Literature DB >> 26510985

Modified HPLC-ESI-MS Method for Glycated Hemoglobin Quantification Based on the IFCC Reference Measurement Procedure and Its Application for Quantitative Analyses in Clinical Laboratories of China.

Zhixin Song1, Baoyuan Xie1, Huaian Ma2, Rui Zhang3, Pengfei Li3, Lihong Liu3, Yuhong Yue3, Jianping Zhang3, Qing Tong3, Qingtao Wang4.   

Abstract

BACKGROUND: The level of glycated hemoglobin (HbA1c ) has been recognized as an important indicator of long-term glycemic control. However, the HbA1c measurement is not currently included as a diagnostic determinant in China. Current study aims to assess a candidate modified International Federation of Clinical Chemistry reference method for the forthcoming standardization of HbA1c measurements in China.
METHODS: The HbA1c concentration was measured using a modified high-performance liquid chromatography-electrospray ionization-mass spectrometry (HPLC-ESI-MS) method. The modified method replaces the propylcyanide column with a C18 reversed-phase column, which has a lower cost and is more commonly used in China, and uses 0.1% (26.5 mmol/l) formic acid instead of trifluoroacetic acid. Moreover, in order to minimize matrix interference and reduce the running time, a solid-phase extraction was employed. The discrepancies between HbA1c measurements using conventional methods and the HPLC-ESI-MS method were clarified in clinical samples from healthy people and diabetic patients. Corresponding samples were distributed to 89 hospitals in Beijing for external quality assessment.
RESULTS: The linearity, reliability, and accuracy of the modified HPLC-ESI-MS method with a shortened running time of 6 min were successfully validated. Out of 89 hospitals evaluated, the relative biases of HbA1c concentrations were < 8% for 74 hospitals and < 5% for 60 hospitals. Compared with other conventional methods, HbA1c concentrations determined by HPLC methods were similar to the values obtained from the current HPLC-ESI-MS method.
CONCLUSION: The HPLC-ESI-MS method represents an improvement over existing methods and provides a simple, stable, and rapid HbA1c measurement with strong signal intensities and reduced ion suppression.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  diabetes; external quality assessment; glycated hemoglobin; high-performance liquid chromatography; mass spectrometry

Mesh:

Substances:

Year:  2015        PMID: 26510985      PMCID: PMC6807171          DOI: 10.1002/jcla.21879

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  16 in total

Review 1.  Specimen materials, target values and commutability for external quality assessment (proficiency testing) schemes.

Authors:  W Greg Miller
Journal:  Clin Chim Acta       Date:  2003-01       Impact factor: 3.786

2.  Development of an isotope dilution mass spectrometry assay for HbA1c based on enzyme-cleaved peptide analysis.

Authors:  Ken Iguchi; Toyofumi Nakanishi; Ayako Miyazaki; Akira Shimizu; Akira Ota
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2004-06-25       Impact factor: 3.205

Review 3.  Carbohydrate metabolism in non-insulin-dependent diabetes mellitus.

Authors:  S Dinneen; J Gerich; R Rizza
Journal:  N Engl J Med       Date:  1992-09-03       Impact factor: 91.245

4.  Why commutability matters.

Authors:  W Greg Miller; Gary L Myers; Robert Rej
Journal:  Clin Chem       Date:  2006-04       Impact factor: 8.327

Review 5.  Traceability in laboratory medicine.

Authors:  Hubert W Vesper; Linda M Thienpont
Journal:  Clin Chem       Date:  2009-04-09       Impact factor: 8.327

6.  Reference materials and commutability.

Authors:  Hubert W Vesper; W Gregory Miller; Gary L Myers
Journal:  Clin Biochem Rev       Date:  2007-11

7.  Approved IFCC reference method for the measurement of HbA1c in human blood.

Authors:  Jan-Olof Jeppsson; Uwe Kobold; John Barr; Andreas Finke; Wieland Hoelzel; Tadao Hoshino; Kor Miedema; Andrea Mosca; Pierluigi Mauri; Rita Paroni; Linda Thienpont; Masao Umemoto; Cas Weykamp
Journal:  Clin Chem Lab Med       Date:  2002-01       Impact factor: 3.694

Review 8.  Glycohemoglobin: a primary predictor of the development or reversal of complications of diabetes mellitus.

Authors:  U Krishnamurti; M W Steffes
Journal:  Clin Chem       Date:  2001       Impact factor: 8.327

9.  Quantitative targeted biomarker assay for glycated haemoglobin by multidimensional LC using mass spectrometric detection.

Authors:  Estela del Castillo; Maria Montes-Bayón; Elena Añón; Alfredo Sanz-Medel
Journal:  J Proteomics       Date:  2010-08-05       Impact factor: 4.044

10.  Modification of the IFCC reference measurement procedure for determination of HbA 1c by HPLC-ESI-MS.

Authors:  Patricia Kaiser; Theodorus Akerboom; László Dux; Hans Reinauer
Journal:  Ger Med Sci       Date:  2006-09-21
View more
  1 in total

1.  An improved reference method for serum cations measurement by ion chromatography.

Authors:  Jihua Zou; Min Shen; Man Zhang; Minmin Tu; Rui Feng; Ying Yan; Bingde Zou
Journal:  J Clin Lab Anal       Date:  2018-03-30       Impact factor: 2.352