Literature DB >> 26787677

High variability in strain estimation errors when using a commercial ultrasound speckle tracking algorithm on tendon tissue.

Åsa Fröberg1, Mattias Mårtensson2, Matilda Larsson2, Birgitta Janerot-Sjöberg3, Jan D'Hooge4, Anton Arndt5.   

Abstract

BACKGROUND: Ultrasound speckle tracking offers a non-invasive way of studying strain in the free Achilles tendon where no anatomical landmarks are available for tracking. This provides new possibilities for studying injury mechanisms during sport activity and the effects of shoes, orthotic devices, and rehabilitation protocols on tendon biomechanics.
PURPOSE: To investigate the feasibility of using a commercial ultrasound speckle tracking algorithm for assessing strain in tendon tissue.
MATERIAL AND METHODS: A polyvinyl alcohol (PVA) phantom, three porcine tendons, and a human Achilles tendon were mounted in a materials testing machine and loaded to 4% peak strain. Ultrasound long-axis cine-loops of the samples were recorded. Speckle tracking analysis of axial strain was performed using a commercial speckle tracking software. Estimated strain was then compared to reference strain known from the materials testing machine. Two frame rates and two region of interest (ROI) sizes were evaluated.
RESULTS: Best agreement between estimated strain and reference strain was found in the PVA phantom (absolute error in peak strain: 0.21 ± 0.08%). The absolute error in peak strain varied between 0.72 ± 0.65% and 10.64 ± 3.40% in the different tendon samples. Strain determined with a frame rate of 39.4 Hz had lower errors than 78.6 Hz as was the case with a 22 mm compared to an 11 mm ROI.
CONCLUSION: Errors in peak strain estimation showed high variability between tendon samples and were large in relation to strain levels previously described in the Achilles tendon. © The Foundation Acta Radiologica 2016.

Entities:  

Keywords:  Achilles tendon; Speckle tracking; strain; ultrasound

Mesh:

Year:  2016        PMID: 26787677     DOI: 10.1177/0284185115626471

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  3 in total

1.  The challenges of measuring in vivo knee collateral ligament strains using ultrasound.

Authors:  Laura C Slane; Josh A Slane; Jan D'hooge; Lennart Scheys
Journal:  J Biomech       Date:  2017-07-31       Impact factor: 2.712

2.  Altered patterns of displacement within the Achilles tendon following surgical repair.

Authors:  Åsa Fröberg; Ann-Sophie Cissé; Matilda Larsson; Mattias Mårtensson; Michael Peolsson; Tomas Movin; Anton Arndt
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-12-21       Impact factor: 4.342

3.  Evaluation of tissue displacement and regional strain in the Achilles tendon using quantitative high-frequency ultrasound.

Authors:  Stijn Bogaerts; Catarina De Brito Carvalho; Lennart Scheys; Kaat Desloovere; Jan D'hooge; Frederik Maes; Paul Suetens; Koen Peers
Journal:  PLoS One       Date:  2017-07-20       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.