| Literature DB >> 26644423 |
Joshua S Agranat1, Nicole R Kitos2, Deborah S Jacobs3.
Abstract
BACKGROUND/AIMS: To determine the impact of prosthetic replacement of the ocular surface ecosystem (PROSE) treatment at 5 years.Entities:
Keywords: Contact lens; Cornea; Ocular surface; Prosthesis; Vision
Mesh:
Year: 2015 PMID: 26644423 PMCID: PMC5013114 DOI: 10.1136/bjophthalmol-2015-307483
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638
Figure 1Candidacy for prosthetic replacement of the ocular surface ecosystem (PROSE) treatment, device dispensed status and 5-year wearing status.
Reasons for discontinuation of wear
| Reason | Number of patients | Distorted surface:OSD | Mean duration of wear (years) |
|---|---|---|---|
| Discomfort | 7 | 4:3 | 0.6 |
| Limited improvement of VA | 5 | 1:4 | 2.2 |
| Insertion/removal too difficult | 4 | 1:3 | 0.8 |
| Cured | 4 | 2:2 | 2.3 |
| Ophthalmic complications | 3 | 2:1 | 0.7 |
| ‘Stopped working’ | 3 | 0:3 | 0.7 |
| Lost/broken | 2 | 1:1 | 4.9 |
| Disease progression | 2 | 0:2 | 0.8 |
| Cosmetic issues | 1 | 0:1 | 0.2 |
| Systemic disease | 1 | 0:1 | 3.9 |
Median duration of wear before discontinuation was 1.0 year. There is no relationship between reasons stopped wearing and primary diagnosis (r=−0.02145, p=0.9072).
OSD, ocular surface disease; VA, visual acuity.
Figure 2Graph showing Visual Function Questionnaire (VFQ-25) changes in regard to wearing status. Those wearing had significant improvement in visual function at 6 months (Δ=+23 points, p<0.001, mean=82, n=87) with no significant decline at 5 years (Δ=−4 points, p=0.22, mean=78, n=89). Those no longer wearing at 5 years had improvement in visual function at 6 months (Δ=+14 points, p=0.08, mean=72, n=32) with no significant decline at 5 years as well (Δ=−2 points, p=0.65, mean=70, n=32). However, the magnitude of the change for the no longer wearing group was not as great. Further, there remained significant difference (p=0.029) in visual function (8 points) between those wearing at 5 years and those no longer wearing.
VFQ-25 scores delineated by wearing status and primary disease category: distorted corneal surface and ocular surface disease (OSD)
| Distorted corneal surface | OSD | Total | ||||
|---|---|---|---|---|---|---|
| Wearing at 5 years? | Yes | No | Yes | No | Yes | No |
| N=48 | N=9 | N=41 | N=23 | N=89 | N=32 | |
| Baseline VFQ-25 score | 62.40 | 64.44 | 54.90 | 55.83 | 58.94 | 58.25 |
| 6-month VFQ-25 score | 86.00 | 68.67 | 82.07 | 73.67 | 84.42 | 72.24 |
| 5-year VFQ-25 score | 82.64 | 76.91 | 73.70 | 67.03 | 78.46 | 69.75 |
| 5-year change VFQ-25 score | +20.24 | +12.47 | +18.80 | +11.20 | +19.52 | +11.5 |
| Baseline GHS score | 68.75 | 61.11 | 57.32 | 59.78 | 63.48 | 60.15 |
| 6-month GHS score | 73.08 | 66.67 | 68.52 | 53.33 | 71.24 | 57.14 |
| 5-year GHS score | 68.90 | 50.00 | 60.42 | 52.38 | 64.94 | 51.72 |
| 5-year change GHS score | +0.15 | −11.11 | +3.10 | −7.40 | +1.46 | −8.43 |
VFQ-25, NEI VFQ-25 composite score, Visual Function Questionnaire's main categories; GHS, NEI VFQ self-reported general health status.