RATIONALE: We reported previously, in porcine coronary arteries, that the novel biodegradable PowerStent Absorb paclitaxel-eluting stent had improved and sustained structural strength and functional performance at one month post-implantation.
OBJECTIVE: To report the stent performance at 6-month follow-up. METHODS AND
RESULTS: Six PowerStent Absorb and six TAXUS stents were randomly implanted in the left anterior descending and right coronary arteries of six Tibet miniature pigs. Quantitative coronary angiography (QCA) and intravascular ultrasound (IVUS) images were obtained at the time of implantation (T0) and at 6 months (T6). Two animals were sacrificed at T6 for histopathological evaluation. At T6, QCA showed that the mean luminal vascular diameter (mLD) between the PowerStent and the TAXUS stents were similar (2.36 ± 0.38 vs. 2.61 ± 0.31, respectively). Based on the IVUS analysis, the mLD and the mean lumen cross-sectional area (mCSA) in the PowerStent-treated arteries were similar between T0 and T6 (mLD: 2.74 ± 0.13 vs. 2.70 ± 0.20 and mCSA: 6.81 ± 0.62 mm2 vs. 6.68 ± 0.94 mm2). Histopathology showed that the PowerStent stents were well apposed to the vessel wall with no recoil, strut fracture and thrombus formation. The stents were fully covered with a layer of endothelial cells.
CONCLUSIONS: At six-month post-implantation, the PowerStent Absorb stents maintained their structural strength and functional performance. The development of restenosis was controlled, no stent thrombosis was observed and the stents were fully re-endothelialized. These results suggest the PowerStent Absorb stent is safe and effective for up to 6 months when implanted in porcine coronary arteries.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26502644     DOI: 10.1166/jbn.2015.2102

Source DB:  PubMed          Journal:  J Biomed Nanotechnol        ISSN: 1550-7033            Impact factor:   4.099


  2 in total

1.  Effect of novel bioresorbable scaffold composed of poly-L-lactic acid and amorphous calcium phosphate nanoparticles on inflammation and calcification of surrounding tissues after implantation.

Authors:  Gaoke Feng; Chaoshi Qin; Xin Yi; Jinggang Xia; Jingjing Chen; Xiyu Chen; Tao Chen; Xuejun Jiang
Journal:  J Mater Sci Mater Med       Date:  2018-07-17       Impact factor: 3.896

Review 2.  Surface engineering at the nanoscale: A way forward to improve coronary stent efficacy.

Authors:  Aleena Mary Cherian; Shantikumar V Nair; Vijayakumar Maniyal; Deepthy Menon
Journal:  APL Bioeng       Date:  2021-06-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.