Literature DB >> 23813451

Passive-fixation lead failure rates and long-term patient mortality in subjects implanted with Sprint Fidelis electrodes.

Dirk Vollmann1, Sandra Woronowicz, Lukasz Kmiec, Klaus Jung, Dieter Zenker, Joachim Seegers, Samuel Sossalla, Marc Dorenkamp, Christian Sohns, Lars Lüthje, Gerd Hasenfuss, Markus Zabel.   

Abstract

AIMS: To evaluate passive-fixation lead failure rates and long-term patient survival in subjects implanted with Sprint Fidelis electrodes. METHODS AND
RESULTS: We identified 748 subjects who received a Sprint Fidelis (n = 429; Medtronic models 6948: 94.8%, 6949: 2.6%, 6930: 1.9%, 6931: 0.7%) or a Sprint 'non-Fidelis' implantable cardioverter defibrillator lead (n = 319, Medtronic models 6944: 68.6%, 6947: 17.9%, 6942: 7.8%, 6943: 3.4%, 6945: 2.2%) at our centre between 1998 and 2008. Kaplan-Meier patient survival was lower in the Fidelis group than in the Control cohort (68.4 vs. 77.0% at 5 years, P = 0.0061), but multivariate analyses revealed no significant association between mortality and implanted lead type. Passive-fixation lead failure rate at 5 years was 14.4% (95% confidence interval (CI) [9.2, 19.3]) in the Fidelis (n = 414) group and 1.8% (95% CI [0.0-3.8]) in the Control (n = 241) cohort (P < 0.001 upon multivariate comparison).
CONCLUSION: Failure rates of passive-fixation Sprint Fidelis leads are increased and similar to those previously reported for active-fixation Fidelis electrodes. Despite the elevated risk for lead failure and its potential sequelae, the Sprint Fidelis has no obvious impact on long-term mortality.

Entities:  

Keywords:  ICD; Lead failure; Mortality; Passive fixation

Mesh:

Year:  2013        PMID: 23813451     DOI: 10.1093/europace/eut185

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  1 in total

Review 1.  Transvenous Implantable Cardioverter-Defibrillator (ICD) Lead Performance: A Meta-Analysis of Observational Studies.

Authors:  Rui Providência; Daniel B Kramer; Dominic Pimenta; Girish G Babu; Laura A Hatfield; Adam Ioannou; Jan Novak; Robert G Hauser; Pier D Lambiase
Journal:  J Am Heart Assoc       Date:  2015-10-30       Impact factor: 5.501

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.