Freek Nijhoff1, Pieter R Stella1, Maartje S Troost1, Anouar Belkacemi2, Hendrik M Nathoe1, Michiel Voskuil1, Mariam Samim1, Pieter A Doevendans1, Pierfrancesco Agostoni3,4.
Abstract
BACKGROUND/Entities:
Keywords: Drug-eluting balloon; In-stent restenosis; Paclitaxel; Percutaneous coronary intervention
Mesh:
Substances:
Year: 2015 PMID: 26541146 PMCID: PMC4841852 DOI: 10.1007/s00392-015-0934-0
Source DB: PubMed Journal: Clin Res Cardiol ISSN: 1861-0684 Impact factor: 5.460
Device characteristics of the drug-eluting balloons
| DIOR | IN.Pact Falcon | |
|---|---|---|
| Manufacturer | EuroCor GmbH, Germany | Medtronic Vascular Inc., USA |
| Balloon type | Semi-compliant | Semi-compliant |
| Balloon diameters available, mm | 2.0, 2.25, 2.5, 2.75, 3.0, 3.5, 4.0 | 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0 |
| Balloon lengths available, mm | 15, 20, 25, 30 | 14, 20, 30, 40 |
| Coated drug | Paclitaxel | Paclitaxel |
| Drug | Paclitaxel | Paclitaxel |
| Loading dose | 3 µg/mm2 | 3 µg/mm2 |
| Excipient | Shellac | Urea |
| Coating procedure | Micro-pipetting | Not specified |
| Inflation time, s | 30–60 | 30–60 |
| CE marking | 2007 | 2009 |
CE Conformité Européene
Baseline patient and lesion characteristics
| DIOR ( | IN.Pact Falcon ( |
| |
|---|---|---|---|
| Patient characteristics | |||
| Age, years | 66.6 ± 10.27 | 65.3 ± 9.69 | 0.66 |
| Male gender | 14 (70) | 17 (68) | 0.89 |
| Diabetes mellitus | 4 (20) | 6 (24) | 0.75 |
| Hypertension | 11 (55) | 14 (56) | 0.95 |
| Hyperlipidemia | 13 (65) | 7 (28) | 0.01 |
| Current smoker | 3 (15) | 4 (16) | 0.93 |
| Family history of cardiovascular disease | 10 (50) | 12 (48) | 0.89 |
| Previous myocardial infarction | 10 (50) | 14 (56) | 0.69 |
| Previous coronary artery bypass grafting | 3 (15) | 3 (12) | 0.77 |
| Lesion characteristics | |||
| Target vessel | 0.22 | ||
| Left anterior descending | 10 (50) | 12 (48) | |
| Ramus circumflex | 6 (30) | 3 (12) | |
| Right coronary artery | 4 (20) | 10 (40) | |
| Pattern of restenosisa | 0.47 | ||
| Focal body | 8 (40) | 5 (20) | |
| Multifocal | 0 | 0 | |
| Diffuse in-stent | 8 (40) | 15 (60) | |
| Proliferative | 3 (15) | 3 (12) | |
| Occlusive | 1 (5) | 2 (8) | |
| Index stent type | 0.44 | ||
| Bare-metal stent | 15 (75) | 22 (88) | |
| Drug-eluting stent | 5 (25) | 3 (12) | |
| Index stent diameter, mm | 2.97 ± 0.38 | 3.01 ± 0.47 | 0.64 |
| Index stent length, mm | 33.5 ± 14.5 | 28.6 ± 13.3 | 0.26 |
aClassified according to the Mehran classification
Procedural features
| DIOR ( | IN.Pact Falcon ( |
| |
|---|---|---|---|
| Predilatation with standard balloon | 20 (100) | 24 (96) | 0.37 |
| Predilatation balloon diameter, mm | 3.1 ± 0.4 | 2.9 ± 0.4 | 0.07 |
| Predilatation balloon length, mm | 17.4 ± 5.4 | 20.7 ± 7.9 | 0.26 |
| Predilatation pressure, ATM | 14.8 ± 4.5 | 14.2 ± 5.0 | 0.71 |
| >1 DEB used per lesion | 2 (10) | 6 (24) | 0.27 |
| DEB diameter, mm | 3.2 ± 0.4 | 3.1 ± 0.3 | 0.34 |
| 2.50 | 2 (10) | 1 (4) | |
| 2.75 | 0 | 3 (12) | |
| 3.00 | 9 (45) | 13 (52) | |
| 3.50 | 8 (40) | 8 (32) | |
| 4.00 | 1 (5) | 0 | |
| DEB length, mm | 27.3 ± 9.4 | 32.4 ± 12.0 | 0.10 |
| DEB inflation pressure, ATM | 10.8 ± 3.0 | 11.3 ± 3.0 | 0.54 |
| DEB inflation time, s | 58.0 ± 5.3 | 52.0 ± 13.2 | 0.18 |
| Postdilatation with standard balloon | 3 (15) | 1 (4) | 0.31 |
| Maximum balloon diameter to index stent diameter ratio | 1.09 ± 0.08 | 1.07 ± 0.09 | 0.34 |
| Additional stenting | 1 (5) | 2 (8) | 0.69 |
| Angiographic success | 20 (100) | 25 (100) | |
| Device success | 18 (90) | 23 (92) | 0.69 |
| Procedural success | 20 (100) | 25 (100) |
Quantitative angiography and fractional flow reserve measurements
| DIOR ( | IN.Pact Falcon ( |
| |
|---|---|---|---|
| Preprocedural | |||
| Reference vessel diameter, mm | 2.32 ± 0.51 | 2.35 ± 0.46 | 0.82 |
| Minimal lumen diameter, mm | 0.59 ± 0.28 | 0.58 ± 0.38 | 0.98 |
| Diameter stenosis, % | 75.0 ± 12.5 | 75.3 ± 16.1 | 0.94 |
| Lesion length, mm | 23.7 ± 9.5 | 26.4 ± 12.6 | 0.52 |
| Fractional flow reserve | 18 (90) | 22 (88) | |
| Distal of the stent | 0.65 ± 0.11 | 0.58 ± 0.17 | 0.31 |
| In-stent gradient | 0.33 ± 0.12 | 0.37 ± 0.18 | 0.54 |
| Postprocedural | |||
| Minimal lumen diameter, mm | 1.88 ± 0.62 | 1.83 ± 0.47 | 0.79 |
| Diameter stenosis, % | 20.3 ± 9.17 | 27.5 ± 15.9 | 0.20 |
| Acute gain, mm | 1.29 ± 0.50 | 1.26 ± 0.61 | 0.85 |
| Residual binary stenosis | 0 | 2 (8) | 0.50 |
| Fractional flow reserve | 20 (100) | 25 (100) | |
| Distal of the stent | 0.93 ± 0.05 | 0.92 ± 0.05 | 0.35 |
| In-stent gradient | 0.05 ± 0.03 | 0.06 ± 0.04 | 0.40 |
| Follow-up | 18 (90) | 23 (92) | |
| Minimal lumen diameter in-stent, mm | 1.46 ± 0.68 | 1.83 ± 0.62 | 0.064 |
| Minimal lumen diameter in-segment, mm | 1.41 ± 0.66 | 1.69 ± 0.56 | 0.11 |
| Diameter stenosis in-stent, % | 40.1 ± 23.9 | 26.0 ± 18.3 | 0.049 |
| Diameter stenosis in-segment, % | 41.3 ± 22.6 | 30.7 ± 16.2 | 0.083 |
| Late lumen loss in-stent, mm | 0.41 ± 0.54 | 0.01 ± 0.43 | 0.026 |
| Late lumen loss in-segment, mm | 0.36 ± 0.48 | -0.03 ± 0.43 | 0.014 |
| Binary restenosis in-stent | 7 (39) | 4 (17) | 0.16 |
| Binary restenosis in-segment | 7 (39) | 4 (17) | 0.16 |
| Fractional flow reserve | 18 (90) | 23 (92) | |
| Distal of the stent | 0.84 ± 0.13 | 0.92 ± 0.07 | 0.029 |
| In-stent gradient | 0.13 ± 0.12 | 0.05 ± 0.05 | 0.002 |
| 6-months clinical outcome | 20 (100) | 25 (100) | |
| Cardiac death | 0 | 0 | |
| Myocardial infarction | 0 | 0 | |
| Stent thrombosis | 0 | 0 | |
| Target lesion revascularization | 7 (35) | 2 (8) | 0.057 |
Fig. 1Illustrative OCT imaging of the same coronary segment for each time point (baseline, post-PCI and follow-up). Severe in-stent restenosis at baseline (a and d). After PCI, lumen enlargement with neointimal disruption and (micro) dissections are observed (b and e), caused by the mechanical effect of DEB angioplasty. Follow-up shows complete healing of the dissections with a moderate increase (C) and limited decrease in neointima (F)
Optical coherence tomographic cross-section analysis
| DIOR ( | IN.pact falcon ( |
| |
|---|---|---|---|
| Preprocedure | 16 (80) | 19 (76) | |
| Stent length analyzed, mm | 23.9 [20.1–30.0] | 22.0 [15.4–35.7] | 0.29 |
| Mean lumen area, mm2 | 3.1 [2.3–4.5] | 3.6 [2.7–4.5] | 0.32 |
| Minimal lumen area, mm2 | 1.1 [0.7–1.4] | 1.3 [0.9–1.8] | 0.13 |
| Mean stent area, mm2 | 6.5 [5.6–8.5] | 7.2 [6.2–8.9] | 0.31 |
| Minimal stent area, mm2 | 5.3 [3.6–7.2] | 5.4 [4.4–7.1] | 0.32 |
| Mean neointimal area stenosis, % | 50.7 [35.4–66.1] | 53.0 [43.7–58.9] | 0.90 |
| Maximum neointimal area stenosis, % | 80.4 [74.2–88.7] | 82.8 [73.6–85.9] | 0.79 |
| Lumen volume, mm3 | 89.7 [45.8–103.4] | 77.4 [60.5–129.1] | 0.92 |
| Stent volume, mm3 | 156 [132–227] | 176 [124–217] | 0.92 |
| Neointimal volume, mm3 | 66.4 [53.8–134] | 87.4 [69.4–102] | 0.53 |
| Neointima occupied stent volume, % | 51.5 [34.9–65.1] | 53.2 [40.8–58.1] | 0.95 |
| Postprocedure | 20 (100) | 25 (100) | |
| Stent length analyzed, mm | 23.9 [19.7–29.9] | 21.8 [15.6–32.0] | 0.40 |
| Mean lumen area, mm2 | 5.0 [4.1–6.8] | 6.2 [5.5–7.1] | 0.093 |
| Minimal lumen area, mm2 | 3.3 [2.6–4.7] | 4.7 [3.0–5.4] | 0.10 |
| Mean stent area, mm2 | 7.4 [6.0–9.7] | 9.8 [8.8–11.0] | 0.016 |
| Minimal stent area, mm2 | 6.1 [4.1–8.5] | 8.2 [7.5–9.2] | 0.024 |
| Mean neointimal area stenosis, % | 28.3 [24.4–40.4] | 33.5 [30.3–36.8] | 0.58 |
| Maximum neointimal area stenosis, % | 47.8 [37.8–55.2] | 48.3 [43.1–50.8] | 0.96 |
| Neointima occupied stent volume, % | 28.2 [24.8–40.2] | 33.6 [29.0–37.4] | 0.38 |
| Follow-up | 17 (85) | 22 (88) | |
| Stent length analyzed, mm | 23.8 [21.2–31.7] | 22.4 [18.1–32.8] | 0.28 |
| Mean lumen area, mm2 | 4.6 [3.4–6.1] | 6.0 [5.2–7.8] | 0.008 |
| Minimal lumen area, mm2 | 2.4 [1.8–3.7] | 4.0 [3.1–6.0] | 0.009 |
| Mean stent area, mm2 | 7.6 [6.1–9.5] | 9.1 [7.5–11.5] | 0.066 |
| Minimal stent area, mm2 | 6.5 [4.4–8.0] | 7.7 [5.3–9.8] | 0.16 |
| Mean neointimal area stenosis, % | 42.8 [23.7–55.3] | 31.6 [24.9–37.5] | 0.011 |
| Maximum neointimal area stenosis, % | 66.4 [49.9–76.6] | 47.7 [37.3–60.7] | 0.010 |
| Neointima occupied stent volume, % | 41.7 [26.0–55.1] | 30.5 [23.7–36.5] | 0.047 |
Fig. 2Cumulative distribution of in-segment late lumen loss for the DIOR and IN.PACT Falcon
Percentage changes between postprocedure and follow-up
| DIOR ( | IN.pact falcon ( | Crude | Adjusted | |
|---|---|---|---|---|
| Quantitative coronary angiography | 18 (90) | 23 (92) | ||
| Minimal lumen diameter change, % | –10.4 [–42.3 to 4.7] | 4.8 [−7.7 to 16.8] | 0.034 | 0.26 |
| Diameter stenosis change, % | 28.6 [−7.1 to 187] | 2.6 [−49.7 to 67.6] | 0.032 | 0.32 |
| Optical coherence tomography | 17 (85) | 21 (84) | ||
| Minimal lumen area change, % | −30.2 [−49.5 to 2.6] | −13.4 [−21.8 to 19.4] | 0.048 | 0.097 |
| Maximal neointimal area change, % | 33.8 [6.2 to 72.0] | −8.9 [−21.0 to 33.0] | 0.002 | 0.007 |
| Maximal neointimal area stenosis change, % | 35.8 [8.7 to 59.1] | 14.6 [−21.7 to 36.6] | 0.014 | 0.009 |
| Lumen volume change, % | −14.6 [−34.0 to 1.7] | 2.89 [−14.0 to 18.6] | 0.011 | 0.026 |
| Stent volume change, % | −0.7 [−3.0 to 2.7] | −1.6 [−6.9 to 5.9] | 0.67 | 0.77 |
| Neointimal volume change, % | 27.2 [1.1 to 58.6] | −15.8 [−36.7 to 28.3] | 0.006 | 0.028 |
| Fractional flow reserve | 17 (85) | 22 (88) | ||
| FFR stent gradient change, % | 69.0 [0.0 to 238] | −40.8 [−58.9 to 18.8] | 0.003 | 0.46 |
* Adjusted for differences in hypercholesterolemia, postdilatation and postprocedure OCT minimal lumen area
Fig. 3Percentage changes in postprocedure and follow-up in stent, neointima and lumen volumes derived from optical coherence tomography (negative values represent a decrease and positive values an increase in volumes during follow-up)