Automatic 3D reconstruction of electrophysiology catheters from two-view monoplane C-arm image sequences.

Christoph Baur1, Fausto Milletari2, Vasileios Belagiannis1, Nassir Navab1,3, Pascal Fallavollita1.   

Abstract

PURPOSE: Catheter guidance is a vital task for the success of electrophysiology interventions. It is usually provided through fluoroscopic images that are taken intra-operatively. The cardiologists, who are typically equipped with C-arm systems, scan the patient from multiple views rotating the fluoroscope around one of its axes. The resulting sequences allow the cardiologists to build a mental model of the 3D position of the catheters and interest points from the multiple views.
METHOD: We describe and compare different 3D catheter reconstruction strategies and ultimately propose a novel and robust method for the automatic reconstruction of 3D catheters in non-synchronized fluoroscopic sequences. This approach does not purely rely on triangulation but incorporates prior knowledge about the catheters. In conjunction with an automatic detection method, we demonstrate the performance of our method compared to ground truth annotations.
RESULTS: In our experiments that include 20 biplane datasets, we achieve an average reprojection error of 0.43 mm and an average reconstruction error of 0.67 mm compared to gold standard annotation.
CONCLUSIONS: In clinical practice, catheters suffer from complex motion due to the combined effect of heartbeat and respiratory motion. As a result, any 3D reconstruction algorithm via triangulation is imprecise. We have proposed a new method that is fully automatic and highly accurate to reconstruct catheters in three dimensions.

Entities:  

Keywords:  Arrhythmias; C-arm fluoroscopy; Catheter 3D reconstruction; Electrophysiology; Probabilistic graphical model

Mesh:

Year:  2015        PMID: 26615429     DOI: 10.1007/s11548-015-1325-8

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  11 in total

1.  Three-dimensional guide-wire reconstruction from biplane image sequences for integrated display in 3-D vasculature.

Authors:  Shirley A M Baert; Everine B van de Kraats; Theo van Walsum; Max A Viergever; Wiro J Niessen
Journal:  IEEE Trans Med Imaging       Date:  2003-10       Impact factor: 10.048

2.  Temporal estimation of the 3d guide-wire position using 2d X-ray images.

Authors:  Marcel Brückner; Frank Deinzer; Joachim Denzler
Journal:  Med Image Comput Comput Assist Interv       Date:  2009

3.  Real-time x-ray fluoroscopy-based catheter detection and tracking for cardiac electrophysiology interventions.

Authors:  YingLiang Ma; Nicolas Gogin; Pascal Cathier; R James Housden; Geert Gijsbers; Michael Cooklin; Mark O'Neill; Jaswinder Gill; C Aldo Rinaldi; Reza Razavi; Kawal S Rhode
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

4.  Accuracy assessment of catheter guidance technology in electrophysiology procedures: a comparison of a new 3D-based fluoroscopy navigation system to current electroanatomic mapping systems.

Authors:  Felix Bourier; Rebecca Fahrig; Paul Wang; Pasquale Santangeli; Klaus Kurzidim; Norbert Strobel; Teri Moore; Cameron Hinkel; Amin Al-Ahmad
Journal:  J Cardiovasc Electrophysiol       Date:  2013-09-19

5.  Fully automatic catheter localization in C-arm images using ł1-sparse coding.

Authors:  Fausto Milletari; Vasileios Belagiannis; Nassir Navab; Pascal Fallavollita
Journal:  Med Image Comput Comput Assist Interv       Date:  2014

6.  Semi-automatic catheter reconstruction from two views.

Authors:  Matthias Hoffmann; Alexander Brost; Carolin Jakob; Felix Bourier; Martin Koch; Klaus Kurzidim; Joachim Hornegger; Norbert Strobel
Journal:  Med Image Comput Comput Assist Interv       Date:  2012

7.  Toward multiple catheters detection in fluoroscopic image guided interventions.

Authors:  Liron Yatziv; Mathieu Chartouni; Saurabh Datta; Guillermo Sapiro
Journal:  IEEE Trans Inf Technol Biomed       Date:  2012-02-28

8.  Automatic detection of multiple and overlapping EP catheters in fluoroscopic sequences.

Authors:  Fausto Milletari; Nassir Navab; Pascal Fallavollita
Journal:  Med Image Comput Comput Assist Interv       Date:  2013

9.  Relationship between contact force sensing technology and medium-term outcome of atrial fibrillation ablation: a multicenter study of 600 patients.

Authors:  Julian W E Jarman; Sandeep Panikker; Moloy DAS; Gareth J Wynn; Waqas Ullah; Andrianos Kontogeorgis; Shouvik K Haldar; Preya J Patel; Wajid Hussain; Vias Markides; Dhiraj Gupta; Richard J Schilling; Tom Wong
Journal:  J Cardiovasc Electrophysiol       Date:  2015-02-11

10.  Is single-view fluoroscopy sufficient in guiding cardiac ablation procedures?

Authors:  Pascal Fallavollita
Journal:  Int J Biomed Imaging       Date:  2010-03-24
View more
  3 in total

1.  Real-time microrobot posture recognition via biplane X-ray imaging system for external electromagnetic actuation.

Authors:  Phu Bao Nguyen; Byungjeon Kang; D M Bappy; Eunpyo Choi; Sukho Park; Seong Young Ko; Jong-Oh Park; Chang-Sei Kim
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-08-20       Impact factor: 2.924

2.  Automatic quadriceps and patellae segmentation of MRI with cascaded U2 -Net and SASSNet deep learning model.

Authors:  Ruida Cheng; Marion Crouzier; François Hug; Kylie Tucker; Paul Juneau; Evan McCreedy; William Gandler; Matthew J McAuliffe; Frances T Sheehan
Journal:  Med Phys       Date:  2021-11-22       Impact factor: 4.506

3.  3D localization from 2D X-ray projection.

Authors:  Dagmar Bertsche; Volker Rasche; Wolfgang Rottbauer; Ina Vernikouskaya
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-07-11       Impact factor: 3.421

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.