Axial displacement of external and internal implant-abutment connection evaluated by linear mixed model analysis.

Hyon-Woo Seol, Seong-Joo Heo, Jai-Young Koak, Seong-Kyun Kim, Shin-Koo Kim.   

Abstract

PURPOSE: To analyze the axial displacement of external and internal implant-abutment connection after cyclic loading.
MATERIALS AND METHODS: Three groups of external abutments (Ext group), an internal tapered one-piece-type abutment (Int-1 group), and an internal tapered two-piece-type abutment (Int-2 group) were prepared. Cyclic loading was applied to implant-abutment assemblies at 150 N with a frequency of 3 Hz. The amount of axial displacement, the Periotest values (PTVs), and the removal torque values(RTVs) were measured. Both a repeated measures analysis of variance and pattern analysis based on the linear mixed model were used for statistical analysis. Scanning electron microscopy (SEM) was used to evaluate the surface of the implant-abutment connection.
RESULTS: The mean axial displacements after 1,000,000 cycles were 0.6 μm in the Ext group, 3.7 μm in the Int-1 group, and 9.0 μm in the Int-2 group. Pattern analysis revealed a breakpoint at 171 cycles. The Ext group showed no declining pattern, and the Int-1 group showed no declining pattern after the breakpoint (171 cycles). However, the Int-2 group experienced continuous axial displacement. After cyclic loading, the PTV decreased in the Int-2 group, and the RTV decreased in all groups. SEM imaging revealed surface wear in all groups.
CONCLUSION: Axial displacement and surface wear occurred in all groups. The PTVs remained stable, but the RTVs decreased after cyclic loading. Based on linear mixed model analysis, the Ext and Int-1 groups' axial displacements plateaued after little cyclic loading. The Int-2 group's rate of axial displacement slowed after 100,000 cycles.

Entities:  

Mesh:

Year:  2015        PMID: 26574863     DOI: 10.11607/jomi.3857

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  2 in total

1.  Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

2.  Effect of lateral oblique cyclic loading on microleakage and screw loosening of implants with different connections.

Authors:  Hakimeh Siadat; Hossain Najafi; Marzieh Alikhasi; Babak Falahi; Elaheh Beyabanaki; Farid Zayeri
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2018-09-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.