Francesco Dell'Olio, Donato Conteduca, Caterina Ciminelli, Mario Nicola Armenise.
Abstract
A multi-analyte biosensing platform with ultra-high resolution ( = 0.2 ng/mL),-which is appropriate for the detection in the human serum of a wide range of biomarkers, e.g. those allowing the lung cancer early diagnosis, has been designed. The platform is based on a new configuration of planar ring resonator. The very strong light-matter interaction enabled by the micro-cavity allows a record limit-of-detection of 0.06 pg/mm(2), five times better than the state-of-the-art. The device with footprint = 2,200 μm(2) for each ring, due to its features, has the potential to be integrated in lab-on-chip microsystems for large-scale screenings of people with high risk of developing cancer.Entities:
Year: 2015 PMID: 26561129 DOI: 10.1364/OE.23.028593
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894