Clinical evaluation of the Shin-Nippon SRW-5000 autorefractor in adults: an update.

Edward A H Mallen1, Bernard Gilmartin2, James S Wolffsohn2, Sei-ichi Tsujimura3.   

Abstract

PURPOSE: The Shin-Nippon SRW-5000 is an open view autorefractor that superseded the Canon R-1 autorefractor in the mid-1990 s and has been used widely in optometry and vision science laboratories. It has been used to measure refractive error, accommodation responses both statically and dynamically, off-axis refractive error, and adapted to measure pupil size. This paper presents an overview of the original 2001 clinical evaluation of the SRW-5000 in adults (Mallen et al., Ophthal Physiol Opt 2001; 21: 101) and provides an update on the use and modification of the instrument since the original publication. RECENT
FINDINGS: The SRW-5000 instrument, and the family of devices which followed, have shown excellent validity, repeatability, and utility in clinical and research settings. The instruments have also shown great potential for increased research functionality following a number of modifications.
SUMMARY: The SRW-5000 and its derivatives have been, and continue to be, of significant importance in our drive to understand myopia progression, myopia control techniques, and oculomotor function in human vision.
© 2015 The Authors Ophthalmic & Physiological Optics © 2015 The College of Optometrists.

Entities:  

Keywords:  Grand Seiko; Shin-Nippon SRW-5000; accommodation measurement; open-view autorefractor

Mesh:

Year:  2015        PMID: 26497294     DOI: 10.1111/opo.12254

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  5 in total

Review 1.  Clinical application of accommodating intraocular lens.

Authors:  You-Ling Liang; Song-Bai Jia
Journal:  Int J Ophthalmol       Date:  2018-06-18       Impact factor: 1.779

2.  Comparison of cycloplegic refraction between Grand Seiko autorefractor and Retinomax autorefractor in the Vision in Preschoolers-Hyperopia in Preschoolers (VIP-HIP) Study.

Authors:  Gui-Shuang Ying; Maureen G Maguire; Marjean Taylor Kulp; Elise Ciner; Bruce Moore; Maxwell Pistilli; Rowan Candy
Journal:  J AAPOS       Date:  2017-05-18       Impact factor: 1.220

3.  Validation of the PowerRef 3 for Measuring Accommodation: Comparison With the Grand Seiko WAM-5500A Autorefractor.

Authors:  Alyssa M Gehring; Jennifer X Haensel; Molly K Curtiss; Tawna L Roberts
Journal:  Transl Vis Sci Technol       Date:  2022-10-03       Impact factor: 3.048

4.  Six Year Refractive Change among White Children and Young Adults: Evidence for Significant Increase in Myopia among White UK Children.

Authors:  Sara J McCullough; Lisa O'Donoghue; Kathryn J Saunders
Journal:  PLoS One       Date:  2016-01-19       Impact factor: 3.240

5.  Comparison of Two Wavefront Autorefractors: Binocular Open-Field versus Monocular Closed-Field.

Authors:  Gonzalo Carracedo; Carlos Carpena-Torres; Laura Batres; Maria Serramito; Anahí Gonzalez-Bergaz
Journal:  J Ophthalmol       Date:  2020-01-03       Impact factor: 1.909