Archwire diameter effect on tooth alignment with different bracket-archwire combinations.

Mona A Montasser1, Ludger Keilig2, Christoph Bourauel3.   

Abstract

INTRODUCTION: Our objective was to evaluate the effect of the diameter of the archwire on tooth alignment with different bracket-archwire combinations.
METHODS: The materials included 2 categories of orthodontic brackets (1) conventional ligating brackets (Victory Series [3M Unitek, Monrovia, Calif], Mini-Taurus [Rocky Mountain Orthodontics, Denver, Colo], and Synergy [Rocky Mountain Orthodontics]) and (2) self-ligating brackets (SmartClip [3M Unitek], a passive self-ligating bracket; Time3 [American Orthodontics, Sheboygan, Wis], an active self-ligating bracket; and SPEED [Strite Industries, Cambridge, Ontario, Canada], an active self-ligating bracket). All brackets had a nominal 0.022-in slot size. The brackets were combined with Therma-Ti 0.014-in and Therma-Ti 0.016-in titanium memory archwires (American Orthodontics). The archwires were tied to the conventional brackets with both stainless steel ligatures (0.010 in) and elastomeric rings. Each bracket-archwire combination was tested 20 times with the orthodontic measurement and simulation system built in a temperature-controlled chamber where the temperature was kept at 37°C (±1°C) during testing. The malocclusion simulated in the study represented a maxillary central incisor displaced 2 mm gingivally (x-axis) and 2 mm labially (z-axis).
RESULTS: The incisogingival corrections achieved by the 0.014-in archwire combined with the brackets used ranged from 40% to 95%; the corrections by the 0.016-in wire were 55% to 95%. The labiolingual corrections achieved by the 0.014-in archwire combined with the brackets used ranged from 10% to 100%, and the corrections of the 0.016-in archwires ranged from 15% to 100%.
CONCLUSIONS: Increasing the diameter from 0.014 to 0.016 in increased the correction achieved by up to 15% in certain bracket-archwire combinations, but it decreased the correction by up to 25% in other combinations.
Copyright © 2016 American Association of Orthodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26718381     DOI: 10.1016/j.ajodo.2015.06.026

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  5 in total

1.  Super-elasticity in vitro assessment of CuNiTi wires according to their Austenite finish temperature and the imposed displacement.

Authors:  Noémie Copelovici; Maï-Linh Tran; François Lefebvre; Pascal Laheurte; Delphine Wagner
Journal:  Angle Orthod       Date:  2022-05-01       Impact factor: 2.684

2.  Evaluation of the load system produced by a single intrusion bend in a maxillary lateral incisor bracket with different alloys.

Authors:  Ricardo Lima Shintcovsk; Roberto Soares da Silva Júnior; Larry White; Lidia Parsekian Martins; Renato Parsekian Martins
Journal:  Angle Orthod       Date:  2018-05-15       Impact factor: 2.079

3.  Evaluation of the force generated by gradual deflection of orthodontic wires in conventional metallic, esthetic, and self-ligating brackets.

Authors:  Manoela Fávaro Francisconi; Guilherme Janson; José Fernando Castanha Henriques; Karina Maria Salvatore de Freitas
Journal:  J Appl Oral Sci       Date:  2016 Sep-Oct       Impact factor: 2.698

4.  Buccolingual Inclination Control of Upper Central Incisors of Aligners: A Comparison with Conventional and Self-Ligating Brackets.

Authors:  Maria Francesca Sfondrini; Paola Gandini; Tommaso Castroflorio; Francesco Garino; Luca Mergati; Krizia D'Anca; Federico Trovati; Andrea Scribante
Journal:  Biomed Res Int       Date:  2018-11-29       Impact factor: 3.411

5.  The Influence of Friction on Design of the Type of Bracket and Its Relation to OHRQoL in Patients Who Use Multi-Bracket Appliances: A Randomized Clinical Trial.

Authors:  Adriana González-Sáez; Laura Antonio-Zancajo; Javier Montero; Alberto Albaladejo; María Melo; Daniele Garcovich; Alfonso Alvarado-Lorenzo
Journal:  Medicina (Kaunas)       Date:  2021-02-17       Impact factor: 2.430

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.