Bobby L Boyanton1, Elizabeth M Darnell2, Anne E Prada2, Dana M Hansz2, Barbara Robinson-Dunn3.
Abstract
The Lyra Direct strep assay was compared to culture for its ability to detect Streptococcus group A and β-hemolytic groups C/G using rapid antigen-negative pharyngeal specimens (n = 161). The Lyra assay correctly detected all β-hemolytic streptococci (group A, n = 19; group C/G, n = 5). In batch mode, the Lyra assay reduced intralaboratory turnaround time by 60% (18.1 h versus 45.0 h) but increased hands-on time by 96% (3 min 16 s versus 1 min 40 s per specimen).Entities:
Mesh:
Year: 2015 PMID: 26491174 PMCID: PMC4702764 DOI: 10.1128/JCM.02405-15
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948