Maren Dietrich1, Nicole Finocchiaro1, Sebastian Olszweski1, Jutta Arens2, Thomas Schmitz-Rode1,2, Joerg Sachweh1, Stefan Jockenhoevel1, Christian G Cornelissen1,3.
Abstract
OBJECTIVE: This study focusses on the development of a biomimetic oxygenator test device. Due to limited biocompatibility, current oxygenators do not allow mid- to long-term therapy. Tissue engineering uses autologous cell sources to overcome the immunogenic barriers of biomaterials. Surface coating with endothelial cells might improve hemocompatibility and thus prevent immunogenic reactions of the body. In this study this concept is applied to endothelialise a gas-permeable membrane to develop a biomimetic oxygenator test-device (ENDOXY).Entities:
Mesh:
Year: 2015 PMID: 26682907 PMCID: PMC4684320 DOI: 10.1371/journal.pone.0142961
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1The ENDOXY system set up as a bioreactor for application of defined shear stress on endothelial cells.
1 –gas chamber, 2 –medium / blood chamber, 3 –Lumox™ slide, 4 –microscope.
Fig 2Schematic drawing of the ENDOXY system set up in bioreactor mode (A) and oxygenator mode (B).
Flow rates generated in the ENDOXY during dynamic cell culture.
| Day of dynamic culture | Flow rate (ml/min) | Shear stress (dyn/cm²) |
|---|---|---|
| 1 | 20 | 2.5 |
| 2 | 30 | 3.7 |
| 3 | 40 | 4.9 |
Fig 3Comparative light-microscopic picture of HUVECs cultured on Lumox™ slides for 7 days.
A –Static control. B—Shear stress stimulated cells. Scale bar = 200 μm.
Fig 4On-line microscopy of endothelial cells cultured in the ENDOXY system during the first 36 hours.
Scale bar = 100 μm.
Fig 5Fluorescence staining against CD31 and vWF for dynamically and statically cultured endothelial cells incubated with either cell culture medium or blood.
Static incubation incubation with blood was done for 90 minutes whereas blood was perfused for 180 minutes. All samples stain positive for CD31 and vWF, proving endothelial lineage. Blood incubation does not alter marker expression. Scale bar = 100 μm.
Gas transfer measurement of uncoated Lumox™ slide—blood gas analysis.
| Flow rate | Pre / Post ENDOXY | pO2 (mmHg) | sO2 (%) | pCO2 (mmHg) | Hemoglobin (g/dl) | |
|---|---|---|---|---|---|---|
| 10 ml / min | Pre | 54.6 ± 0.8 | 68.6 ± 0.7 | p = 0.03 | 53.2 ± 1.0 | 10.8 ± 0.5 |
| Post | 57.5 ± 1.8 | 72.8 ± 2.0 | 52.2 ± 1.3 | 10.6 ± 0.2 | ||
| 7.5 ml / min | Pre | 68.6 ± 26.2 | 66.1 ± 1.0 | n.s. | 33.1 ± 16.0 | 14.2 ± 0.1 |
| Post | 53.2 ± 0.6 | 67.9 ± 1.2 | 51.6 ± 0.6 | 13.9 ± 1.0 | ||
| 5 ml / min | Pre | 54.6 ± 0.2 | 68.1 ± 0.3 | n.s. | 50.1 ± 1.0 | 13.5 ± 0.8 |
| Post | 57.5 ± 1.7 | 70.6 ± 1.6 | 33.0 ± 14.9 | 12.6 ± 1.0 | ||
| 2.5 ml / min | Pre | 53.2 ± 0.2 | 68.2 ± 0.3 | p = 0.02 | 48.0 ± 0.5 | 18.1 ± 0.4 |
| Post | 54.2 ± 1.2 | 70.5 ± 1.0 | 47.7 ± 0.3 | 17.8 ± 0.4 | ||
Blood gas analysis on blood samples acquired from the gas transfer test loop at the ENDOXY’s inlet (Pre) and outlet (Post). Based on results of the T-Test on oxygen saturation values, gas transfer calculation was performed for flow rates of 10 ml / min and 2.5 ml / min. n.s. = not significant.
Gas transfer measurement of cell coated Lumox™ slide—blood gas analysis.
| Flow rate | Pre / Post ENDOXY | pO2 (mmHg) | sO2 (%) | pCO2 (mmHg) | Hemoglobin (g/dl) | |
|---|---|---|---|---|---|---|
| 10 ml / min | Pre | 41.0 ± 0.3 | 58.3 ± 1.4 | n.s. | 52.4 ± 3.8 | 12.8 ± 1.0 |
| Post | 41.2 ± 0.2 | 59.9 ± 0.3 | 47.8 ± 0.1 | 11.7 ± 0.1 | ||
| 5 ml / min | Pre | 40.8 ± 0.5 | 57.9 ± 0.3 | n.s. | 50.4 ± 4.6 | 12.5 ± 0.9 |
| Post | 40.8 ± 1.0 | 57.8 ± 0.8 | 51.4 ± 2.3 | 12.7 ± 1.4 | ||
| 2.5 ml / min | Pre | 40.8 ± 0.6 | 58.4 ± 0.6 | n.s. | 50.9 ± 0.3 | 12.1 ± 0.3 |
| Post | 41.2 ± 0.9 | 59.8 ± 1.4 | 50.3 ± 0.3 | 12.1 ± 0.1 | ||
Blood gas analysis on blood samples acquired from the gas transfer test loop at the ENDOXY’s inlet (Pre) and outlet (Post). The difference of oxygen saturation at the inlet and outlet is statistically not significant, prohibiting further analysis. n.s. = not significant.
Gas transfer of uncoated Lumox™ slide.
| Flow rate | Measured gas transfer | Maximum of gas transfer |
|---|---|---|
| 10 ml / min | 8.50 × 10−5 ± 6.85 × 10−5 | 14.11 x 10−5 |
| 2.5 ml / min | 1.21 × 10−5 ± 2.04 x 10−5 | 14.11 x 10−5 |
Calculated gas transfer through uncoated Lumox™ slides. Results show high standard deviations. Still, the measured gas transfer at 10 ml / min blood flow demonstrates an efficient system setup as the measured transfer is near the theoretical maximum.
Gas transfer measurement of uncoated Lumox™ slide after redesign—blood gas analysis.
| Flow rate | Pre / Post ENDOXY | pO2 (mmHg) | sO2 (%) | pCO2 (mmHg) | Hemoglobin (g/dl) | |
|---|---|---|---|---|---|---|
| 3 ml / min | Pre | 40.7 ± 0.7 | 69.5 ± 1.2 | p < 0.05 | 46.9 ± 0.7 | 11.2 ± 0.2 |
| Post | 59.7 ± 5.1 | 87.7 ± 1.2 | 47.1 ± 0.6 | 13.9 ± 0.3 | ||
| 2 ml / min | Pre | 42.1 ± 0.3 | 65.1 ± 0.7 | p < 0.05 | 49.7 ± 0.2 | 10.7 ± 0.1 |
| Post | 53.9 ± 3.7 | 85.4 ± 2.5 | 49.2 ± 0.6 | 13.9 ± 0.4 | ||
| 1 ml / min | Pre | 41.9 ± 0.9 | 62.1 ± 2.4 | p < 0.05 | 50.3 ± 0.8 | 10.0 ± 0.4 |
| Post | 53.1 ± 1.3 | 91.4 ± 2.4 | 50.4 ± 0.5 | 11.8 ± 0.4 | ||
Blood gas analysis on blood samples acquired from the gas transfer test loop at the ENDOXY’s inlet (Pre) and outlet (Post). Based on results of the T-Test on oxygen saturation values, gas transfer calculation was performed for all flow rates.
Gas transfer of uncoated Lumox™ slide after redesign.
| Flow rate | Measured gas transfer | Maximum of gas transfer |
|---|---|---|
| 3 ml / min | 7.27 x 10−5 ± 2.4 x 10−5 | 14.11 × 10−5 |
| 2 ml / min | 9.11 x 10−5 ± 1.2 x 10−5 | 14.11 × 10−5 |
| 1 ml / min | 6.28 x 10−5 ± 5.6 x 10−5 | 14.11 × 10−5 |
Calculated gas transfer through uncoated Lumox™ slides, measured in the redesigned ENDOXY system.
Gas transfer measurement of coated Lumox™ slide after redesign—blood gas analysis.
| Flow rate | Pre / Post ENDOXY | pO2 (mmHg) | sO2 (%) | pCO2 (mmHg) | Hemoglobin (g/dl) | |
|---|---|---|---|---|---|---|
| 3 ml / min | Pre | 37.0 ± 0.7 | 52.4 ± 0.9 | p < 0.05 | 50.7 ± 1.8 | 11.9 ± 0.7 |
| Post | 39.7 ± 1.1 | 58.7 ± 2.4 | 48.0 ± 2.6 | 10.9 ± 1.0 | ||
| 2 ml / min | Pre | 35.1 ± 0.2 | 46.5 ± 0.6 | p < 0.05 | 55.8 ± 0.4 | 11.8 ± 0.1 |
| Post | 41.6 ± 1.4 | 60.6 ± 3.4 | 55.9 ± 0.9 | 12.8 ± 0.1 | ||
| 1 ml / min | Pre | 35.6 ± 0.1 | 49.5 ± 0.6 | p < 0.05 | 52.5 ± 0.7 | 11.4 ± 0.2 |
| Post | 50.6 ± 2.2 | 74.7 ± 2.1 | 52.3 ± 0.5 | 13.0 ± 0.4 | ||
Blood gas analysis on blood samples acquired from the gas transfer test loop at the ENDOXY’s inlet (Pre) and outlet (Post). Based on results of the T-Test on oxygen saturation values, gas transfer calculation was performed for all flow rates.
Gas transfer of coated Lumox™ slide after redesign.
| Flow rate | Measured gas transfer | Maximum of gas transfer |
|---|---|---|
| 3 ml / min | 9.54 × 10−6 ± 4.08 x 10−5 | 14.11 x 10−5 |
| 2 ml / min | 8.46 × 10−5 ± 1.62 x 10−5 | 14.11 x 10−5 |
| 1 ml / min | 7.64 × 10−5 ± 8.18 x 10−6 | 14.11 x 10−5 |
Calculated gas transfer through coated Lumox™ slides, measured in the redesigned ENDOXY system. Results show a high standard deviation for a blood flow of 3 ml / min. At a blood flow of 1 ml / min and 2 ml / min, gas transfer through the cell-coated membrane is detectable.