Literature DB >> 26522766

Repeatability of QuantiFERON-TB gold in-tube assay results near cut-off points.

Kwang Sook Woo1, Jae Lim Choi1, Bo Ram Kim1, Jin Yeong Han1, Jung Man Kim1, Kyeong Hee Kim2.   

Abstract

Entities:  

Mesh:

Year:  2016        PMID: 26522766      PMCID: PMC4697350          DOI: 10.3343/alm.2016.36.1.76

Source DB:  PubMed          Journal:  Ann Lab Med        ISSN: 2234-3806            Impact factor:   3.464


× No keyword cloud information.
Dear Editor, Interferon (IFN)-γ release assay (IGRA) is a functional assay used to measure T cell responses to Mycobacterium tuberculosis-specific antigens in whole blood or blood-derived mononuclear cells [12]. The QuantiFERON-TB Gold In-Tube (QFT-GIT; Cellestis, Carnegie, Australia) is a commercially available IGRA. Significant within-subject variability in QFT-GIT results has been reported in tuberculosis (TB) screening [34]. Variability in QFT-GIT results for the retesting of the same patient sample has been observed in a low TB-incidence setting. Twenty-eight (8%) of 366 patients showed results different from those of the initial QFT-GIT, and 24 (85.7%) of these 28 patients had an initial TB response between 0.25 and 0.80 IU/mL [5]. To explore the clinical impact of QFT-GIT results with borderline TB response in an intermediate TB-incidence setting, we specifically assessed the repeatability of QFT-GIT near cut-off points in routine laboratory practice. We prospectively included 748 samples referred for TB screening between March, 2013 and February 2014 in the Dong-A University Hospital, Busan, Korea. This study was approved by the Institutional Review Board of Dong-A University Hospital. The QFT-GIT was performed according to the manufacturer's instructions. A TB response between 0.25 and 0.80 IU/mL was adopted as a borderline range [5]. After the initial test, the residual samples with borderline range of TB response, were stored and retested. According to the manufacturer's interpretation criteria, we compared and analyzed the first and second results. Bland-Altman plots were used for quantitative results. Data analysis was performed by using MedCalc Software (ver. 12.6.1, MedCalc Software, Mariakerke, Belgium). Eighty-five of the 748 patients showed a TB response in the range of 0.25-0.80 IU/mL. Among these 85 patients, 25 (29.4%) were diagnosed as having TB, 28 (33%) had suspected symptoms of TB, and 32 (37.6%) were examined by QFT-GIT to screen for latent TB infection. The median age was 56 yr, and 47 (55.3%) were males. The first QFT-GIT result was positive in 49 patients (57.6%), negative in 35 (41.2%), and indeterminate in 1 (1.2%). Seventy-one patients (83.5%) showed concordance between the results of the first and second QFT-GIT. Of the 14 patients with discordant results (14/85, 16.4%), three showed positive conversion, eight showed negative reversion, one with an indeterminate result in the first QFT-GIT showed a negative result in the second QFT-GIT, and two with positive results in the first QFT-GIT showed indeterminate results in the second QFT-GIT. The discordant results were most frequent (5/17, 29.5%) in the group with second-quartile TB response where the values included the cut-off value of 0.35 IU/mL. However, the likelihood ratio for subsequent discordant results was not statistically significant (P=0.48; Table 1). Bland-Altman plots of the differences in the first and second responses for nil, mitogen, and TB responses are shown in Fig. 1; the mean difference in the mitogen response was the highest.
Table 1

Distribution of discordant results based on the initial TB response

Initial TB response (IU/mL)Total (N = 85)Discordant result (N = 14)Likelihood ratio (95% CI)
Q1 0.25-0.31273 (11.1)0.63 (0.22-1.82)
Q2 0.32-0.38175 (29.4)2.11 (0.88-5.05)
Q3 0.39-0.54204 (20.0)1.27 (0.50-3.23)
Q4 0.55-0.80212 (9.5)0.53 (0.14-2.04)

Values are expressed as numbers (%). The subsequent discordant result according to initial TB response was not statistically significant (P=0.48, Chi-squared test).

Abbreviations: TB, tuberculosis; Q, quartile; CI, confidence interval.

Fig. 1

Bland-Altman plots for the first-second response differences for nil (A), mitogen (B), and TB responses (C); the analysis showed mean differences of -0.02 IU/mL, -0.2 IU/mL, and 0.03 IU/mL, respectively. The lower and upper limits of agreement (mean difference±1.96 SD) were -0.5 IU/mL and 0.5 IU/mL for the nil response, -2.6 IU/mL and 2.2 IU/mL for the mitogen response, and -0.5 IU/mL and 0.6 IU/mL for the TB response, respectively. Mitogen1 indicated initial result for initial mitogen stimulation and mitogen2 indicated second result for second mitogen stimulation.

Abbreviation: TB, tuberculosis.

The repeatability and reproducibility of QFT-GIT have important clinical considerations because the test uses a single cutoff point to distinguish between positive and negative results with one-time testing. The low positive result for TB response (0.35-0.59 IU/mL) could be supposed to revert to negative in the retest on the basis of inherent variability of the QFT-GIT using a linear mixed effects model [5]. The conversion and reversion may have occurred in patients undergoing serial testing, and this could result in unnecessary preventive therapy while screening for latent TB infection [67]. In this study, we observed that a considerable portion of the QFT-GIT results near the cut-off points showed discordance on retesting of the same sample. Thus, it might be worth introducing a borderline decision range for quantitative TB antigen response in QFT-GIT. In addition, we found that the mitogen response would be more dynamic than nil and TB responses, consistent with those of a previous study involving serial QFT-GIT assays in 299 patients [8]. Although we could not examine subjects with a TB response <0.25 IU/mL and >0.8 IU/mL, nearly two-thirds of the discordant results were observed between the cut-off points minus 0.03 IU/mL and plus 0.19 IU/mL for the TB response (0.32-0.54 IU/mL). This range considerably overlaps a low positive zone (0.35-0.59 IU/mL) defined by a previous study on QFT-GIT variability [5]. In conclusion, we suggest that to improve the clinical interpretation of QFT-GIT results, the results near the cut-off points in particular should be interpreted cautiously, and the establishment of a borderline decision range for quantitative TB antigen response in QFT-GIT should also be considered.
  8 in total

1.  Within-subject variability of Mycobacterium tuberculosis-specific gamma interferon responses in German health care workers.

Authors:  Felix C Ringshausen; Albert Nienhaus; José Torres Costa; Heiko Knoop; Stephan Schlösser; Gerhard Schultze-Werninghaus; Gernot Rohde
Journal:  Clin Vaccine Immunol       Date:  2011-05-18

2.  Updated guidelines for using Interferon Gamma Release Assays to detect Mycobacterium tuberculosis infection - United States, 2010.

Authors:  Gerald H Mazurek; John Jereb; Andrew Vernon; Phillip LoBue; Stefan Goldberg; Kenneth Castro
Journal:  MMWR Recomm Rep       Date:  2010-06-25

Review 3.  Reproducibility of interferon gamma (IFN-γ) release Assays. A systematic review.

Authors:  Saloua Tagmouti; Madeline Slater; Andrea Benedetti; Sandra V Kik; Niaz Banaei; Adithya Cattamanchi; John Metcalfe; David Dowdy; Richard van Zyl Smit; Nandini Dendukuri; Madhukar Pai; Claudia Denkinger
Journal:  Ann Am Thorac Soc       Date:  2014-10

4.  Significance of interferon-gamma response to mitogen in serial QuantiFERON-TB Gold In-Tube assay of routine laboratory practice.

Authors:  Kwang-Sook Woo; Jae-Lim Choi; Bo-Ram Kim; Ji-Eun Kim; Byoung-Gwon Kim; Hyuck Lee; Kyeong-Hee Kim
Journal:  Clin Chim Acta       Date:  2014-01-03       Impact factor: 3.786

Review 5.  Gamma interferon release assays for detection of Mycobacterium tuberculosis infection.

Authors:  Madhukar Pai; Claudia M Denkinger; Sandra V Kik; Molebogeng X Rangaka; Alice Zwerling; Olivia Oxlade; John Z Metcalfe; Adithya Cattamanchi; David W Dowdy; Keertan Dheda; Niaz Banaei
Journal:  Clin Microbiol Rev       Date:  2014-01       Impact factor: 26.132

6.  Monthly follow-ups of interferon-γ release assays among health-care workers in contact with patients with TB.

Authors:  Jong Sun Park; Ji Sun Lee; Min Young Kim; Chang Hoon Lee; Ho Il Yoon; Sang-Min Lee; Chul-Gyu Yoo; Young Whan Kim; Sung Koo Han; Jae-Joon Yim
Journal:  Chest       Date:  2012-12       Impact factor: 9.410

7.  Test variability of the QuantiFERON-TB gold in-tube assay in clinical practice.

Authors:  John Z Metcalfe; Adithya Cattamanchi; Charles E McCulloch; Justin D Lew; Ngan P Ha; Edward A Graviss
Journal:  Am J Respir Crit Care Med       Date:  2012-10-26       Impact factor: 21.405

8.  IGRAs--the gateway to T cell based TB diagnosis.

Authors:  Hilary S Whitworth; Melanie Scott; David W Connell; Bianca Dongés; Ajit Lalvani
Journal:  Methods       Date:  2013-01-04       Impact factor: 3.608

  8 in total
  4 in total

1.  Optimization and Interpretation of Serial QuantiFERON Testing to Measure Acquisition of Mycobacterium tuberculosis Infection.

Authors:  Elisa Nemes; Virginie Rozot; Hennie Geldenhuys; Nicole Bilek; Simbarashe Mabwe; Deborah Abrahams; Lebohang Makhethe; Mzwandile Erasmus; Alana Keyser; Asma Toefy; Yolundi Cloete; Frances Ratangee; Thomas Blauenfeldt; Morten Ruhwald; Gerhard Walzl; Bronwyn Smith; Andre G Loxton; Willem A Hanekom; Jason R Andrews; Maria D Lempicki; Ruth Ellis; Ann M Ginsberg; Mark Hatherill; Thomas J Scriba
Journal:  Am J Respir Crit Care Med       Date:  2017-09-01       Impact factor: 21.405

2.  Adults with Mycobacterium tuberculosis infection and pre-diabetes have increased levels of QuantiFERON interferon-gamma responses.

Authors:  Matthew J Magee; Susanna L Trost; Argita D Salindri; Genet Amere; Cheryl L Day; Neel R Gandhi
Journal:  Tuberculosis (Edinb)       Date:  2020-04-02       Impact factor: 3.131

3.  The use of a borderline zone for the interpretation of interferon-gamma release assay results for serial screening of healthcare workers.

Authors:  Jae Hyeon Park; Namhee Kim; Hyunwoong Park; Taek Soo Kim; Sang-Won Park; Eun Youn Roh; Jong Hyun Yoon; Sue Shin
Journal:  PLoS One       Date:  2020-06-30       Impact factor: 3.240

4.  Performance Evaluation of a New Automated Chemiluminescent Immunoanalyzer-Based Interferon-Gamma Releasing Assay AdvanSure I3 in Comparison With the QuantiFERON-TB Gold In-Tube Assay.

Authors:  Jin Ju Kim; Younhee Park; Dasom Choi; Hyon Suk Kim
Journal:  Ann Lab Med       Date:  2020-01       Impact factor: 3.464

  4 in total