Lead Integrity Alert Is Useful for Assessment of Performance of Biotronik Linox Leads.

Christian Steinberg1, Gareth J Padfield1, Edna Hahn1, Sheila Flavelle1, Cheryl McILROY1, Otto VAN Bremen1, John A Yeung-Lai-Wah1, Charles R Kerr1, Marc W Deyell1, Stanley K Tung1, Jason G Andrade1, Matthew T Bennett1, Jamil G Bashir1, Andrew D Krahn1, Santabhanu Chakrabarti1.   

Abstract

INTRODUCTION: Medtronic's Lead Integrity Alert (LIA) software algorithm is useful for detecting abnormal parameters across various ICD-lead families. However, its utility in the assessment of the Biotronik Linox™ family of high-voltage (HV) leads is unknown.
METHODS: We conducted a retrospective cohort study to assess the performance of the LIA algorithm to detect abnormalities and lead failure in Linox ICD-leads. All LIA-enabled Medtronic devices connected to an active Linox lead were included. The alerts were adjudicated by 2 blinded electrophysiologists and correlated with clinical data.
RESULTS: Between 2008 and 2012, data from 208 patients with 564 patient-years of follow-up were available for analysis. The median follow-up duration was 32 (IQR 21-41 months). Twenty-one LIA triggers were noted in 20 different patients. The median delay until a positive LIA was 32 months (IQR 21-41 months) postimplant with a 5-year lead survival free from LIA of 76%. Ninety-five percent (19/20) LIA alerts were true lead failures. The most common LIA triggers were short V-V intervals (85%) and nonsustained ventricular tachycardia (85%). Abrupt changes of the ICD-lead impedance occurred in 5/20 triggers. Inappropriate ICD-shocks were strongly associated with a positive LIA (30% vs. 7.4%; P = 0.006). Of the explanted Linox leads 53% had visible abnormalities. The sensitivity, specificity, and positive predictive value for lead failure in the presence of a LIA trigger were 87%, 99.5%, and 95.2%, respectively.
CONCLUSIONS: A positive LIA trigger in Biotronik Linox ICD-leads is highly predictive of lead failure. LIA is useful in ongoing surveillance of lead performance.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  ICD-lead, lead integrity alert; Linox; defibrillator; lead failure; shock; sudden death

Mesh:

Year:  2015        PMID: 26471861     DOI: 10.1111/jce.12832

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  2 in total

Review 1.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08

Review 2.  Externalized conductors and insulation failure in Biotronik defibrillator leads: History repeating or a false alarm?

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Clin Cases       Date:  2017-02-16       Impact factor: 1.337