Abstract
A new highly specific and sensitive monoclonal antibody (MAb) to T-2 toxin (T-2) was produced, providing an IC50 value of 1.02 ng/mL and negligible cross-reactivity (CR) to other related mycotoxins. Based on the new MAb, a lateral-flow immunochromatographic assay (LFIA) using colloidal gold (CG) and fluorescent microspheres (FMs) as labels was proposed for T-2. Under the optimized conditions, in rapid qualitative assay, the cut-off values of the CG-LFIA were 400 μg/kg in rice and 50 μg/L in fresh milk, and the cut-off values of the FMs-LFIA were 100 μg/kg in both rice and chicken feed. For the quantitative assay with the FMs-LFIA, the limit of detection (LOD) were 0.23 μg/kg and 0.41 μg/kg in rice and chicken feed, respectively, and the average recoveries ranged from 80.2% to 100.8% with the coefficient of variation (CV) below 10.8%. In addition, we found that the CG-LFIA could tolerate the matrix effect of fresh milk better than the FMs-LFIA, while the FMs-LFIA could tolerate the matrix effect of chicken feed better than CG-LFIA under the same experimental conditions. These results provide a certain reference for the selection of appropriate labels to establish a rapid LFIA in various biological samples.Entities:
Keywords: T-2 toxin; colloidal gold; fluorescent microsphere; lateral-flow immuno-chromatographic assay; monoclonal antibody
Mesh:
Substances:
Year: 2015 PMID: 26729073 PMCID: PMC6273043 DOI: 10.3390/molecules21010027
Source DB: PubMed Journal: Molecules ISSN: 1420-3049 Impact factor: 4.411
Figure 1Standard curves. (A) represents the results of the immunization from the six BALB/c mice; and (B) represents the standard curve of T-2 in PBS.
Figure 2LFIA optimization. (A) Different NC membranes; (B) Amount of MAb for conjugation with CG; (C) FMs diameter; (D) Amount MAb loaded onto FMs; (E) Amounts of EDC; (F) Incubation time.
Figure 3Detection of T-2 in PBS by the CG-LFIA (A); and FMs-LFIA (B).
Figure 4Calibration curves for T-2 in PBS, rice and chicken feed by the FMs-LFIA.
LOD, IC50, IC20–IC80 and R2 of T-2 in PBS, rice and chicken feed by FMs-LFIA.
| PBS | Rice | Chicken Feed | |
|---|---|---|---|
| LOD | 0.28 | 0.23 | 0.41 |
| IC50 (ng/mL) | 1.58 | 1.78 | 1.78 |
| IC20~IC80 (ng/mL) | 0.28–8.9 | 0.23–13.7 | 0.41–7.8 |
| R2 | 0.990 | 0.987 | 0.989 |
Figure 5Detection of T-2 in rice (A); fresh milk (C); and chicken feed (E) by the CG-LFIA; and in rice (B); fresh milk (D); and chicken feed (F) by the FMs-LFIA, respectively.
Recoveries and CV values for T-2 in rice and chicken feed.
| Sample | Spiked (μg/kg) | Test (μg/kg) | Recoveries (%) | CV (%) |
|---|---|---|---|---|
| Rice | 5 | 4.748 | 94.9 | 5.42 |
| 10 | 10.08 | 100.8 | 7.51 | |
| 20 | 19.22 | 96.1 | 9.78 | |
| Chicken Feed | 5 | 4.02 | 80.2 | 10.8 |
| 10 | 8.901 | 89 | 9.87 | |