Literature DB >> 26703162

Amplitude and phase estimator combined with the Wiener postfilter for medical ultrasound imaging.

Ali Mohades Deylami1, Babak Mohammadzadeh Asl2.   

Abstract

BACKGROUND: The adaptive amplitude and phase estimator (APES) has been introduced in medical ultrasound imaging to calculate the amplitude of the desired signal more robustly than other adaptive beamformers like minimum variance (MV). This beamformer minimizes the optimization problem of MV by replacing the estimated array covariance matrix by the interferences plus noise covariance matrix. On the other hand, the Wiener postfilter as a post-weighting factor, which will be multiplied to the final weight vector of the beamformer, estimates the power of the desired signal and the power of the interferences plus noise to improve the contrast.
METHOD: The proposed method is a combination of the APES beamformer with the Wiener postfilter which uses the capabilities of the APES beamformer for accurate estimation of the amplitude of the desired signal and the Wiener postfilter in suppressing sidelobes. Specifically, we used the interferences plus the noise covariance matrix estimated in the APES beamformer to obtain an APES-based Wiener postfilter and obtained the APES + Wiener weight vector by multiplying the APES-based Wiener postfilter to the standard APES weight vector.
RESULTS: To evaluate the proposed APES + Wiener beamformer, we tested the proposed method on simulated and experimental datasets. The results of a simulated wire phantom demonstrate that the proposed beamformer can resolve two point scatterers better than the standard APES beamformer, even if the points are placed near each other. Simulating a cyst phantom shows that the APES + Wiener beamformer improves the contrast of the resulting images by about 4.5 dB by estimating the interior of the cyst better than the standard APES.
CONCLUSION: The evaluation of the proposed beamformer on an experimental dataset confirms the results of simulations, in which the proposed beamformer improves the resolution and contrast in comparison with the standard APES beamformer.

Entities:  

Keywords:  Adaptive beamformer; Medical ultrasound imaging; Minimum variance; Phase and amplitude estimator; Wiener postfilter

Mesh:

Year:  2015        PMID: 26703162     DOI: 10.1007/s10396-015-0671-z

Source DB:  PubMed          Journal:  J Med Ultrason (2001)        ISSN: 1346-4523            Impact factor:   1.314


  9 in total

1.  Adaptive imaging using the generalized coherence factor.

Authors:  Pai-Chi Li; Meng-Lin Li
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2003-02       Impact factor: 2.725

2.  High range resolution ultrasonographic vascular imaging using frequency domain interferometry with the Capon method.

Authors:  Hirofumi Taki; Kousuke Taki; Takuya Sakamoto; Makoto Yamakawa; Tsuyoshi Shiina; Motoi Kudo; Toru Sato
Journal:  IEEE Trans Med Imaging       Date:  2011-10-06       Impact factor: 10.048

3.  Wiener beamforming and the coherence factor in ultrasound imaging.

Authors:  Carl-Inge Nilsen; Sverre Holm
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2010-06       Impact factor: 2.725

4.  Eigenspace-based minimum variance beamforming applied to medical ultrasound imaging.

Authors:  Babak Mohammadzadeh Asl; Ali Mahloojifar
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2010-11       Impact factor: 2.725

5.  Adaptive beamforming applied to medical ultrasound imaging.

Authors:  Johan-Fredrik Synnevåg; Andreas Austeng; Sverre Holm
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2007-08       Impact factor: 2.725

6.  Benefits of minimum-variance beamforming in medical ultrasound imaging.

Authors:  Johan-Fredrik Synnevag; Andreas Austeng; Sverre Holm
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2009-09       Impact factor: 2.725

7.  Minimum variance beamforming combined with adaptive coherence weighting applied to medical ultrasound imaging.

Authors:  Babak Mohammadzadeh Asl; Ali Mahloojifar
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2009-09       Impact factor: 2.725

8.  Contrast enhancement and robustness improvement of adaptive ultrasound imaging using forward-backward minimum variance beamforming.

Authors:  Babak Mohammadzadeh Asl; Ali Mahloojifar
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2011-04       Impact factor: 2.725

9.  Apodization and windowing functions.

Authors:  Kevin J Parker
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2013-06       Impact factor: 2.725

  9 in total
  2 in total

1.  Twofold minimum variance beamforming for enhanced ultrasound imaging.

Authors:  Sayed Mahmoud Sakhaei; Seyede Elham Shamsian
Journal:  J Med Ultrason (2001)       Date:  2017-04-21       Impact factor: 1.314

2.  Generalized sidelobe canceler beamforming combined with Eigenspace-Wiener postfilter for medical ultrasound imaging.

Authors:  Jin Yang; Xiaodong Chen; Huaiyu Cai; Yi Wang
Journal:  Technol Health Care       Date:  2022       Impact factor: 1.205

  2 in total