Localization errors in mantle-field irradiation for Hodgkin's disease.

M Hulshof1, L Vanuytsel, W Van den Bogaert, E van der Schueren.   

Abstract

The frequency and magnitude of localization errors detected by verification films, during mantle-field irradiation for Hodgkin's disease, were assessed. A total of 216 treatment set-ups was studied. The first verification film, at the start of treatment, showed a localization error of over 1 cm in 13% of cases, leading to a critical margin between the shielding block and the tumor bearing area in 9% of the treatment set-ups. After the first correction, an adequate treatment set-up was obtained in 60% of cases, whereas after two corrections the adequacy rate was 84%. Verification films are therefore very useful in detecting clinically important localization errors and in monitoring and checking subsequent corrections. An additional advantage exists because the first verification film can be used as a final check of the simulation plan itself.

Entities:  

Mesh:

Year:  1989        PMID: 2674080     DOI: 10.1016/0360-3016(89)90124-7

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  Reproducibility of irregular radiation fields for malignant lymphoma.

Authors:  U Mock; K Dieckmann; A M Molitor; U Haverkamp; R Pötter
Journal:  Strahlenther Onkol       Date:  1998-10       Impact factor: 3.621

2.  Patient-specific daily pretreatment setup protocol using electronic portal imaging for radiation therapy.

Authors:  Michael H Wittmer; Thomas M Pisansky; Jon J Kruse; Michael G Herman
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

3.  Guide to clinical use of electronic portal imaging.

Authors:  M G Herman; J J Kruse; C R Hagness
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.