Literature DB >> 26475731

Intra- and interobserver reliability and intra-catheter reproducibility using frequency domain optical coherence tomography for the evaluation of morphometric stent parameters and qualitative assessment of stent strut coverage.

Lisbeth Antonsen1, Per Thayssen2, Anders Junker2, Karsten Tange Veien2, Henrik Steen Hansen2, Knud Nørregaard Hansen2, Mikkel Hougaard2, Lisette Okkels Jensen2.   

Abstract

PURPOSE: Frequency-domain optical coherence tomography (FD-OCT) is a high-resolution imaging tool (~10-15 μm), which enables near-histological in-vivo images of the coronary vessel wall. The use of the technique is increasing, both for research- and clinical purposes. This study sought to investigate the intra- and interobserver reliability, as well as the intra-catheter reproducibility of quantitative FD-OCT-assessment of morphometric stent parameters and qualitative FD-OCT-evaluation of strut coverage in 10 randomly selected 6-month follow-up Nobori® biolimus-eluting stents (N-BESs).
METHODS: Ten N-BESs (213 cross sectional areas (CSAs) and 1897 struts) imaged with OCT 6 months post-implantation were randomly selected and analyzed by 2 experienced analysts, and the same 10 N-BESs were analyzed by one of the analysts 3 months later. Further, 2 consecutive pullbacks randomly performed in another 10 N-BESs (219 CSAs and 1860 struts) were independently assessed by one of the analysts.
RESULTS: The intraobserver variability with regard to relative difference of mean luminal area and mean stent area at the CSA-level was very low: 0.1%±1.4% and 0.5%±3.2%. Interobserver variability also proved to be low: -2.1%±3.3% and 2.1%±4.6%, and moreover, very restricted intra-catheter variation was observed: 0.02%±6.8% and -0.18%±5.2%. The intraobserver-, interobserver- and intra-catheter reliability for the qualitative evaluation of strut coverage was found to be: kappa (κ)=0.91 (95% confidence interval (CI): 0.88-0.93, p<0.01), κ=0.88 (95% CI: 0.85-0.91, p<0.01), and κ=0.73 (95% CI: 0.68-0.78, p<0.01), respectively.
CONCLUSIONS: FD-OCT is a reproducible and reliable imaging tool for quantitative evaluation of stented coronary segments, and for qualitative assessment of strut coverage.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Optical coherence tomography; Reproducibility; Stent morphometry; Stent strut coverage

Mesh:

Substances:

Year:  2015        PMID: 26475731     DOI: 10.1016/j.carrev.2015.08.010

Source DB:  PubMed          Journal:  Cardiovasc Revasc Med        ISSN: 1878-0938


  1 in total

1.  Reproducibility of optical coherence tomography in vein grafts used for coronary revascularization.

Authors:  Øystein Pettersen; Elżbieta Pociask; Krzysztof P Malinowski; Magdalena Slezak; Knut Hegbom; Rune Wiseth; Dag Ole Nordhaug
Journal:  Cardiol J       Date:  2018-11-16       Impact factor: 2.737

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.