| Literature DB >> 26618177 |
Ryosuke Mihara1, Nobuyasu Komasawa1, Sayuri Matsunami1, Toshiaki Minami1.
Abstract
BACKGROUND: Videolaryngoscopes may not be useful in the presence of hematemesis or vomitus. We compared the utility of the Macintosh laryngoscope (McL), which is a direct laryngoscope, with that of the Pentax-AWS Airwayscope (AWS) and McGRATH MAC (McGRATH), which are videolaryngoscopes, in simulated hematemesis and vomitus settings.Entities:
Mesh:
Year: 2015 PMID: 26618177 PMCID: PMC4651703 DOI: 10.1155/2015/806243
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Simulated hematemesis and vomitus settings. (a) Simulated hematemesis and vomitus. (b) A representative laryngoscopic view with three laryngoscopes in each simulation condition (normal, hematemesis, and vomitus).
Tracheal intubation success rates for McL, AWS, and McGRATH in normal, hematemesis, and vomitus settings.
| Normal simulation | Hematemesis simulation | Vomitus simulation |
| |
|---|---|---|---|---|
| McL | 17/17 | 15/17 | 17/17 | 0.125 |
| AWS | 17/17 | 10/17 | 12/17 | <0.001 |
| McGRATH | 17/17 | 16/17 | 17/17 | 0.361 |
|
| 0.978 | 0.021 | 0.003 |
AWS, Pentax-AWS Airwayscope; McGRATH, McGRATH MAC; McL, Macintosh laryngoscope.
Numerator: number of participants who were successfully intubated.
Denominator: number of participants for whom tracheal intubation was attempted.
Differences were analyzed using chi-square test.
Figure 2Box-and-whisker plot (median, IQR, and range) of time required for simulated tracheal intubation in hematemesis and vomitus settings using each laryngoscope. Results are expressed as means ± SDs and were analyzed using one-way analysis of variance. (a) Normal setting, (b) hematemesis, and (c) vomitus setting. AWS, Pentax-AWS Airwayscope; McGRATH, McGRATH MAC; McL, Macintosh laryngoscope.
Figure 3Box-and-whisker plot (median, IQR, and range) of visual analog scale scores for difficulty of simulated tracheal intubation in hematemesis and vomitus settings using each laryngoscope. (a) Normal setting, (b) hematemesis setting, and (c) vomitus setting. Results are expressed as means ± SDs and were analyzed using one-way analysis of variance. AWS, Pentax-AWS Airwayscope; McGRATH, McGRATH MAC; McL, Macintosh laryngoscope.