| Literature DB >> 26703509 |
Karen Grove1, Garry Condon1, Barbara C Erny1, David F Chang1, Terry Kim2.
Abstract
UNLABELLED: We describe a new complication associated with capsular tension ring (CTR) implantation after placement of capsule retractors. We report 3 cases from 3 different surgeons of inadvertent threading of a CTR through a capsule retractor loop. In each case, the distal loop was opened or amputated to facilitate hook removal. We report this case series to alert cataract surgeons to a potential complication of using CTRs and capsule retractors together and to offer potential strategies for preventing and managing an intracapsular entanglement. FINANCIAL DISCLOSURE: Dr. Kim is a consultant to Alcon Surgical, Inc. Dr. Chang receives personal fees from Abbott Medical Optics, Inc. Dr. Condon receives personal fees from Alcon Surgical, Inc. and Microsurgical Technology. No other author has a financial or proprietary interest in any material or method mentioned.Entities:
Mesh:
Year: 2015 PMID: 26703509 DOI: 10.1016/j.jcrs.2015.10.012
Source DB: PubMed Journal: J Cataract Refract Surg ISSN: 0886-3350 Impact factor: 3.351