Development and validation of reversed phase high-performance liquid chromatography method for estimation of lercanidipine HCl in pure form and from nanosuspension formulation.

Ankita D Chonkar1, Renuka S Managuli1, Josyula Venkata Rao2, Nayanabhirama Udupa3.   

Abstract

AIM: Quantitative estimation of lercanidipine HCl in bulk material as well as from nanosuspension formulations via a developed reverse phase HPLC method.
MATERIALS AND METHODS: Optimized chromatographic condition was used to achieve separation on a Kromasil (100-5c18 250 × 4.6 mm) column using Shimadzu HPLC system. The mobile phase consisted of a mixture of acetate buffer (20 mM pH 4.5) and acetonitrile in the ratio of 10:90, v/v. It is pumped through the chromatographic system at a flow rate of 1 ml/min. The detection was carried out at 240 nm using ultraviolet-visible spectrophotometry detector. The method was validated as per Q2 (R1) guidelines, and suitability of the developed method was established by optimized nanosuspension formulation.
RESULTS: The method is specific to lercanidipine (RT: 7.7 min), and has ability to resolve the analyte peak from excipient interferences. It is linear (regression coefficient: 0.9993), accurate (average recovery: 100%), and passed all the system suitability requirements.
CONCLUSION: Developed method was found applicable for evaluation of drug content, content uniformity, and analyzing samples of dissolution studies of nanosuspension.

Entities:  

Keywords:  Lercanidipine HCl; liquid chromatography; nanosuspension; validation

Year:  2015        PMID: 26792958      PMCID: PMC4702276          DOI: 10.4103/0976-0105.170586

Source DB:  PubMed          Journal:  J Basic Clin Pharm        ISSN: 0976-0113


Introduction

Lercanidipine, 5-O-(1-3,3-diphenylpropyl(methyl)amino] -2-methylpropan-2-yl) 3-O-methyl 2,6-dimethyl-4- (3-nitrophenyl)-l, 4-dihydropyridine-3,5-dicarboxylate is a dihydropyridine calcium antagonist [Figure 1]. It is used for the management of Stage I and Stage II hypertension and is also perhaps useful in relieving angina pectoris.[12]
Figure 1

Structure of lercanidipine hydrochloride

Structure of lercanidipine hydrochloride Lercanidipine HCl is soluble in dimethylformamide, dichloromethane, and methanol.[3] It has a pKa value of about 6.83.[4] It is extremely lipophilic, possess octanol: water partition coefficient (Log P) value of about 6.4.[1] Preclinical and clinical findings propose that lercanidipine may have protective effects on the kidneys, cardiovascular system, and target organs. Lercanidipine (10 mg/day) produces a smooth antihypertensive effect without unfavorable hemodynamic or sympathetic effects due to its vascular selectivity. Lercanidipine emerged as a flexible choice for antihypertensive treatment across a wide range of patients due to its favorable efficacy and safety profile. Lercanidipine attains maximum plasma concentration within 2–3 h after oral administration and exhibits a slow onset of action. However, it is extensively metabolized by cytochrome P450 3A4.[5] Its low water solubility (5 μg/ml), poor permeability, extensive first pass metabolism, and food dependent absorption result in its low bioavailability of 10%.[2] There are few HPLC methods described in literature for the analysis of lercanidipine hydrochloride, its impurities and degradation products in bulk as well as in commercial tablets.[67] In the present research work, nanosuspensions of lercanidipine hydrochloride were formulated using a wide range of surfactants and polymers as nanosuspension stabilizers and analyzed by proposed method. As lercanidipine is a poorly bioavailable drug, we may expect continuing the research work, in the context of its solubility and permeation enhancement using a broad range of excipients. Therefore, the proposed method has been developed and validated with a definite aim of having a method, which is simple in operation, cost-effective, and able to analyze bulk material, uniformity of dosage, in vitro release samples, and to detect small changes that may occur when drug is processed with various excipients. The method is validated as per ICH Q2 (R1) (validation of analytical procedures: Text and Methodology) guideline.[8] Typical validation characteristics such as accuracy, precision, repeatability, intermediate precision, specificity, limit of detection, limit of quantification (LOQ), linearity, range, and robustness were evaluated.

Materials and Methods

Instrumentation

A Shimadzu HPLC system (LC 2010 HT) equipped with degassing unit, low-pressure gradient unit, pump unit, ultra-fast autosampler, column oven, and an ultraviolet-visible spectrophotometry detector with a thermostatted flow cell was utilized. LC solution software (version 1.24 SP 1, It is the software by SHIMADZU for processing of chromatographic data. Make and model of HPLC system used mentioned under Instrumentation section) was used for data acquisition and system suitability calculations. In addition, Equitron bath sonicator, Millipore Direct Q-3 water purification system, and electronic balances (Denver Instruments, USA and Sartorius, India) were used in the study. Chromatographic parameters used for the determination of lercanidipine HCl are given in Table 1.
Table 1

Chromatographic parameters for determination of lercanidipine HCl

Chromatographic parameters for determination of lercanidipine HCl

Materials

Lercanidipine HCl hemihydrate obtained as gift sample from Lupin Limited, Research Park, Pune. HPLC grade methanol and acetonitrile were obtained from Finar Chemicals, Gujarat (India). All the surfactants and/or polymers were procured from Loba Chemie Ltd. Mumbai and Sigma-Aldrich Bengaluru, India. Lercanidipine (LR) nanosuspensions were prepared by precipitation technique. 50 mg of Lercanidipine was dissolved in sufficient volume (500 µl) of methanol. Several polymers and/or surfactants such as PEG (polyethylene glycol) 400, HPMC (hydroxypropyl methyl cellulose) E15, PVA (poly vinyl alcohol), sodium alginate, methyl cellulose, HPMC E5, and SLS (sodium lauryl sulphate) were used singly or in a combination to act as a stabilizer. The polymers/surfactants were dissolved in water (10 ml) separately. The resulting mixture was kept under high-speed homogenization (Polytron PT 3100) for 15 min at 10000 rpm. The temperature of the stabilizer solution was maintained at 10°C. The drug solution was added all at once in stabilizer solution kept under high-speed homogenization. After complete addition of the drug solution, homogenization was continued for 15 min to get nanosuspension. The developed method was validated for the parameters such as accuracy, precision, repeatability, intermediate precision, specificity, LOD, LOQ, linearity, range, system suitability, and robustness as described below. The linearity of an analytical method is its capability to obtain a test result, which has a certain mathematical relationship to the concentration of analytes. A standard solution of lercanidipine HCl (500 µg/mL) was prepared by dissolving exactly weighed 5 mg of lercanidipine HCl in 10 ml methanol. Different volumes of stock solution were transferred into 10 ml volumetric flasks separately and diluted with mobile phase to yield 5.0–25.0 µg/ml concentration range. Each dilution was prepared in duplicate. Areas for five injections were determined and graph was prepared. Slope and intercept were estimated.

Method precision (repeatability)

Method precision (repeatability) is the result of the method working over a short time interval under the identical conditions (inter-assay precision). Six replicates of standard solution of lercanidipine HCl (10.0 μg/mL) were analyzed, and chromatograms were recorded. The mean area and % relative standard deviation (RSD) for those injections were calculated.

Intermediate precision

Intermediate precision of the method was checked by repeating the entire procedure on the next day for 3 replicates of standard solution of lercanidipine HCl (10.0 μg/mL). The % RSD for response area and retention time was calculated.

Formulation precision

Prepared nanosuspensions were assessed using the present analytical method. Theoretical drug content of nanosuspension is 5 mg/ml. 1 ml aliquot of nanosuspension was removed from 6 individual batches of nanosuspension and transferred to 10 ml volumetric flasks separately. Volume was made up with methanol and mixed thoroughly using vortex mixer to extract the drug completely. These solutions were filtered through 0.45 µm cellulose membrane filter paper and diluted appropriately to get the concentration of 10 μg/ml. These 6 solutions were injected continuously and calculated the RSD for area and retention time.

Specificity of the method

To determine the specificity of the method in existence of excipients, matrix consisting of several excipients used in final formulation was prepared in 10 ml of mobile phase and solution was filtered through 0.45 μm cellulose membrane filter. Necessary dilutions were made and 20 μl of this solution was injected on column, and peak response was noted.

Accuracy

The developed analytical method was validated for its accuracy in determining the drug content from solution, from the excipient blend, and from formulation.[9]

Recovery from drug solution

The accuracy of the method was performed by recovery study. Accuracy was executed in triplicate for various concentrations of lercanidipine equivalent to 50%, 100%, and 150% of the standard amount. Samples were injected into the HPLC system as per the test procedure. The average % recovery of lercanidipine was calculated.[9]

Recovery from excipient blend (assay by spiking)

Recovery studies from excipients blend was performed by spiking a definite amount of drug solution (50, 100, and 150% of assay concentration) in excipient matrix. The solutions were prepared in methanol in triplicate. The samples were mixed thoroughly using vortex mixer and allowed to dry in a dark place. The dried blend was then reconstituted with mobile phase in 10 ml volumetric flask and this solution was sonicated for 30 min in a bath sonicator. The solutions were filtered through 0.45 μm cellulose membrane filter. Necessary dilutions were made and 20 μl of these solutions were injected. Percentage of drug recovered was calculated using a standard curve prepared. Retention time and peak shape were noted in the presence and absence of excipient blend.[910]

Recovery from formulation

Prepared nanosuspensions were evaluated using the present analytical method to determine the drug content. Theoretical drug content of nanosuspension is 5 mg/ml. Different aliquots of 1 ml, 2 ml, and 3 ml were taken from nanosuspensions and transferred to 10 ml volumetric flask separately. Volume was made up with methanol and mixed thoroughly using vortex mixer to extract the drug completely. These solutions were filtered through 0.45 µm cellulose membrane filter paper and diluted appropriately to get the concentrations of 5 µg/ml, 10 µg/ml, and 15 µg/ml (50, 100, and 150% of assay concentration).

Filter validation

Filter validation was performed by analyzing solutions at 5 and 40 μg/ml (the lowest and highest concentrations of the working solutions). The solutions were analyzed in duplicate after filtration through 0.45 μm cellulose membrane filter. The results were compared with the unfiltered sample injected at the same concentration levels.[10]

System suitability

Data from six injections (at 100% assay concentration) were used for computing system suitability parameters such as tailing factor and number of theoretical plates using software.[9]

Limit of detection and limit of quantification

The LOD and LOQ were calculated on the basis of response and slope of regression equation. LOD = 3.3 × σ/S and LOD = 10 × σ/S, where σ = the standard deviation of the response, S = the slope of the calibration curve of analyte.[9111213] The robustness of an analytical procedure is an extent of its ability to remain unaffected by small, but deliberate changes in method parameters and provides a clue of its reliability during normal usage. The factors chosen for this study were detection wavelength (nm), temperature (°C), flow rate (ml/min), mobile phase (percentage acetonitrile), and pH of the acetate buffer. These factors were changed as per manner shown in Table 2.[9]
Table 2

Variations in method parameters for robustness studies

Variations in method parameters for robustness studies Influence of change in the above factors on analysis parameters such as response area, retention time, tailing factor, and theoretical plates were determined by calculating correlation coefficient using “one-factor response surface method” (Stat-Ease Design Expert® Software Version 9).

Results and Discussion

The chromatogram of lercanidipine acquired by developed method is indicated in Figure 2. Lercanidipine elutes at retention time 7.73 min with an average tailing factor of 0.8.
Figure 2

Representative chromatogram of lercanidipine using a developed method

Representative chromatogram of lercanidipine using a developed method

Linearity

Table 3 indicates the regression statistics obtained for linearity test. The linearity of an analytical method is its capability to produce test results that are directly proportional to the concentration of analyte within a given range [Figure 3]. The method was found linear in the expected concentration range. The regression coefficient was found to be 0.9993, demonstrating its suitability for analysis.
Table 3

Regression statistics for lercanidipine HCl

Figure 3

Linearity plot for lercanidipine HCl

Regression statistics for lercanidipine HCl Linearity plot for lercanidipine HCl

Repeatability

The precision of an analytical method is the extent of agreement between individual test results when the method is applied repetitively to multiple sampling of homologous sample. The method passed the test for repeatability, as percent RSD (0.7%) of the area of the peaks of six replicates injection at 100% assay concentration was within the limits of 2%. % RSD of response area and retention time was calculated and presented in Table 4.[9]
Table 4

Results of precision

Results of precision The procedure followed for assay method in method precision was repeated on the next day. As the percent RSD Table 4 of the response areas obtained was within the limits of 2%; method passed the test for intermediate precision. Therefore, proposed analytical technique has a good intermediate precision.[9] Specificity is the capability of an analytical method to assess unambiguously the analyte in the presence of other components that are present in the sample matrix. The representative chromatogram [Figure 4] of excipient blend indicates that excipients do not interfere with the drug peak [Figure 2], which shows the specificity of the method for lercanidipine.
Figure 4

Chromatogram of excipient blend

Chromatogram of excipient blend When the developed method was used for the estimation of lercanidipine in dosage form, it was evaluated for the spectral purity in the diode array detector. The purity of the peak constituting for lercanidipine passes the test. The total peak purity index (1.000000) was found to be greater than the single point threshold (0.999714) [Figure 5]. It indicates that the method is highly specific and no other components were co-eluting with the analytes.
Figure 5

Peak purity curve of lercanidipine HCl

Peak purity curve of lercanidipine HCl System suitability test is an essential part of chromatographic methods and performed to confirm that the resolution and reproducibility of the system are suitable for the analysis to be done. The results of system suitability are given in Table 5. All the values for the system suitability parameters are well within the acceptance criteria.[9]
Table 5

System suitability parameters

System suitability parameters The accuracy of an analytical method determines the closeness of test results obtained by the method to the true value. It can be determined by applying analytical procedure to an analyte of known purity (for a drug substance) or by recovery studies, where known amount of standard is spiked in the placebo (for the drug product). The results of accuracy studies from drug solution, excipient matrix, and formulation are shown in Tables 6 and 7, and it is observed that the method is accurate within desired range. Low % RSD values of drug recovery from formulation at each concentration level indicate uniformity of drug content in dosage form[9].
Table 6

Accuracy/recovery data for lercanidipine HCl from solution

Table 7

Accuracy/recovery data for lercanidipine HCl from excipient matrix and formulation

Accuracy/recovery data for lercanidipine HCl from solution Accuracy/recovery data for lercanidipine HCl from excipient matrix and formulation According to the study carried out, the LOD and LOQ of lercanidipine were found 0.219 µg/ml and 0.663 µg/ml, respectively. The RSD obtained at lower concentration (0.68%) and higher concentration (0.59%) shows suitability of the cellulose membrane filter for the dissolution sample filtration, as the RSD is <2%.[10] Effects of change in factors such as detection wavelength (nm), temperature (°C), flow rate (ml/min), mobile phase (percentage acetonitrile), and pH of the acetate buffer on analysis parameters such as response area, retention time, tailing factor, and theoretical plates were observed in robustness studies and results obtained were indicated in Table 8.
Table 8

Effect of changes in independent factors on chromatographic profile

Effect of changes in independent factors on chromatographic profile Effects of changes in independent factors on chromatographic profile were calculated by determining correlation coefficient using “One-Factor Response Surface Method” (Stat-ease Design Expert 9) [Table 9]. All the independent factors are listed in different columns and all the analysis parameters are listed in different rows. Correlation coefficient of particular factor and analysis parameter is indicated by the intersection of respective column and row. The value of correlation coefficient always lies between −1 and +1.
Table 9

Correction coefficients to show the effect of the different variables studied on chromatographic profile of lercanidipine HCl

Correction coefficients to show the effect of the different variables studied on chromatographic profile of lercanidipine HCl Positive value of correlation coefficient indicates a positive association. Negation value of correction coefficient indicates a negative association. The value of correlation coefficient in between ± 0.70 and ± 1 indicates a strong association. Therefore, it can be observed from the above table that change in column temperature has a strong influence on the parameters such as retention time and number of theoretical plates. Change in detection wavelength does not influence tailing factor at all. Flow rate has a strong influence on all the analysis parameters. The value of correlation coefficient in between ± 0.50 and ± 0.70 indicates a moderate association. pH of acetate buffer and peak area have a moderate positive association. The value of correlation coefficient in between ± 0.30 to ± 0.50 indicates a low association. The value of correlation coefficient in between - 0.30 to + 0.30 indicates no association. This indicated that there is a no correlation between column temperature and mean peak area.

Conclusion

A simple and quick analytical method has been developed to be useful in routine to determine lercanidipine in bulk and in its dosage forms. The method proposed by HPLC to determine lercanidipine in nanosuspension has been proved in a linear, precise, accurate, specific way, and robust about the wavelength, flow rate, mobile phase, and temperature.

Financial support and sponsorship

Dr. T. M. A. Pai Endowment Chair, Manipal University, Manipal 576 104.

Conflicts of interest

There are no conflicts of interest.
  6 in total

1.  A validated high performance liquid chromatographic method for analysis of nicotine in pure form and from formulations.

Authors:  Kaustubh R Tambwekar; Ritesh B Kakariya; Sanjay Garg
Journal:  J Pharm Biomed Anal       Date:  2003-07-14       Impact factor: 3.935

2.  A selective HPLC method for determination of lercanidipine in tablets.

Authors:  A Alvarez-Lueje; S Pujol; J A Squella; L J Núñez-Vergara
Journal:  J Pharm Biomed Anal       Date:  2003-02-05       Impact factor: 3.935

3.  Development and validation of a new HPLC analytical method for the determination of alprazolam in tablets.

Authors:  P Pérez-Lozano; E García-Montoya; A Orriols; M Miñarro; J R Ticó; J M Suñé-Negre
Journal:  J Pharm Biomed Anal       Date:  2004-03-10       Impact factor: 3.935

4.  A Stability-indicating HPLC Method for Assay of Lercanidipine Hydrochloride in Tablets and for Determining Content Uniformity.

Authors:  H O Kaila; M A Ambasana; R S Thakkar; H T Saravaia; A K Shah
Journal:  Indian J Pharm Sci       Date:  2010-05       Impact factor: 0.975

Review 5.  Lercanidipine in hypertension.

Authors:  Claudio Borghi
Journal:  Vasc Health Risk Manag       Date:  2005

6.  Development and validation of reversed-phase high-performance liquid chromatography method for estimation of rizatriptan benzoate in oral strip formulations.

Authors:  S T Bhagawati; M Sreenivasa Reddy; Kiran Avadani; B S Muddukrishna; Swapnil J Dengale; Krishnamurthy Bhat
Journal:  J Basic Clin Pharm       Date:  2014-12
  6 in total
  2 in total

1.  Analysis of PK11195 concentrations in rodent whole blood and tissue samples by rapid and reproducible chromatographic method to support target-occupancy PET studies.

Authors:  Agnė Stadulytė; Carlos José Alcaide-Corral; Tashfeen Walton; Christophe Lucatelli; Adriana Alexandre S Tavares
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2019-04-12       Impact factor: 3.205

2.  Development of three ecological spectroscopic methods for analysis of betrixaban either alone or in mixture with lercanidipine: greenness assessment.

Authors:  Amal A El-Masry; Dalia R El-Wasseef; Manal Eid; Ihsan A Shehata; Abdallah M Zeid
Journal:  R Soc Open Sci       Date:  2022-02-02       Impact factor: 2.963

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.