Novel technique to measure horizontal corneal diameter using a Scheimpflug image on the Pentacam.

Youngje Sung1, Sokjoong Chung1, Sang Min Nam2.   

Abstract

PURPOSE: A new method to measure the horizontal corneal diameter using a Scheimpflug image (SFI) was introduced and compared with the automated iris camera photo (ICP) method on the Pentacam.
METHODS: Each one eye of 88 subjects without significant abnormality in the cornea was included. In SFI, diameters were measured manually using the built-in software tool with the edge-filter option. ICP value read from the overview display. Repeatability was determined from the first two measurements of each eye. The accuracy was assessed using Bland-Altman plot in comparison with caliper measurements.
RESULTS: The measurement failure rate of ICP was aggravated by pupil dilation, while that of SFI was unrelated to pupil dilation. Repeatability coefficients of ICP and SFI were 0.1mm and 0.12mm, respectively. The mean difference between ICP values and caliper values was not significant. SFI values were greater than caliper measurements by a mean of 0.13mm.
CONCLUSIONS: New SFI was reliable and could complement ICP in dilated pupil. The measurement failure rates of Pentacam need to be improved.
Copyright © 2015 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Horizontal white-to-white corneal diameter; Iris camera; Reliability; Scheimpflug image

Mesh:

Year:  2015        PMID: 26628338     DOI: 10.1016/j.clae.2015.11.003

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  2 in total

1.  Short-time evaluation on intraocular scattering after implantable collamer lens implantation for correcting high myopia.

Authors:  Zhe Yu; Jun Li; Hui Song
Journal:  BMC Ophthalmol       Date:  2020-06-17       Impact factor: 2.209

2.  Associations between anterior segment biometry and high axial myopia in 3438 cataractous eyes in the Chinese population.

Authors:  Ao Miao; Yating Tang; Xiangjia Zhu; Dongjin Qian; Tianyu Zheng; Yi Lu
Journal:  BMC Ophthalmol       Date:  2022-02-12       Impact factor: 2.209

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.