Abstract
In the present study a simple, accurate, precise, economical and specific UV-spectrophotometric method for estimation of besifloxacin in bulk and in different pharmaceutical formulation has been developed. The drug shows maximum λmax289 nm in distilled water, simulated tears and phosphate buffer saline. The linearity range of developed methods were in the range of 3-30 μg/ml of drug with a correlation coefficient (r(2)) 0.9992, 0.9989 and 0.9984 with respect to distilled water, simulated tears and phosphate buffer saline, respectively. Reproducibility by repeating methods as %RSD were found to be less than 2%. The limit of detection in different media was found to be 0.62, 0.72 and 0.88 μg/ml, respectively. The limit of quantification was found to be 1.88, 2.10, 2.60 μg/ml, respectively. The proposed method was validated statically according to International Conference on Harmonization guidelines with respect to specificity, linearity, range, accuracy, precision and robustness. The proposed methods of validation were found to be accurate and highly specific for the estimation of besifloxacin in different pharmaceutical formulations.Entities:
Keywords: Besifloxacin; ICH guidelines; Simulated tears; UV-Spectrophotometric method; Validation
Year: 2015 PMID: 26664055 PMCID: PMC4649791 DOI: 10.4103/0250-474x.164784
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1Chemical structure of besifloxacin.
Fig. 2Ultraviolet spectrum in different media. Scanned stacked spectrum graph of BSF in (a) distilled water (b) simulated tears and in (c) phosphate buffer saline.
Fig. 3Overlay ultraviolet spectrum in different media.
Overlay spectrum graph report of BSF in distilled water, in simulated tears and in phosphate buffer saline.
Fig. 4Calibration curve of besifloxacin in different media.
Calibration curve of besifloxacin in distilled water , in simulated tears and in phosphate buffer saline
REGRESSION, ANALYSIS AND SYSTEM SUITABLITY PARAMETERS OF BSF FROM UV-SPECTROPHOTOMETRIC METHOD
ACCURACY
RANGE
PRECISION REPEATABILITY
INTERMEDIATE PRECISION
ROBUSTNESS