Literature DB >> 26701690

Corneal powers measured with a rotating Scheimpflug camera.

Kristian Næser1, Giacomo Savini2, Jesper Flethøj Bregnhøj1.   

Abstract

BACKGROUND/AIM: Keratometry measures the anterior corneal curvature only. Corneal power is calculated by multiplication with the keratometric refractive index, which takes into account the average negative posterior corneal power. The aim of this study was to calculate and compare various expressions for total corneal power assessed with Scheimpflug camera techniques, which also measure the posterior corneal curvature.
METHODS: We used the Pentacam rotating Scheimpflug camera to measure the equivalent power, total corneal refractive power (based on Snell's law ray tracing), and simulated keratometry (keratometric refractive index=1.3375) over the central 3.0 mm zone in 951 eyes. The keratometric refractive index of the equivalent power and the total corneal refractive power was calculated as the ratio between these values and the anterior corneal curvature.
RESULTS: The equivalent power, total corneal refractive power, and simulated keratometry all differed statistically significantly (analysis of variance, p<0.001) and averaged 42.26 (±1.46) dioptres (D), 42.78 (±1.51) D and 43.42 (±1.49) D. The calculated keratometric refractive indices for equivalent and total corneal refractive power averaged 1.3284 (±0.0009) and 1.3324 (±0.0015). The error of using these calculated keratometric refractive indices rather than the measured values for equivalent and total corneal refractive power averaged 0 (±0.11 D) and -0.01 D (±0.19).
CONCLUSIONS: Pentacam rotating camera assessment of total corneal power over the central 3.0 mm zone differed significantly for simulated keratometry, equivalent power and Snell's law ray tracing. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/

Keywords:  Cornea; Optics and Refraction; Vision

Mesh:

Year:  2015        PMID: 26701690     DOI: 10.1136/bjophthalmol-2015-307474

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  8 in total

1.  Corneal power measurement with a new aberrometer/corneal topographer in eyes after small incision lenticule extraction for myopia.

Authors:  Chao Pan; Weina Tan; Yanjun Hua; Xiaohua Lei
Journal:  Int Ophthalmol       Date:  2019-05-27       Impact factor: 2.031

2.  Accuracy of optimized Sirius ray-tracing method in intraocular lens power calculation.

Authors:  Li-Qing Wei; Ying-Hui Fu; Wei-Hua Pan; Li Nie; Yu Chen; Gui-Fang Liu; Zhen-Bin Qian
Journal:  Int J Ophthalmol       Date:  2022-02-18       Impact factor: 1.779

3.  Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging.

Authors:  Giacomo Savini; Leonardo Taroni; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  Eye (Lond)       Date:  2020-11-02       Impact factor: 4.456

4.  Comprehensive evaluation of total corneal refractive power by ray tracing in predicting corneal power in eyes after small incision lenticule extraction.

Authors:  Chao Pan; Weina Tan; Yanjun Hua; Xiaohua Lei
Journal:  PLoS One       Date:  2019-06-06       Impact factor: 3.240

5.  Corneal power changes with Scheimpflug rotating camera after hyperopic LASIK.

Authors:  Woong-Joo Whang; Young-Sik Yoo; Choun-Ki Joo
Journal:  Medicine (Baltimore)       Date:  2018-12       Impact factor: 1.817

6.  Analysis of total corneal astigmatism with a rotating Scheimpflug camera in keratoconus.

Authors:  Jinho Kim; Woong-Joo Whang; Hyun-Seung Kim
Journal:  BMC Ophthalmol       Date:  2020-12-03       Impact factor: 2.209

7.  A Multicenter Study of the Distribution Pattern of Posterior-To-Anterior Corneal Curvature Radii Ratio in Chinese Myopic Patients.

Authors:  Changting Tang; Qiaowei Wu; Baoyi Liu; Guanrong Wu; Jing Fan; Yijun Hu; Honghua Yu
Journal:  Front Med (Lausanne)       Date:  2021-12-20

8.  Scheimpflug analysis of corneal power changes after hyperopic small incision lenticule extraction.

Authors:  Dan Fu; Jianmin Shang; Xiaoyu Zhang; Lingling Niu; Tian Han; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2021-07-20       Impact factor: 2.209

  8 in total