Doppler color flow mapping of simulated in vitro regurgitant jets: evaluation of the effects of orifice size and hemodynamic variables.

I A Simpson1, L M Valdes-Cruz, D J Sahn, A Murillo, T Tamura, K J Chung.   

Abstract

The spatial distribution of simulated regurgitant jets imaged by Doppler color flow mapping was evaluated under constant flow and pulsatile flow conditions. Jets were simulated through latex tubings of 3.2, 4.8, 6.35 and 7.9 mm by varying flow rates from 137 to 1,260 cc/min. Color jet area was linearly related to flow rate at each orifice (r = 0.96, SEE = 3.4; r = 0.99, SEE = 1.6; r = 0.97, SEE = 2.3; r = 0.97, SEE = 3.2, respectively), but significantly higher flow rates were required to maintain the same maximal spatial distribution of the jet at the larger regurgitant orifices. Constant flow jets were also simulated through needle orifices of 0.2, 0.5 and 1 mm, with a known total volume (5 cc) injected at varying flow rates and with differing absolute volumes injected at the same flow rate (0.2, 1.0 and 2.0 cc/s, respectively). Again, maximal color jet area was linearly related to flow rate at each orifice (r = 0.97, SEE = 2.3; r = 0.97, SEE = 2.4; r = 0.92, SEE = 3.9, respectively), but was not related to the absolute volume of regurgitation. Color encoding of regurgitant jets on Doppler color flow maps was demonstrated to be highly dependent on velocity and, hence, driving pressure, such that color encoding was obtained from a constant flow jet injected at a velocity of 4 m/s through an orifice of 0.04 mm diameter with flow rates as low as 0.008 cc/s. Mitral regurgitant jets were also simulated in a physiologic in vitro pulsatile flow model through three prosthetic valves with known regurgitant orifice sizes (0.2, 0.6 and 2.0 mm2). For each regurgitant orifice size, color jet area at each was linearly related to a regurgitant pressure drop (r = 0.98, SEE = 0.15; r = 0.97, SEE = 0.20; r = 0.97, SEE = 0.23, respectively), regurgitant stroke volume (r = 0.77, SEE = 0.55; r = 0.94, SEE = 0.30; r = 0.91, SEE = 0.41, respectively) and peak regurgitant flow rate (r = 0.98, SEE = 0.16; r = 0.97, SEE = 0.21; r = 0.93, SEE = 0.37, respectively), but the spatial distribution of the regurgitant jets was most highly dependent on the regurgitant pressure drop. Jet kinetic energy calculated from the summation of the individual pixel intensities integrated over the jet area was closely related to driving pressure (r = 0.84), but integration of the power mode area times pixel intensities provided the best estimation of regurgitant stroke volume (r = 0.80).(ABSTRACT TRUNCATED AT 400 WORDS)

Entities:  

Mesh:

Year:  1989        PMID: 2647815     DOI: 10.1016/0735-1097(89)90284-2

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  9 in total

Review 1.  Colour Doppler flow mapping.

Authors:  I A Simpson; A J Camm
Journal:  BMJ       Date:  1990-01-06

2.  Three-dimensional ultrasound imaging model of mitral valve regurgitation: design and evaluation.

Authors:  Stephen H Little; Stephen R Igo; Marti McCulloch; Craig J Hartley; Yukihiko Nosé; William A Zoghbi
Journal:  Ultrasound Med Biol       Date:  2008-02-06       Impact factor: 2.998

3.  Computational simulations of mitral regurgitation quantification using the flow convergence method: comparison of hemispheric and hemielliptic formulae.

Authors:  J Hopmeyer; E Whitney; D A Papp; M S Navathe; R A Levine; Y H Kim; A P Yoganathan
Journal:  Ann Biomed Eng       Date:  1996 Sep-Oct       Impact factor: 3.934

4.  How to quantitate valve regurgitation by echo Doppler techniques. British Society of Echocardiography.

Authors:  I A Simpson; M A de Belder; A Kenny; M Martin; P Nihoyannopoulos
Journal:  Br Heart J       Date:  1995-05

5.  Doppler ultrasound and the apparently normal heart.

Authors:  A B Houston
Journal:  Br Heart J       Date:  1993-02

Review 6.  New technology in echocardiography I: Doppler techniques.

Authors:  W N McDicken; P R Hoskins; C M Moran; G R Sutherland
Journal:  Heart       Date:  1996-06       Impact factor: 5.994

7.  Influence of colour Doppler echocardiography on the ultrasonic assessment of congenital heart disease: a prospective study.

Authors:  J P Gnanapragasam; A B Houston; W B Doig; R Fraser; S Lilley; E Murtagh; G Olafsson
Journal:  Br Heart J       Date:  1991-09

8.  Cardiovascular manifestations of Marfan's syndrome: improved evaluation by transoesophageal echocardiography.

Authors:  I A Simpson; M A de Belder; T Treasure; A J Camm; C W Pumphrey
Journal:  Br Heart J       Date:  1993-02

Review 9.  Mitral regurgitation after transcatheter aortic valve replacement.

Authors:  Francesco Nappi; Antonio Nenna; Irina Timofeeva; Christos Mihos; Federico Gentile; Massimo Chello
Journal:  J Thorac Dis       Date:  2020-05       Impact factor: 2.895

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.