Literature DB >> 26580296

Head-locking durability of fixed and variable angle locking screws under repetitive loading.

Mark Lenz1,2, Dieter Wahl1, Ivan Zderic1, Boyko Gueorguiev1, Jesse B Jupiter3, Stephan M Perren1,4.   

Abstract

Polyaxial locking screws are increasingly applied in fracture fixation. To investigate the durability of the head-locking mechanism, the removal torque of variable angle (VA) and fixed angle (FA) stainless steel and titanium locking screws was investigated without and after a cyclic loading test. Stainless steel (St) and titanium (Ti) 2.4 mm orthogonally inserted FA screws and 2.4 mm VA screws inserted in different inclinations (0°-15°) (n = 6 per group) were locked at 0.8 Nm. Removal torque was determined without (W) and after (A) cyclic loading (sinusoidal load, 5 Hz, constant amplitude of 25 N, up to 10'000 cycles, or failure). Significant differences in-between the groups were detected by Student's t-test (p < 0.05). Except VA Ti in 0deg and FA, all groups exhibited a drop in removal torque below the insertion torque without and after cyclic testing. The removal torque was (St: FA W:0.81 Nm ± 0.04 A:0.72Nm ± 0.04; VA0deg W:0.73 Nm ± 0.04 A:0.65 Nm ± 0.05; VA15deg W:0.51 Nm ± 0.05 A:0.50 Nm ± 0.08; Ti: FA W:0.82 Nm ± 0.03 A:0.70 Nm ± 0.04; VA0deg W:0.80 Nm ± 0.02 A:0.72 Nm ± 0.05; VA15deg W:0.55 Nm ± 0.03 A:0.54 Nm ± 0.06). In all groups, the removal torque after cyclic testing did not drop below 16% of the removal torque without cyclic testing. No head loosening was observed after cyclic testing. Stainless steel and titanium 2.4 mm fixed and variable angle locking screws provide a stable and lasting head-locking mechanism.
© 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 34:949-952, 2016. © 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  fixed angle locking; locking durability; polyaxial locking; screw removal torque; variable angle locking

Mesh:

Year:  2015        PMID: 26580296     DOI: 10.1002/jor.23108

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  2 in total

1.  Half-threaded holes markedly increase the fatigue life of locking plates without compromising screw stability.

Authors:  Ching-Kong Chao; You-Lin Chen; Jinn Lin
Journal:  Bone Joint Res       Date:  2020-10-10       Impact factor: 5.853

2.  Locking plate constructs in subtrochanteric fixation: a biomechanical comparison of LCP screws and AO-nuts.

Authors:  T P A Baltes; A J van der Veen; L Blankevoort; J C E Donders; P Kloen
Journal:  J Clin Orthop Trauma       Date:  2020-12-31