Donald I Galen1,2.
Abstract
BACKGROUND: Uterine fibroids occur singly or as multiple benign tumors originating in the myometrium. Because they vary in size and location, the approach and technique for their identification and surgical management vary. Reference images, such as ultrasound images, magnetic resonance images, and sonohystograms, do not provide real-time intraoperative findings.Entities:
Mesh:
Year: 2015 PMID: 26471917 PMCID: PMC4608284 DOI: 10.1186/s12938-015-0086-5
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Fig. 1Surgeon’s view of real-time operative screens (from left-to-right): laparoscopic view of radiofrequency ablation handpiece inserted into a large intramural fibroid, Acessa Guidance screens, and Acessa System screen prior to deployment of electrode array
Fig. 2Acessa Guidance screens. (Left) Laparoscopic ultrasound view of Acessa handpiece tip entering fibroid from left to right within the two (purple) virtual tracking lines of the target zone; (right) computer-generated view of the relationship of the ultrasound beam and trajectory of the handpiece tip into uterus and fibroid
User (physician) preference testing: summary response statistics
| Subjects, N | Mean scorea | |
|---|---|---|
| Targeting statements | ||
| The transducer and handpiece dynamic animation were helpful in targeting the fibroid | 5 | 5 ± 0 |
| The use of TAG enabled the ability to reach the fibroid faster | 5 | 5 ± 0 |
| Knowing the handpiece trajectory helped to target the fibroid | 5 | 5 ± 0 |
| The transducer and handpiece dynamic animation helped visualize the positions of the actual transducer and handpiece in the cavity | 5 | 5 ± 0 |
| The use of TAG provided confidence when targeting the fibroid | 5 | 5 ± 0 |
| Feature statements | ||
| The center tip-to-plane meter was useful in staying in plane | 4 | 4.75 ± 0.5 |
| The double green ring in the trajectory was useful in determining the location of the tip | 5 | 3.6 ± 0.5 |
| The purple Targeting Zone was useful in determining where the tip intersected the ultrasound plane | 5 | 5 ± 0 |
| While entering the fibroid and deploying the electrodes, it was easier to visualize the electrodes if the targeting zone was turned off | 5 | 4.2 ± 0.8 |
| The messages below the two windows provided useful information | 5 | 3.8 ± 0.4 |
TAG targeting animation guidance
aStatements were evaluated according to the following scoring system: (1) strongly disagree, (2) disagree, (3) neither disagree nor agree, (4) agree, (5), strongly agree