier; F Laborde
Journal:  Arch Mal Coeur Vaiss       Date:  2002-05

2.  Development of aorta-to-right atrial fistula following closure of secundum atrial septal defect using the Amplatzer septal occluder.

Authors:  David S Chun; Mark W Turrentine; Ahmad Moustapha; Mark H Hoyer
Journal:  Catheter Cardiovasc Interv       Date:  2003-02       Impact factor: 2.692

3.  Life-threatening isometric-exertion related cardiac perforation 5 years after Amplatzer atrial septal defect closure: should isometric activity be limited in septal occluder holders?

Authors:  Francesco Santini; Mohammed Morjan; Francesco Onorati; Giorgio Morando; Giuseppe Faggian; Alessandro Mazzucco
Journal:  Ann Thorac Surg       Date:  2012-02       Impact factor: 4.330

Review 4.  Erosions, erosions, and erosions! Device closure of atrial septal defects: how safe is safe?

Authors:  Karim Diab; Damien Kenny; Ziyad M Hijazi
Journal:  Catheter Cardiovasc Interv       Date:  2012-08-01       Impact factor: 2.692

5.  Aorta-to-left atrial fistula developing after surgical removal of an atrial septal occlusion device eight years after original implantation.

Authors:  Kisha Beg; Larry A Latson; Gosta Pettersson; Lee Wallace; Athar M Qureshi
Journal:  World J Pediatr Congenit Heart Surg       Date:  2015-04

6.  Secundum atrial septal defect. Nonoperative closure during cardiac catheterization.

Authors:  T D King; S L Thompson; C Steiner; N L Mills
Journal:  JAMA       Date:  1976-06-07       Impact factor: 56.272

7.  Late aortic perforation with a fractured ATRIASEPT II device resulting in life-threatening tamponade.

Authors:  Francisco José Hernández Pérez; José Antonio Fernández Díaz; Carlos García Montero; Arturo García Touchard; Juan Francisco Oteo Domínguez; José Ramón Domínguez Puente; Javier Goicolea Ruigómez
Journal:  EuroIntervention       Date:  2013-08-22       Impact factor: 6.534

Review 8.  Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder.

Authors:  Abhay Divekar; Tidimogo Gaamangwe; Nasir Shaikh; Michael Raabe; John Ducas
Journal:  J Am Coll Cardiol       Date:  2005-04-19       Impact factor: 24.094

9.  The influence of morphological changes in amplatzer device on the atrial and aortic walls following transcatheter closure of atrial septal defects.

Authors:  Masataka Kitano; Satoshi Yazaki; Hisashi Sugiyama; Osamu Yamada
Journal:  J Interv Cardiol       Date:  2009-02       Impact factor: 2.279

10.  Erosion by the Amplatzer septal occluder: experienced operator opinions at odds with manufacturer recommendations?

Authors:  Howaida G El-Said; John W Moore
Journal:  Catheter Cardiovasc Interv       Date:  2009-06-01       Impact factor: 2.692

View more
  8 in total

1.  Comparison of Figulla Flex® and Amplatzer™ devices for atrial septal defect closure: A meta-analysis.

Authors:  Alvaro Aparisi; Roman J Arnold; Hipólito Gutiérrez; Ana Revilla; Ana Serrador; Benigno Ramos; Tania Rodriguez-Gabella; Alberto Campo; Carlos Baladrón; Itziar Gómez; Manuel Carrasco-Moraleja; José A San Roman; Ignacio J Amat-Santos
Journal:  Cardiol J       Date:  2020-04-24       Impact factor: 2.737

2.  Coronary Sinus Defect Following Transcatheter Closure of ASD Using Amplatzer Septal Occluder: Potential Erosion by the Device.

Authors:  Bassel Mohammad Nijres; Maytham Al-Kubaisi; John Bokowski; Ra-Id Abdulla; Sawsan Awad
Journal:  Pediatr Cardiol       Date:  2017-04-10       Impact factor: 1.655

Review 3.  Atrial septal defect (ASD) device trans-catheter closure: limitations.

Authors:  Alessia Faccini; Gianfranco Butera
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

4.  Late Erosion of Percutaneous ASD Occlusion Device.

Authors:  Ronan W Cusack; Gearoid Fitzgerald; Mark W Kennedy
Journal:  Eur J Case Rep Intern Med       Date:  2020-12-24

5.  Cocoon devices for transcatheter closure of atrial septal defect and patent ductus arteriosus in children: Single center experience.

Authors:  Hyam Mahmoud; Alin Nicolescu; Cristina Filip; Gabriela Duica; Georgiana Nicolae; Eliza Elena Cinteza
Journal:  Medicine (Baltimore)       Date:  2019-03       Impact factor: 1.817

6.  Simplified percutaneous closure of patent foramen ovale and atrial septal defect with use of plain fluoroscopy: Single operator experience in 110 consecutive patients.

Authors:  Antonis S Manolis; Spyridon Koulouris; Efthymia Rouska; John Pyrros
Journal:  Indian Heart J       Date:  2017-08-08

7.  Transcatheter atrial septal defect closure in an infant (body weight 6.4 kg) using the GORE CARDIOFORM septal occluder (GCSO).

Authors:  Roman Scheidmann; Thomas Paul; Matthias Sigler
Journal:  Mol Cell Pediatr       Date:  2017-11-03

Review 8.  Echocardiographic approach to catheter closure of atrial septal defects: patient selection, procedural guidance and post-procedural checks.

Authors:  Alexandra Savis; John Simpson
Journal:  Echo Res Pract       Date:  2018-03-27