Effect of using a Planecta™ port with a three-way stopcock on the natural frequency of blood pressure transducer kits.

Shigeki Fujiwara1, Keiichi Tachihara2, Satoshi Mori2, Kentaro Ouchi2, Chizuko Yokoe3, Uno Imaizumi3, Yoshinari Morimoto4, Yoichiro Miki5, Izumi Toyoguchi6, Kazu-Ichi Yoshida3, Takeshi Yokoyama2.   

Abstract

Blood pressure transducer kits are equipped with two types of Planecta™ ports-the flat-type Planecta™ port (FTP) and the Planecta™ port with a three-way stopcock (PTS). We reported that FTP application decreased the natural frequency of the kits. However, Planecta™ is an invaluable tool as it prevents infection, ensures technical simplicity, and excludes air. Hence, an ideal Planecta™ port that does not decrease the frequency characteristics is required. As a first step in this direction, we aimed to assess the influence of PTSs on the natural frequency of blood transducer kits. A DTXplus transducer kit (DT4812J; Argon Medical Devices, TX, USA) was used along with ≥1 PTSs (JMS, Hiroshima, Japan), and the frequency characteristics were assessed. The natural frequency and damping coefficient of each kit were obtained by using frequency characteristics analysis software, and these parameters were evaluated by plotting them on Gardner's chart. Regardless of whether one or two PTSs were inserted, the natural frequency of the kits only slightly decreased (from 42.5 to 41.1 Hz, when 2 PTSs were used). Thus, the frequency characteristics of the kits with PTSs were adequate for pressure monitoring. The insertion of ≥2 FTPs in pressure transducer kits should be avoided, as they markedly decrease the natural frequency and lead to underdamping. However, the effect of PTS insertion in pressure transducer kits on the frequency characteristics is minimal. Thus, we found that the use of PTS markedly improved the frequency characteristics as compared to the use of FTP.

Entities:  

Keywords:  Blood pressure transducer kits; Frequency characteristics; Natural frequency; Planecta™ with a three-way stopcock

Mesh:

Year:  2015        PMID: 26467334     DOI: 10.1007/s10877-015-9795-3

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  7 in total

1.  Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems.

Authors:  Stefano Romagnoli; Salvatore Mario Romano; Sergio Bevilacqua; Chiara Lazzeri; Gian Franco Gensini; Carlo Pratesi; Diego Quattrone; Daniele Dini; Angelo Raffaele De Gaudio
Journal:  J Crit Care       Date:  2010-10-30       Impact factor: 3.425

2.  Recommendation of a clinical impulse response analysis for catheter calibration-dumping coefficient and natural frequency are incomplete parameters for clinical evaluation.

Authors:  Hiroaki Watanabe; Shin-ichi Yagi; Akiyoshi Namiki
Journal:  J Clin Monit Comput       Date:  2006-03-07       Impact factor: 2.502

3.  Evaluation of dynamic performance in liquid-filled catheter systems for measuring invasive blood pressure.

Authors:  M Todorovic; E W Jensen; C Thøgersen
Journal:  Int J Clin Monit Comput       Date:  1996-08

4.  Microbial contamination of arterial infusions used for hemodynamic monitoring: a randomized trial of contamination with sampling through conventional stopcocks versus a novel closed system.

Authors:  S Crow; S A Conrad; C Chaney-Rowell; J W King
Journal:  Infect Control Hosp Epidemiol       Date:  1989-12       Impact factor: 3.254

5.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

6.  An evaluation of blood pressure measurement.

Authors:  W B Runciman; A J Rutten; A H Ilsley
Journal:  Anaesth Intensive Care       Date:  1981-11       Impact factor: 1.669

7.  Effect of planecta and ROSE™ on the frequency characteristics of blood pressure-transducer kits.

Authors:  Shigeki Fujiwara; Yoshifumi Kawakubo; Satoshi Mori; Keiichi Tachihara; Izumi Toyoguchi; Takeshi Yokoyama
Journal:  J Clin Monit Comput       Date:  2014-12-17       Impact factor: 2.502

  7 in total
  3 in total

1.  Influence of the marvelous™ three-way stopcock on the natural frequency and damping coefficient in blood pressure transducer kits.

Authors:  Shigeki Joseph Luke Fujiwara; Keiichi Tachihara; Satoshi Mori; Kentaro Ouchi; Shoko Itakura; Michiko Yasuda; Takashi Hitosugi; Uno Imaizumi; Yoichiro Miki; Izumi Toyoguchi; Kazu-Ichi Yoshida; Takeshi Yokoyama
Journal:  J Clin Monit Comput       Date:  2017-01-10       Impact factor: 2.502

Review 2.  Journal of Clinical Monitoring and Computing 2016 end of year summary: cardiovascular and hemodynamic monitoring.

Authors:  Bernd Saugel; Karim Bendjelid; Lester A Critchley; Steffen Rex; Thomas W L Scheeren
Journal:  J Clin Monit Comput       Date:  2017-01-07       Impact factor: 2.502

3.  Frequency characteristics of pressure transducer kits with inserted pressure-resistant extension tubes.

Authors:  Shigeki Fujiwara; Satoshi Mori; Keiichi Tachihara; Takeshi Yamamoto; Chizuko Yokoe; Uno Imaizumi; Yoshinari Morimoto; Yoichiro Miki; Izumi Toyoguchi; Kazu-Ichi Yoshida; Takeshi Yokoyama
Journal:  J Clin Monit Comput       Date:  2016-03-05       Impact factor: 2.502

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.