Literature DB >> 26687376

In vitro experiments of vessel wall apposition between the Enterprise and Enterprise 2 stents for treatment of cerebral aneurysms.

Kenichi Kono1, Tomoaki Terada2.   

Abstract

BACKGROUND: A closed-cell stent called Enterprise has been used for stent-assisted coil embolization of cerebral aneurysms. The Enterprise stent tends to cause kinks and vessel wall malposition in curved vessels and may cause thromboembolic complications. We evaluated vessel wall apposition of a new closed-cell stent, Enterprise 2, compared with a previous Enterprise stent, using curved vascular silicone models.
METHODS: The Enterprise or Enterprise 2 stent was deployed in curved vascular models with various radii of approximately 5 to 10 mm. Stent deployment was performed 25 times in each stent. A push-pull technique was used to minimize incomplete wall apposition. To evaluate conformity of stents, gaps between a stent and a vessel wall were measured.
RESULTS: The gap ratio (gap / a wall diameter) was 15 % ± 17 % (mean ± standard deviation) and 41 % ± 15 % with the Enterprise 2 stent and the Enterprise stent, respectively. Taking gap ratios and radii of vessel curvature into consideration, the Enterprise 2 stent had significantly better wall apposition than the Enterprise stent (p = 0.005). In the same radius of vessel curvature, the Enterprise 2 stent had approximately half of the gap compared with the Enterprise stent. There were no significant differences in vessel straightening effects between the two stents.
CONCLUSIONS: The Enterprise 2 stent has better wall apposition in curved vessels than the Enterprise stent. The gap between a vessel wall and the Enterprise 2 stent is approximately half that of the Enterprise stent. However, gaps and kinks are still present in curved vessels with a small radius. Caution should be taken for kinks and malposition in acutely curved vessels, such as the siphon of the internal carotid artery.

Entities:  

Keywords:  Closed-cell stents; Enterprise 2 stent; Silicone model; Stent-assisted coil embolization; Wall apposition

Mesh:

Substances:

Year:  2015        PMID: 26687376     DOI: 10.1007/s00701-015-2666-3

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  3 in total

1.  First Experiences with the New Enterprise2® Stent.

Authors:  C Herweh; S Nagel; J Pfaff; C Ulfert; M Wolf; M Bendszus; M Möhlenbruch
Journal:  Clin Neuroradiol       Date:  2016-09-27       Impact factor: 3.649

2.  The Enterprise2 Stent for Endovascular Treatment of Intracranial Aneurysms: Short-Term Results From a Single Center Experience.

Authors:  Linhui Chen; Chaobo Zheng; Jiong Wu; Jie Gong; Yuhai Gao; Shu Wan
Journal:  Front Neurol       Date:  2020-12-03       Impact factor: 4.003

3.  Effects of Enterprise 2 stent-assisted coil embolization for wide-necked intracranial aneurysms.

Authors:  Yangyang Zhou; Qichen Peng; Shiqing Mu
Journal:  Chin Neurosurg J       Date:  2022-10-02