Evaluation of the performance of the OneTouch Select Plus blood glucose test system against ISO 15197:2013.

Steven Setford1, Antony Smith1, David McColl1, Mike Grady1, Krisna Koria1, Hilary Cameron1.   

Abstract

OBJECTIVES: Assess laboratory and in-clinic performance of the OneTouch Select(®) Plus test system against ISO 15197:2013 standard for measurement of blood glucose.
METHODS: System performance assessed in laboratory against key patient, environmental and pharmacologic factors. User performance was assessed in clinic by system-naïve lay-users. Healthcare professionals assessed system accuracy on diabetes subjects in clinic.
RESULTS: The system demonstrated high levels of performance, meeting ISO 15197:2013 requirements in laboratory testing (precision, linearity, hematocrit, temperature, humidity and altitude). System performance was tested against 28 interferents, with an adverse interfering effect only being recorded for pralidoxime iodide. Clinic user performance results fulfilled ISO 15197:2013 accuracy criteria. Subjects agreed that the color range indicator clearly showed if they were low, in-range or high and helped them better understand glucose results.
CONCLUSION: The system evaluated is accurate and meets all ISO 15197:2013 requirements as per the tests described. The color range indicator helped subjects understand glucose results and supports patients in following healthcare professional recommendations on glucose targets.

Entities:  

Keywords:  ISO 15197:2013; blood glucose measurement; color range indicator; self-monitoring; test strip

Mesh:

Substances:

Year:  2015        PMID: 26488139     DOI: 10.1586/17434440.2015.1102049

Source DB:  PubMed          Journal:  Expert Rev Med Devices        ISSN: 1743-4440            Impact factor:   3.166


  3 in total

1.  Evidence From a Long-Term, Systematic Post-Market Surveillance Program: Clinical Performance of a Hematocrit-Insensitive Blood Glucose Test Strip.

Authors:  Steven Setford; Stuart Phillips; Mike Grady
Journal:  J Diabetes Sci Technol       Date:  2019-02-07

2.  Gender Differences in Heart Rate Variability Among Individuals Undergoing Regular Resistance Training Preliminary observations.

Authors:  Ellen S V Ramalho; Estevam L Souza-Junior; Marciane Magnani; Valdir A Braga
Journal:  Sultan Qaboos Univ Med J       Date:  2017-06-20

3.  Patient and Health-care Professional Satisfaction with a New, Simple, High Accuracy Blood Glucose Meter with Color Range Indicator.

Authors:  Laurence B Katz; Mike Grady; Lorna Stewart; Hilary Cameron; Pamela A Anderson; Anish Desai
Journal:  Indian J Endocrinol Metab       Date:  2017 Mar-Apr
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.