Literature DB >> 26599665

Response to Standardized MR Terminology and Reporting of Implants and Devices as Recommended by the American College of Radiology Subcommittee on MR Safety.

Terry O Woods1, Jana G Delfino1, Mitchell J Shein1.   

Abstract

The U.S. Food and Drug Administration (FDA) continually works toward the goal of safety. For patients with magnetic resonance (MR) Conditional devices, safety is achieved when MR Conditional labeling is clear and accessible and can be unambiguously interpreted and applied. The FDA supports the three facets of standardization listed by the American College of Radiology (ACR) Subcommittee on MR Safety in their special report: (a) standardization in terminology and reporting of spatial gradient magnetic fields associated with MR systems; (b) standardization in reporting of ferromagnetic testing results for implants and devices; and (c) standardization, consistency, and clarity in radiofrequency power deposition guidelines and terminology. While the FDA is in agreement with the ACR Subcommittee on MR Safety that patient safety is of primary concern, the authors disagree with the Subcommittee on several important points and offer a point-by-point response to the Subcommittee's four recommendations. (©) RSNA, 2015.

Entities:  

Mesh:

Year:  2015        PMID: 26599665     DOI: 10.1148/radiol.2015151108

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  1 in total

1.  A Web-based searchable system to confirm magnetic resonance compatibility of implantable medical devices in Japan: a preliminary study.

Authors:  Yasuhiro Fujiwara; Hitoshi Fujioka; Tomoko Watanabe; Maiko Sekiguchi; Ryuji Murakami
Journal:  Radiol Phys Technol       Date:  2017-07-31
  1 in total