Determining the Minimal Required Radioactivity of 18F-FDG for Reliable Semiquantification in PET/CT Imaging: A Phantom Study.

Ming-Kai Chen1, David H Menard2, David W Cheng3.   

Abstract

UNLABELLED: In pursuit of as-low-as-reasonably-achievable (ALARA) doses, this study investigated the minimal required radioactivity and corresponding imaging time for reliable semiquantification in PET/CT imaging.
METHODS: Using a phantom containing spheres of various diameters (3.4, 2.1, 1.5, 1.2, and 1.0 cm) filled with a fixed (18)F-FDG concentration of 165 kBq/mL and a background concentration of 23.3 kBq/mL, we performed PET/CT at multiple time points over 20 h of radioactive decay. The images were acquired for 10 min at a single bed position for each of 10 half-lives of decay using 3-dimensional list mode and were reconstructed into 1-, 2-, 3-, 4-, 5-, and 10-min acquisitions per bed position using an ordered-subsets expectation maximum algorithm with 24 subsets and 2 iterations and a gaussian 2-mm filter. SUVmax and SUVavg were measured for each sphere.
RESULTS: The minimal required activity (±10%) for precise SUVmax semiquantification in the spheres was 1.8 kBq/mL for an acquisition of 10 min, 3.7 kBq/mL for 3-5 min, 7.9 kBq/mL for 2 min, and 17.4 kBq/mL for 1 min. The minimal required activity concentration-acquisition time product per bed position was 10-15 kBq/mL⋅min for reproducible SUV measurements within the spheres without overestimation. Using the total radioactivity and counting rate from the entire phantom, we found that the minimal required total activity-time product was 17 MBq⋅min and the minimal required counting rate-time product was 100 kcps⋅min.
CONCLUSION: Our phantom study determined a threshold for minimal radioactivity and acquisition time for precise semiquantification in (18)F-FDG PET imaging that can serve as a guide in pursuit of achieving ALARA doses.
© 2016 by the Society of Nuclear Medicine and Molecular Imaging, Inc.

Entities:  

Keywords:  PET; acquisition time; dose reduction; dosimetry; phantom; standard uptake value

Mesh:

Substances:

Year:  2016        PMID: 26769598     DOI: 10.2967/jnmt.115.165258

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  3 in total

1.  Nuclear medicine RIP (radiation induced phobia); improving the image.

Authors:  V Ralph McCready; Sabina Dizdarevic
Journal:  Eur J Nucl Med Mol Imaging       Date:  2018-09-25       Impact factor: 9.236

2.  Variation and repeatability of measured standardized uptake values depending on actual values: a phantom study.

Authors:  Tomohiro Kaneta; Na Sun; Matsuyoshi Ogawa; Hitoshi Iizuka; Tetsu Arisawa; Ayako Hino-Shishikura; Keisuke Yoshida; Tomio Inoue
Journal:  Am J Nucl Med Mol Imaging       Date:  2017-11-01

3.  Reduction of the fluorine-18-labeled fluorodeoxyglucose dose for clinically dedicated breast positron emission tomography.

Authors:  Yoko Satoh; Tetsuro Sekine; Yoshie Omiya; Hiroshi Onishi; Utaroh Motosugi
Journal:  EJNMMI Phys       Date:  2019-11-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.