Accuracy assessment of catheter guidance technology in electrophysiology procedures: a comparison of a new 3D-based fluoroscopy navigation system to current electroanatomic mapping systems.

Felix Bourier1, Rebecca Fahrig, Paul Wang, Pasquale Santangeli, Klaus Kurzidim, Norbert Strobel, Teri Moore, Cameron Hinkel, Amin Al-Ahmad.   

Abstract

BACKGROUND: With increasing complexity in electrophysiology (EP) procedures, the use of electroanatomic mapping systems (EAMS) as a supplement to fluoroscopy has become common practice. This is the first study that evaluates spatial and point localization accuracy for 2 current EAMS, CARTO3(®) (Biosense Webster, Diamond Bar, CA, USA) and EnSite Velocity(®) (St. Jude Medical Inc., St. Paul, MN, USA), and for a novel overlay guidance (OG) software (Siemens AG, Forchheim, Germany) in a phantom experiment. METHODS AND
RESULTS: A C-arm CT scan was performed on an acrylic phantom containing holes and location markers. Spatial accuracy was assessed for each system using distance measurements involving known markers inside the phantom and properly placed catheters. Anatomical maps of the phantom were acquired by each EAMS, whereas the 3D-based OG software superimposed an overlay image of the phantom, segmented from the C-arm CT data set, onto biplane fluoroscopy. Registration processes and landmark measurements quantitatively assessed the spatial accuracy of each technology with respect to the ground truth phantom. Point localization performance was 0.49 ± 0.25 mm in OG, 0.46 ± 0.17 mm in CARTO3(®) and 0.79 ± 0.83 mm in EnSite(®) . The registration offset between virtual visualization and reality was 1.10 ± 0.52 mm in OG, 1.62 ± 0.77 mm in CARTO3(®) and 2.02 ± 1.21 mm in EnSite(®) . The offset to phantom C-arm CT landmark measurements was 0.30 ± 0.26 mm in OG, 0.24 ± 0.21 mm in CARTO3(®) and 1.32 ± 0.98 mm in EnSite(®) .
CONCLUSIONS: Each of the evaluated EP guidance systems showed a high level of accuracy; the observed offsets between the virtual 3D visualization and the real phantom were below a clinically relevant threshold of 3 mm.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  C-arm CT; CARTO; EnSite; atrial fibrillation; catheter ablation; image integration; registration

Mesh:

Year:  2013        PMID: 24102965     DOI: 10.1111/jce.12264

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  9 in total

1.  Automatic 3D reconstruction of electrophysiology catheters from two-view monoplane C-arm image sequences.

Authors:  Christoph Baur; Fausto Milletari; Vasileios Belagiannis; Nassir Navab; Pascal Fallavollita
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-11-28       Impact factor: 2.924

Review 2.  Emerging risk biomarkers in cardiovascular diseases and disorders.

Authors:  Ravi Kant Upadhyay
Journal:  J Lipids       Date:  2015-04-08

Review 3.  Mapping of atrial fibrillation: strategies to understand an enigmatic arrhythmia.

Authors:  Isabel Deisenhofer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-09-13

Review 4.  [Modern mapping technologies : Technical background and clinical use].

Authors:  Felix Bourier; Frédéric Sacher
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-06-26

Review 5.  Catheter Ablation of Atrial Fibrillation Without Radiation Exposure Using A 3D Mapping System.

Authors:  Marco Scaglione; Elisa Ebrille; Francesca Di Clemente; Fiorenzo Gaita; Jason S Bradfield
Journal:  J Atr Fibrillation       Date:  2015-02-28

6.  Rate of acquired pulmonary vein stenosis after ablation of atrial fibrillation referred to electroanatomical mapping systems: Does it matter?

Authors:  Katharina Schoene; Arash Arya; Cosima Jahnke; Ingo Paetsch; Nikolaos Dagres; Andreas Bollmann; Gerhard Hindricks; Philipp Sommer
Journal:  Cardiol J       Date:  2018-09-24       Impact factor: 2.737

7.  Novel three-dimensional imaging approach for cryoballoon navigation and confirmation of pulmonary vein occlusion.

Authors:  Christopher A B Kowalewski; Miguel Rodrigo; Chad Brodt; Francois Haddad; Paul J Wang; Sanjiv M Narayan
Journal:  Pacing Clin Electrophysiol       Date:  2020-02-17       Impact factor: 1.976

8.  Contrast-Based 3D/2D Registration of the Left Atrium: Fast versus Consistent.

Authors:  Matthias Hoffmann; Christopher Kowalewski; Andreas Maier; Klaus Kurzidim; Norbert Strobel; Joachim Hornegger
Journal:  Int J Biomed Imaging       Date:  2016-03-08

9.  CT-fusion-guided transseptal puncture in a patient with atrial fibrillation and absent right superior vena cava.

Authors:  Felix Bourier; Sonia Ammar; Tilko Reents; Gabriele Hessling; Isabel Deisenhofer
Journal:  HeartRhythm Case Rep       Date:  2015-06-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.