Literature DB >> 26680758

Discontinuation rates and acceptability during 1 year of using the intrauterine ball (the SCu380A).

Ellen Wiebe1, James Trussell2.   

Abstract

OBJECTIVE: To characterize method-related discontinuation rates and acceptability over 1 year of SCu380A use. STUDY
DESIGN: Women enrolled into this prospective pilot study underwent SCu380A placement by a single clinician with follow-up at 6-8 weeks and at 3, 6, 9 and 12 months. Outcome measures included reasons for discontinuation and satisfaction.
RESULTS: Fifty-one women had a 12-mm-diameter intrauterine ball (IUB; SCu380A) inserted by a single clinician (EW). Postinsertion ultrasonography showed all the devices to be correctly placed. By 8 weeks, there were 9 expulsions and 2 removals for symptoms. By 12 months, there were 14 expulsions (27%), 8 removals for symptoms (16%), 1 pregnancy and 7 lost to follow-up. Of the remaining 21 women, 15 said that they were satisfied.
CONCLUSION: The high rate of expulsions and removals for symptoms in the first year indicates that this device, as presently designed, is unacceptable as an alternative to the currently available copper intrauterine devices. IMPLICATIONS: More research is needed before we know if it is the spherical design, the size of the device or some other factor that led to the high expulsion rate. There are currently three different IUBs (12, 15 and 18 mm) approved and being investigated.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Acceptability; Discontinuation; Expulsion; Intrauterine device

Mesh:

Year:  2015        PMID: 26680758      PMCID: PMC5515364          DOI: 10.1016/j.contraception.2015.12.002

Source DB:  PubMed          Journal:  Contraception        ISSN: 0010-7824            Impact factor:   3.375


  4 in total

Review 1.  Intrauterine devices and intrauterine systems.

Authors: 
Journal:  Hum Reprod Update       Date:  2008-04-09       Impact factor: 15.610

2.  Complications of the intrauterine device in nulliparous and parous women.

Authors:  H M Veldhuis; A G Vos; A L M Lagro-Janssen
Journal:  Eur J Gen Pract       Date:  2004-09       Impact factor: 1.904

3.  Effects of age, parity, and device type on complications and discontinuation of intrauterine devices.

Authors:  Joelle Aoun; Virginia A Dines; Dale W Stovall; Mihriye Mete; Casey B Nelson; Veronica Gomez-Lobo
Journal:  Obstet Gynecol       Date:  2014-03       Impact factor: 7.661

4.  The IUB, a newly invented IUD: a brief report.

Authors:  Ilan Baram; Ariel Weinstein; James Trussell
Journal:  Contraception       Date:  2013-11-07       Impact factor: 3.375

  4 in total
  3 in total

Review 1.  New developments in long-acting reversible contraception: the promise of intrauterine devices and implants to improve family planning services.

Authors:  David K Turok; Lori M Gawron; Samantha Lawson
Journal:  Fertil Steril       Date:  2016-10-04       Impact factor: 7.329

2.  Discontinuation rates of intrauterine contraception due to unfavourable bleeding: a systematic review.

Authors:  Dustin Costescu; Rajinder Chawla; Rowena Hughes; Stephanie Teal; Martin Merz
Journal:  BMC Womens Health       Date:  2022-03-21       Impact factor: 2.809

3.  Dimensional analysis of the endometrial cavity: how many dimensions should the ideal intrauterine device or system have?

Authors:  Norman D Goldstuck
Journal:  Int J Womens Health       Date:  2018-04-09
  3 in total