| Literature DB >> 26518666 |
Rui Providência1, Daniel B Kramer2, Dominic Pimenta3, Girish G Babu4, Laura A Hatfield5, Adam Ioannou6, Jan Novak7, Robert G Hauser8, Pier D Lambiase9.
Abstract
BACKGROUND: Despite the widespread use of implantable cardioverter-defibrillators (ICDs) in clinical practice, concerns exist regarding ICD lead durability. The performance of specific lead designs and factors determining this in large populations need clarification. METHODS ANDEntities:
Keywords: adverse event; complications; failure; fracture; implantable cardioverter‐defibrillator
Mesh:
Year: 2015 PMID: 26518666 PMCID: PMC4845221 DOI: 10.1161/JAHA.115.002418
Source DB: PubMed Journal: J Am Heart Assoc ISSN: 2047-9980 Impact factor: 5.501
Figure 1Study selection process. ICD indicates implantable cardioverter‐defibrillator.
Main Characteristics of the Studies (n=17) Included in the Meta‐Analysis
| Author, Year | Study Design Follow‐up Duration | Baseline Population | Lead Failure Definition |
|---|---|---|---|
| Kleeman et al 2007 |
Prospective single‐center registry; |
All recipients of a transvenous ICD system (first implantation); | Severe lead failure that required surgical correction. Lead failure required ≥1 of the following: (1) Oversensing unrelated to the cardiac cycle was documented; (2) lead impedance was out of normal range, and a surgical revision was suggested by the experts of the manufacturer; (3) a fracture was observed on x‐ray; (4) evidence existed of a lead failure during electrical testing |
| Eckstein et al 2008 |
Retrospective multicentric (3 sites) study; | All patients implanted with transvenous ICD leads over a decade | Lead‐related problem requiring surgical revision to be corrected. Lead problems were defined as structural problems, including insulation defects and lead fractures, and functional problems, including noise or far‐field sensing, T‐wave oversensing, and others (noise resulting from contact with another lead, unstable impedance measurements, R‐wave reduction, and loss of capture) |
| Hauser et al 2008 |
Retrospective multicentric (7 sites) study; |
All patients with HCM receiving transvenous high‐voltage ICD leads; | Lead undersensed or oversensed normal cardiac electrical activity or could not provide effective electrical therapy (including sensing, pacing, or defibrillation) as the result of a noniatrogenic device‐related defect in the insulation, conductor, electrode, shocking coil, fixation mechanism, or terminal pin, ring, or seal that connects the lead to the ICD pulse generator |
| Borleffs et al 2009 | Retrospective single‐center study; follow‐up: 4330 patient‐years; mean: 2.4 years | All recipients of ICD leads; age 61±13 years; 80% male; dual chamber, 49%; CRT, 36% | At least 1 of the following criteria had to be met to define suspected lead failure (1 and 2) or verified lead failure (3–6): (1) loss of capture or markedly elevated thresholds; (2) loss of sensing, oversensing, or skeletal muscular stimulation; (3) a visible conductor fracture or insulation defect seen at surgery; (4) a change in lead impedance, judged to be caused by conductor or insulation failure; (5) an evident fracture seen on chest roentgenogram; or (6) manufacturer's returned product report confirming the failure |
| Hauser and Hayer 2009 | Prospective multicentric (2 sites) registry; follow‐up: 5390 patient‐years; mean: 1.9 years | All patients implanted with ICD leads | Lead undersensed or oversensed normal cardiac electrical activity or if it could not provide effective electrical therapy including sensing, pacing, or defibrillation. Displacements after the index procedure were considered a failure only if the lead had to be removed from service and was found to have a defective fixation mechanism |
| Morrison et al 2011 | Prospective multicentric (3 sites) study; follow‐up: 8421 patient‐years; mean: 3.2 years | All recipients of the prespecified ICD leads; age 64.5±14.6 years; 77.3% male; CRT, 31.4% | Lead removed from service due to an inability to meet its performance specifications or otherwise perform as intended. A lead failed if (1) it exhibited abnormal impedance; (2) it exhibited electrical noise as manifested by nonphysiological signals on the electrogram or by pulse generator diagnostic data suggesting rapid oversensing, for example, nonphysiological short intervals and/or recurrent nonsustained ventricular tachycardia with intervals usually <220 ms; or (3) it could not sense R waves and/or provide effective electrical therapy due to apparent structural defect such as a conductor fracture or insulation breach |
| Parkash et al 2012 | Randomized controlled trial (RAFT) subanalysis; follow‐up: 5808 patient‐years; mean: 3.3 years | All participants in the trial; age 66.1±9.4 years; 83% male; CRT, 49.7% | If 2 of the following: (1) impedance rise (>50% or 500 Ω in 1 week), (2) short interval count >10 times per day or 300 times per month, and (3) inappropriate shock because of noise recorded on the electrogram |
| Sung et al 2012 | Prospective multicentric (150 sites) US Department of Veteran Affairs registry; follow‐up: 51 592 patient‐years; mean: 3.5 years | All patients implanted with the prespecified leads | If 1 of the following was met: (1) presence of nonphysiological noise not due to external interference such as electromagnetic interference; (2) rise in pace/sense (p/s) conductor impedance to >2000 Ω usually from baseline impedance <1000 Ω or >2× rise in stable baseline impedance; (3) drop in p/s conductor impedance to less than half of the previously stable baseline value or to impedance <200 Ω from baseline impedance >300 Ω; (4) change in superior vena cava or high‐voltage coil impedance to >200 or <25 Ω; and (5) rise in capture threshold to >2 times the previously stable value |
| Abdelhadi et al 2013 | Retrospective multicentric (7 sites) study; follow‐up: 9509 patient‐years; mean: 3.5 years |
Adults implanted with the prespecified ICD leads; age | If the lead did not perform according to its specifications or otherwise function as intended, including electrical malfunction and externalized conductors |
| Ellenbogen et al 2013 | Retrospective multicentric study; follow‐up: 11 424 patient‐years; mean: 0.9 year | All patients with a Medtronic ICD with LIA enabled; age 66.3±14.3 years; 73.1% male; dual chamber, 38.0%; CRT, 45.6% | Lead‐system event, 1 of the following: lead failure, connector issue, dislodgement, perforation, and lead–lead interaction. Lead‐failure was defined based on electrogram tracings, sensing integrity counters, RR intervals, and lead impedance trends |
| Fazal et al 2013 | Retrospective single‐center study; follow‐up: 984 patient‐years; mean: 4.5 years | All patients implanted with the prespecified ICD leads; age 64.4±18.5 years; 83.8% male; dual chamber, 31.0%; CRT, 33.3% | Any of the following: (1) sudden increase in pacing or defibrillation impedance from baseline without alternative explanation; (2) frequent short V–V intervals implying fracture of the conductor, contact between the 2 components generating electrical “noise” (sensing artifact); (3) delivery of inappropriate shock(s) as a consequence of interpretation of these sensing artifacts as ventricular arrhythmia; and (4) failure of effective electrical therapy including sensing, pacing, or defibrillation |
| Rordorf et al 2013 | Retrospective single‐center study; follow‐up: 4515 patient‐years; mean: 5.2 years | All patients implanted with transvenous ICD leads; age 57±13 years; 83% male; dual chamber, 7%; CRT, 29% | Sudden change (≥50% compared with chronic values) in long‐term pacing and high‐voltage impedance and/or electrical noise artifacts from rapid, nonphysiological make‐break potentials recorded on the sensing channel. Lead dislocation or perforation or oversensing of noncardiac potentials, not considered indicative of lead integrity failure (ie, electromagnetic interferences), were not considered in this analysis |
| Verlato et al 2013 | Prospective multicentric (12 sites) registry; follow‐up: 2336 patient‐years; mean: 2.4 years | All patients implanted with the prespecified leads; age 67±10 years; 83.5% male; dual chamber, 17.1%; CRT, 67.3% | In the absence of noise‐induced shocks, lead failure evidence was derived from the analysis of signals such as a significant decrease or increase in lead impedance; presence of false nonsustained ventricular tachycardia episodes; significant increase in the sensing integrity counter, which counts the number of sensed, nonphysiological short ventricular intervals near the ICD blanking period; oversensing or undersensing of the normal cardiac electrical activity, as registered on the right ventricular channel; or any change in electrical parameters which may prevent detection or interruption of potentially lethal arrhythmias |
| Liu et al 2014 | Retrospective single‐center study; follow‐up: 19 237 patient‐years; mean: 3.6 years | All ICD recipients; age 68±13 years; 76.9% male; CRT, 39.1% | Electrical malfunction resulting in lead extraction or replacement with a new ICD lead was defined as abnormal pace sense or high‐voltage impedance values, decrease in R‐wave amplitude, increase in pacing threshold necessitating lead replacement; or the presence of electrical noise leading to inappropriate ICD therapy. Leads demonstrating mechanical failure (eg, cable externalization in Riata leads) with normal electrical function were not classified as lead failures in the present study |
| Vollman et al 2014 | Retrospective single‐center study; follow‐up: −2774 patient‐years; mean: 3.7 years | All patients implanted with the prespecified leads; age 62.3±14.9 years; 81.0% male; CRT, 30.1% | A lead failed if ≥1 of the following criteria applied: (1) sudden rise in long‐term pacing or high‐voltage impedance; (2) electrical noise artifacts as manifested by nonphysiological signals on the electrogram or by device diagnostics; (3) failure to sense R‐waves or ineffective electrical therapy due to an apparent structural lead defect |
| Yanagisawa et al 2014 | Retrospective single‐center study; follow‐up: 779 patient‐years; mean: 3.6 years | All patients implanted with the prespecified ICD leads; age 60.2±15.2 years; 73.4% male; CRT, 38.8% | Nonphysiological high‐rate sensing; sudden change of sense and pace, coil impedance out of normal limits, or inappropriate shock due to sensing of electrical noise artifacts were suggestive of lead failure |
| Kramer et al 2015 | Retrospective multicentric (4 sites) study; follow‐up: 8797 patient‐years; median: 3.2 years | All patients implanted with the prespecified leads; age 65 years (IQR 55–74 years); 74% male | Failure definition included ≥1 of the following: (1) impedance outside the labeled normal range for that model; (2) electrical noise manifest as nonphysiological signals on the electrogram or as pulse generator diagnostic data suggesting rapid oversensing, eg, nonphysiological short intervals and/or recurrent nonsustained ventricular tachycardia with intervals usually <220 ms; (3) increase in pacing threshold or decline in R‐wave amplitude necessitating lead replacement; (4) inability to provide effective therapy due to a lead defect; (5) externalized conductor that breached the outer insulation and appeared outside the lead body on fluoroscopy or radiography; and (6) lead dislodgement, except simple dislodgments without an identified fixation mechanism defect. Functional abnormalities, including exit block and physiological oversensing in an electrically intact lead, were not considered as failures |
CRT indicates cardiac resynchronization therapy; HCM, hypertrophic cardiomyopathy; ICD, implantable cardioverter‐defibrillator; IQR, interquartile range; LIA, lead integrity alert; RAFT, Resynchronization/Defibrillation for Ambulatory Heart Failure trial.
Main Events in the Studies Included in the Meta‐Analysis
| Study | Durata | Endotak Reliance | Fidelis | Quattro | Riata | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Failure Rate | Lead‐Years | Incidence 95% CI | Failure Rate | Lead‐Years | Incidence 95% CI | Failure Rate | Lead‐Years | Incidence 95% CI | Failure Rate | Lead‐Years | Incidence 95% CI | Failure Rate | Lead‐Years | Incidence (95% CI) | |
| Kleeman 2007 | — | — | — | 0/1, 0% | 6.6 | 0 (0–36.93) | 0/1, 0% | 0.27 | 0 (0–93.43) | 0/139, 0% | 155.7 | 0 (0–7.25) | 13/199, 6.53% | 167.2 | 7.78 (4.60–12.85) |
| Eckstein 2008 | — | — | — | 1/159, 0.63% | 1017.6 | 0.10 (0.02–0.55) | — | — | — | 3/103, 2.91% | 659.2 | 0.46 (0.15–1.33) | 0/42, 0% | 268.8 | 0 (0–1.41) |
| Hauser 2008 | — | — | — | 2/89, 2.25% | 247.3 | 0.81 (0.22–2.90) | 4/69, 5.80% | 102.1 | 3.92 (1.53–9.64) | 0/70, 0% | 190.9 | 0 (0–1.97) | 1/35, 2.86% | 40.8 | 2.45 (0.43–12.65) |
| Borleffs 2009 | — | — | — | 15/911, 1.65% | 1954.3 | 0.77 (0.47–1.26) | 16/376, 4.26% | 584.1 | 2.74 (1.69–4.40) | 20/322, 6.21% | 1284.8 | 1.56 (1.01–2.39) | 4/183, 2.19% | 504.1 | 0.79 (0.31–2.02) |
| Hauser 2009 | — | — | — | 7/607, 1.15% | 1241 | 0.56 (0.27–1.16) | 72/848, 8.49% | 1923 | 3.74 (2.98–4.69) | 11/1273, 0.86% | 2013 | 0.55 (0.31–0.98) | 1/90, 1.11% | 213 | 0.47 (0.08–2.61) |
| Morrison 2011 | — | — | — | — | — | — | 85/1030, 8.25% | 2952.7 | 2.88 (2.33–3.55) | 23/1641, 1.40% | 5454.3 | 0.42 (0.28–0.63) | — | — | — |
| Parkash 2012 | — | — | — | — | — | — | 45/818, 5.50% | 2658.5 | 1.69 (1.27–2.26) | 2/969, 0.21% | 3149.25 | 0.06 (0.02–0.23) | — | — | — |
| Sung 2012 | — | — | — | 12/2401, 0.50% | 7580 | 0.16 (0.09–0.28) | 379/5073, 7.47% | 18 255 | 2.08 (1.88–2.30) | 28/6091, 0.46% | 18 748 | 0.15 (0.10–0.22) | 47/1403, 3.35% | 7039 | 0.67 (0.5–0.89) |
| Abdelhadi 2013 | — | — | — | — | — | — | — | — | — | 23/1668, 1.38% | 5304.2 | 0.43 (0.29–0.65) | 47/1081, 4.35% | 4205.1 | 1.12 (0.84–1.48) |
| Ellenbogen 2013 | 10/4179=0.24% | 3343.2 | 0.30 (0.16–0.55) | 24/3654, 0.66% | 3288.6 | 0.73 (0.49–1.08) | 24/1556, 1.54% | 2800.8 | 0.86 (0.58–1.27) | — | — | — | 21/1944, 1.08% | 1749.6 | 1.20 (0.79–1.83) |
| Fazal 2013 | — | — | — | — | — | — | 13/113, 11.50% | 499.5 | 2.61 (1.53–4.40) | — | — | — | 13/106, 12.26% | 484.4 | 2.62 (1.57–4.54) |
| Rordorf 2013 | 0/86=0% | 157.7 | 0 (0–2.38) | 7/204, 3.43% | 663 | 1.06 (0.51–2.16) | 29/190, 15.26% | 680.2 | 4.8 (2.98–6.06) | 0/130, 0% | 650 | 0 (0–0.90) | 16/182, 8.79% | 606.6 | 2.64 (1.63–4.25) |
| Verlato 2013 | — | — | — | — | — | — | 27/508, 5.31% | 1437.6 | 1.88 (1.29–2.72) | 2/468, 0.43% | 898.6 | 0.22 (0.06–0.81) | — | — | — |
| Liu 2014 | 18/828=2.17% | 1904.4 | 0.95 (0.60–1.49) | 26/2190, 1.32% | 10 074 | 0.26 (0.18–0.38) | 47/623, 7.54% | 1682.1 | 2.79 (2.11–3.70) | 11/1020, 1.08% | 3570 | 0.31 (0.17–0.55) | 38/627, 6.06% | 2006.4 | 1.89 (1.38–2.59) |
| Vollmann 2014 | — | — | — | — | — | — | 33/429, 7.69% | 1394.3 | 2.37 (1.69–3.31) | 5/276, 1.81% | 1380 | 0.36 (0.15–0.85) | — | — | — |
| Yanagisawa 2014 | — | — | — | — | — | — | 14/75, 18.67% | 360 | 3.89 (2.33–6.42) | 0/131, 0% | 379.9 | 0 (0–1.00) | 1/8, 12.5% | 39.2 | 2.55 (0.45–13.12) |
| Kramer 2015 | 2/445=0.45% | 1311 | 0.15 (0.04–0.55) | 6/389, 1.54% | 1407 | 0.43 (0.20–0.93) | — | — | — | 17/1818, 0.94% | 6079 | 0.28 (0.17–0.45) | — | — | — |
| Pooled Data | 30/5538=0.54% | 6716.3 | 0.45 (0.31–0.64) | 100/10 605, 0.94% | 27 479.3 | 0.36 (0.30–0.44) | 788/11 709, 6.73% | 35 300.2 | 2.23 (2.08–2.39) | 145/16 119, 0.90% | 49 689.3 | 0.29 (0.25–0.34) | 202/5900, 3.42% | 17 323.6 | 1.17 (1.02–1.34) |
Failure rate in percentage of total leads of each model; incidence in percentage per year.
Figure 2Incidence of lead failure.
Results of the Head‐to‐Head Comparison of the 5 Lead Families (Differences in Mean Incidence and Study Heterogeneity)
| Durata | Endotak Reliance | Fidelis | Sprint Quattro | Riata |
|---|---|---|---|---|
| Durata | 0.14 (−0.48 to 0.77); | −3.12 (−5.77 to −0.47); | −0.25 (−0.75 to 0.24); | −1.25 (−2.01 to −0.49); |
| Endotak Reliance | −2.20 (−3.03 to −1.36); | 0.01 (−0.17 to 0.19); | −0.57 (−1.01 to −0.13); | |
| Fidelis | −2.29 (−2.71 to −1.88); | −1.24 (−2.08 to −0.40); | ||
| Sprint Quattro | −0.73 (−1.29 to −0.16); | |||
| Riata |
Figure 3Sensitivity analysis of recalled versus nonrecalled leads.
Figure 4Sensitivity analysis of 7‐ versus ≥8‐French leads.
Figure 5Meta‐regression analysis of head‐to‐head comparisons of the different lead families. Coeff. indicates coefficient; CRT, cardiac resynchronization therapy.
Figure 6Construction and materials of the different lead families.