Literature DB >> 26709071

Portable capillary electrophoresis instrument with contactless conductivity detection for on-site analysis of small volumes of biological fluids.

Michal Greguš1, František Foret2, Petr Kubáň3.   

Abstract

A novel, easy to use and portable capillary electrophoretic instrument for injection of small volumes of biological fluids equipped with contactless conductivity detection was constructed. The instrument is lightweight (<5 kg), all necessary parts including a tablet computer are accommodated in a plastic briefcase with dimensions 20 cm × 33 cm × 17 cm (w × l × h), allows hydrodynamic injection of small sample volumes and can continuously operate for at least 10 hours. The semi-automated hydrodynamic sample injection is accomplished via a specially designed PMMA interface that is able to repeatedly inject sample aliquots from a sample volume as low as 10 μL, with repeatability of peak areas below 5%. The developed interface and the instrument were optimized for the injection of biological fluids. Practical utility was demonstrated on the determination of formate in blood serum samples from acute methanol intoxication patients and on the analysis of ionic profile (nitrosative stress markers, including nitrite and nitrate) in the exhaled breath condensate from one single exhalation.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Contactless conductivity detection; Exhaled breath condensate; Low sample volume; Methanol intoxication; Portable capillary electrophoresis

Mesh:

Substances:

Year:  2015        PMID: 26709071     DOI: 10.1016/j.chroma.2015.11.088

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  3 in total

1.  Capillary Electrophoresis Coupled to Electrospray Ionization Tandem Mass Spectrometry for Ultra-Sensitive Proteomic Analysis of Limited Samples.

Authors:  Kendall R Johnson; Michal Greguš; James C Kostas; Alexander R Ivanov
Journal:  Anal Chem       Date:  2022-01-04       Impact factor: 6.986

Review 2.  Low-cost and open-source strategies for chemical separations.

Authors:  Joshua J Davis; Samuel W Foster; James P Grinias
Journal:  J Chromatogr A       Date:  2020-12-24       Impact factor: 4.759

3.  [Determination of two quaternary ammonium salts in disinfector by portable capillary electrophoresis device based on smartphone].

Authors:  Yuanyu Wang; Ruihua Zhang; Qiang Zhang; Chengxi Cao; Liuyin Fan; Weiwen Liu
Journal:  Se Pu       Date:  2021-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.