Preliminary evaluation of image quality in a new clinical ToF-PET/MR scanner.

Gaspar Delso1, Mehdi Khalighi1, Marlena Hofbauer2, Miguel Porto2, Patrick Veit-Haibach2, Gustav von Schulthess2.   

Abstract

Entities:  

Year:  2014        PMID: 26501629      PMCID: PMC4544605          DOI: 10.1186/2197-7364-1-S1-A41

Source DB:  PubMed          Journal:  EJNMMI Phys        ISSN: 2197-7364


× No keyword cloud information.
A new ToF-PET/MR scanner has recently been deployed at the University Hospital of Zurich for clinical evaluation purposes. The goal of the present work was to compare the image quality obtained on this system with that of equivalent state-of-the-art standalone scanners. The GE Signa PET/MR is a whole-body scanner combining a 3T wide-bore MR system with a 25 cm PET detector ring based on SiPM technology and mounted on a customized radiofrequency coil. MR image quality of this system was tested by scanning two volunteers: One with a brain protocol and another with a whole-body protocol. The same scans were repeated before and after PET detectors were installed on the system. A baseline measurement was acquired on a Discovery 750w scanner with the same volunteers. PET image quality was tested on patients referred for an FDG PET/CT scan. After 15-20 minutes of PET/CT, the patients were moved to the PET/MR, where an equivalent acquisition was performed. Data were reviewed by nuclear medicine and radiology physicians. As seen in Figure 1, MR image quality in the brain is identical between PET/MR (pre- and post-insert) and MR750w. In whole-body scans, shading on the images due to different receive coil sensitivity were observed. These are explained by the replacement of the bed-mounted posterior array coil by a central matrix array at the scanner iso-center. Figure 2 illustrates the differences in PET, showing equivalent or better quality in the PET/MR data, despite latter acquisition, justified by the increased sensitivity of the scanner.
Figure 1

Coronal 2D MERGE views of a healthy volunteer scanned on a Discovery 750w (left), on the PET/MR before the installation of the PET insert (center) and on the same scanner once the insert was in place and online (right).

Figure 2

Three-dimensional maximum intensity projection of a PET/CT acquisition (left) and posterior PET/MR acquisition (right) of the same patient.

Coronal 2D MERGE views of a healthy volunteer scanned on a Discovery 750w (left), on the PET/MR before the installation of the PET insert (center) and on the same scanner once the insert was in place and online (right). Three-dimensional maximum intensity projection of a PET/CT acquisition (left) and posterior PET/MR acquisition (right) of the same patient. Our results show that the image quality of clinical MR sequences is not affected by the presence of the PET detector insert. PET image quality shows great promise, but will require further testing to confirm this quantitatively.
  4 in total

1.  Simulation study of the second-generation MR-compatible SPECT system based on the inverted compound-eye gamma camera design.

Authors:  Xiaochun Lai; Ling-Jian Meng
Journal:  Phys Med Biol       Date:  2018-02-12       Impact factor: 3.609

2.  Variations in PET/MRI Operations: Results from an International Survey Among 39 Active Sites.

Authors:  Wolfgang Peter Fendler; Johannes Czernin; Ken Herrmann; Thomas Beyer
Journal:  J Nucl Med       Date:  2016-08-11       Impact factor: 10.057

3.  Impact of time-of-flight PET on quantification accuracy and lesion detection in simultaneous 18F-choline PET/MRI for prostate cancer.

Authors:  Urs J Muehlematter; Hannes W Nagel; Anton Becker; Julian Mueller; Kerstin N Vokinger; Felipe de Galiza Barbosa; Edwin E G T Ter Voert; Patrick Veit-Haibach; Irene A Burger
Journal:  EJNMMI Res       Date:  2018-05-31       Impact factor: 3.138

4.  Development of clinical simultaneous SPECT/MRI.

Authors:  Brian F Hutton; Michele Occhipinti; Andre Kuehne; Domokos Máthé; Noémi Kovács; Helmar Waiczies; Kjell Erlandsson; Debora Salvado; Marco Carminati; Giovanni L Montagnani; Susan C Short; Luisa Ottobrini; Pieter van Mullekom; Claudio Piemonte; Tamas Bukki; Zoltan Nyitrai; Zoltan Papp; Kalman Nagy; Thoralf Niendorf; Irene de Francesco; Carlo Fiorini
Journal:  Br J Radiol       Date:  2017-03-07       Impact factor: 3.039

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.