| Literature DB >> 26578461 |
Andreas Thomas1, Christian Görgens1, Sven Guddat1, Detlef Thieme2, Frank Dellanna3, Wilhelm Schänzer1, Mario Thevis1.
Abstract
The analysis of low-molecular-mass peptides in doping controls has become a mandatory aspect in sports drug testing and, thus, the number of samples that has to be tested for these analytes has been steadily increasing. Several peptides <2 kDa with performance-enhancing properties are covered by the list of prohibited substances of the World Anti-Doping Agency including Desmopressin, LH-RH, Buserelin, Triptorelin, Leuprolide, GHRP-1, GHRP-2, GHRP-3, GHRP-4, GHRP-5,GHRP-6, Alexamorelin, Ipamorelin, Hexarelin, ARA-290, AOD-9604, TB-500 and Anamorelin. With the presented method employing direct urine injection into a liquid chromatograph followed by ion-mobility time-of-flight mass spectrometry, a facile, specific and sensitive assay for the aforementioned peptidic compounds is provided. The accomplished sensitivity allows for limits of detection between 50 and 500 pg/mL and thus covers the minimum required performance level of 2 ng/mL accordingly. The method is precise (imprecision <20%) and linear in the estimated working range between 0 and 10 ng/mL. The stability of the peptides in urine was tested, and -20°C was found to be the appropriate storage temperature for sports drug testing. Finally, proof-of-concept was shown by analysing elimination study urine samples collected from individuals having administered GHRP-6, GHRP-2, or LHRH.Entities:
Keywords: Doping control; High-resolution mass spectrometry; Peptide analysis
Mesh:
Substances:
Year: 2015 PMID: 26578461 DOI: 10.1002/jssc.201501060
Source DB: PubMed Journal: J Sep Sci ISSN: 1615-9306 Impact factor: 3.645