Literature DB >> 26614147

In vitro comparison of epidural bacteria filters permeability and screening scanning electron microscopy.

Aysin Sener1, Yuksel Erkin2, Alper Sener3, Aydin Tasdogen2, Esra Dokumaci4, Zahide Elar2.   

Abstract

BACKGROUND AND OBJECTIVES: Epidural catheter bacteria filters are barriers in the patient-controlled analgesia/anaesthesia for preventing contamination at the epidural insertion site. The efficiency of these filters varies according to pore sizes and materials.
METHOD: The bacterial adhesion capability of the two filters was measured in vitro experiment. Adhesion capacities for standard Staphylococcus aureus (ATCC 25923) and Pseudomonas aeruginosa (ATCC 27853) strains of the two different filters (Portex and Rusch) which have the same pore size were examined. Bacterial suspension of 0.5 Mc Farland was placed in the patient-controlled analgesia pump, was filtered at a speed of 5 mL/h. in continuous infusion for 48 h and accumulated in bottle. The two filters were compared with colony counts of bacteria in the filters and bottles. At the same time, the filters and adhered bacteria were monitored by scanning electron microscope.
RESULTS: Electron microscopic examination of filters showed that the Portex filter had a granular and the Rusch filter fibrillary structure. Colony counting from the catheter and bottle showed that both of the filters have significant bacterial adhesion capability (p<0.001). After the bacteria suspension infusion, colony countings showed that the Portex filter was more efficient (p<0.001). There was not any difference between S. aureus and P. aeruginosa bacteria adhesion. In the SEM monitoring after the infusion, it was physically shown that the bacteria were adhered efficiently by both of the filters.
CONCLUSION: The granular structured filter was found statistically and significantly more successful than the fibrial. Although the pore sizes of the filters were same - of which structural differences shown by SEM were the same - it would not be right to attribute the changes in the efficiencies to only structural differences. Using microbiological and physical proofs with regard to efficiency at the same time has been another important aspect of this experiment.
Copyright © 2013 Sociedade Brasileira de Anestesiologia. Published by Elsevier Editora Ltda. All rights reserved.

Entities:  

Keywords:  Bacteria filter; Filtro antibacteriano; Microscópio eletrônico de varredura; Pseudomonas aeruginosa; Scanning electron microscope; Staphylococcus aureus

Mesh:

Year:  2013        PMID: 26614147     DOI: 10.1016/j.bjane.2013.08.004

Source DB:  PubMed          Journal:  Braz J Anesthesiol


  2 in total

1.  The oral microbiome-the relevant reservoir for acute pediatric appendicitis?

Authors:  Carlotta Blod; Nadine Schlichting; Sara Schülin; Anne Suttkus; Nicole Peukert; Catalina Suzana Stingu; Christian Hirsch; Wieland Elger; Martin Lacher; Ulf Bühligen; Steffi Mayer
Journal:  Int J Colorectal Dis       Date:  2017-12-22       Impact factor: 2.571

2.  Rupture of an epidural filter connector during bolus administration of local anesthetic: a case report.

Authors:  Daniel A Nahrwold; Aaron R Muncey; Nasrin N Aldawoodi; Raymond M Evans; Jamie P Hoffman
Journal:  BMC Anesthesiol       Date:  2021-05-12       Impact factor: 2.217

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.